Transcription of CLIA-Compliant Analytical Method Validation Plan and ...
{{id}} {{{paragraph}}}
LRN-C Analytical Method Validation PLAN & TEMPLATE 20131 TABLE OF CONTENTS List of LRN Clients and Requirements for Analytical Method Performance Method Performance Specifications and Validation Analytical Method Validation Plan and TemplateFOR LRN-C LABORATORIESISSUES IN BRIEF: LRN-C Analytical Method Validation PLAN & TEMPLATE DECEMBER 2013 LIST OF ACRONYMSEQA External Quality AssessmentHPLC-MS/MS High-Performance Liquid Chromatography Tandem Mass SpectrometryIS Internal StandardLDT Laboratory Developed Test(s)LOD Limit of DetectionLOQ Limit of QuantitationPK/PD Pharmacokinetic/Pharmacodynamic modelingTEa Total Allowable ErrorULOL Upper Limit of Linearity APHL PUBLIC HEALTH LABORATORIES2 INTRODUCTION The Laboratory Response Networki (LRN)
• calibration verification using linear dynamic range standards, and • analysis of characterized materials with similar concentrations to the sample. Any practice used to “extend” the reportable range must be validated. 6 Note: The calibration model is established in method development and must not be modified during validation or