Local Exhaust Ventilation Testing Protocols
National Institutes of Health Office of Research Services Division of Occupational Health and Safety Local Exhaust Ventilation Testing Protocols Technical Assistance Branch Reviewed June 2011 i National Institutes of Health Office of Research Services Division of Occupational Health and Safety Local Exhaust Ventilation Testing Protocols CONTENTS DOWNDRAFT SINKS FOR SMALL ANIMAL SURGERY ........................................ ................... 1 Introduction.
inward movement of air violent enough to affect the capture of the majority of the generated smoke. Smoke should be released for capture by the downdraft table at a height of five inches. 1.2.3 Recording Results . After testing of each downdraft sink, a section of the "Downdraft Sink Survey Report" is to be ...