Cavitron JET Plus Troubleshooting and Analysis, continuedThis Troubleshooting section is meant for use by qualified Cavitron Service MEASURESI ntermittent scaling power or no scaling power.( continued )9. Foot Control Connect the auxiliary Foot Control cable between the Foot Control and the unit. If the unit will not operate with the auxiliary cable connected, replace the Foot Control. Follow the Cavitron JET Plus Directions for Use and Service Manual instructions for Foot Control heats Insufficient water to cool Air trapped in the Insert water passageway Handpiece cable not supported during Worn insert being Increase the setting on the handpiece lavage control until handpiece runs When the inserts are changed, hold the handpiece in an upright position until the trapped air is removed and the water flows Replace the Cavitron insert and check Loop handpiece cable around arm or support with finger to prevent water Replace wit
No water flow from handpiece with no insert installed. 1. High dental office water pressure. 1. Install a water pressure regulator on the main water supply line and reduce the pressure to 20-40 psi. Water spray from insert is not properly covering the operating area of the activated tip. 1. Improper water flow adjustment. 2. P-style insert ...