baxter s HomeChoice APD System Alarm Troubleshooting* and Prevention TacticsLow Drain Volume Alarm (LDVA)Root Causes Patient reabsorbed more solution than expected or incomplete Last FillPatient empty at start of Initial Drain (I-Drain)I-Drain Alarm setting too high Catheter obstructionPoor catheter positionCycler position too high relative to patient positionFibrin blockageTroubleshooting Check Drain Volume press down arrowChange position to try to drain more fluidCheck for kinks, air or fibrin in patient lineHave you changed your prescription? Have you completed a day drain?Verify cycler position relative to patient supine positionCall dialysis center to help diagnose problem causing LDVAP revention Review or adjust the Initial Drain (I-Drain) Alarm setting if the Last Fill volume, Last Fill dwell time, or if the Last Fill solution type changedDetermine root cause for patient absorbing more solution than expected- Increased sodium intake?
Baxter’s HomeChoice APD System Alarm Troubleshooting* and Prevention Tactics Low Drain Volume Alarm (LDVA) Root Causes • Patient reabsorbed more solution than expected or incomplete Last Fill
Domain:
Source:
Link to this page:
Please notify us if you found a problem with this document:
{{id}} {{{paragraph}}}
PiSector GSM Cellular Wireless Alarm System, Public Address General Alarm for harsh environments, SYSTEM, Alarm, Alarm system, FIRE PROTECTION, Contact ID Format Library, Temperature Controller, OPERATING INSTRUCTIONS, Identiflex 610 Analog / Addressable Fire Alarm, Identiflex 610 analog / addressable fire alarm control