Cavitron JET Plus Troubleshooting and Analysis, continued

Cavitron JET Plus Troubleshooting and Analysis, continuedThis Troubleshooting section is meant for use by qualified Cavitron Service MEASURESI ntermittent scaling power or no scaling power.( continued )9. Foot Control Connect the auxiliary Foot Control cable between the Foot Control and the unit. If the unit will not operate with the auxiliary cable connected, replace the Foot Control. Follow the Cavitron JET Plus Directions for Use and Service Manual instructions for Foot Control heats Insufficient water to cool Air trapped in the Insert water passageway Handpiece cable not supported during Worn insert being Increase the setting on the handpiece lavage control until handpiece runs When the inserts are changed, hold the handpiece in an upright position until the trapped air is removed and the water flows Replace the Cavitron insert and check Loop handpiece cable around arm or support with finger to prevent water Replace with a new Cavitron insert.

This troubleshooting section is meant for use by qualified Cavitron ... Problem on the PC board(s). 1. Purge mode will only operate with the handpiece empty. Remove the insert and re-press the Purge button. 2. Install a JET-Mate™ on the …