1 ACQUITY uplc H-Class System FEATURESD well volume (total System ) <400 L (includes standard 100 L mixer)Integrated leak management Leak sensors, as standard, and safe leak handlingQuantum synchronization Injection synchronization between pump and sample manager enhances retention time reproducibilitySettable flow rate range to mL/min, in mL increments (firmware version and earlier) to mL/min, in mL increments (firmware version ) to mL/min in mL increments (firmware version and later)Maximum operating pressure 15,000 psi up to mL/min, 9000 psi up to mL/min (firmware version and earlier) 15,000 psi up to mL/min, 7800 psi up to mL/min (firmware version and later)pH range 2 to 12 Unattended operation Leak sensors and safe leak handling, full 96-hour diagnostic data display through console softwareCycle time <30 s inject-to-inject
ACQUITY UPLC H-Class System Instrument Specifications 4 COLUMN MANAGEMENT (CM-A AND CM-AUX) Column capacity CM-A: Two columns, as standard (maximum length of 150 mm with filter or guard column)