Cavitron JET Plus Troubleshooting and Analysis, continuedThis Troubleshooting section is meant for use by qualified Cavitron Service MEASURESI ntermittent scaling power or no scaling power.( continued )9. Foot Control Connect the auxiliary Foot Control cable between the Foot Control and the unit. If the unit will not operate with the auxiliary cable connected, replace the Foot Control. Follow the Cavitron JET Plus Directions for Use and Service Manual instructions for Foot Control heats Insufficient water to cool Air trapped in the Insert water passageway Handpiece cable not supported during Worn insert being Increase the setting on the handpiece lavage control until handpiece runs When the inserts are changed, hold the handpiece in an
8. Duckbill filters were removed or installed in reverse. 9. Main air regulator in the Cavitron ® JET Plus™ is set too low. 10. Clogged air tubing in the Cavitron ® JET Plus™ or the handpiece. 7. a. Have the dental office air compressor serviced. b. An air dryer and filter should be installed to remove moisture and contaminants.