Data SheetD-TS-V1018_5 Temperature Measurement Products Spring Loaded Sensor Assembly - V10186 Welded Sensor Assembly - V10187 Sensor Assembly Without Thermowell - V10188 Sensor Only - V10189 Spring Loaded Sensor Assembly -V10186 Welded Sensor Assembly - V10187 Sensor Assembly w/o Thermowell - V10188 Sensor Only - V10189 Design- Can be built from standardizedcomponents- Standard lengths for fewer spare partson stock- Immersion lengths can be selectedindividually- Sensor can be replaced during operation- No welding seams coming into contactwith media Technical Features- Approvals according to FM and ATEXfor intrinsically safe installation of thetransmitter- Thermowell materials and designsadapted to operating conditions- Installation of a transmitter in theconnection head eliminates the need formulti-wire circuit- Interference-immune standard outputsignal 4 to 20 mA Applications- Chemical process engineering- Petroleum/natural gas supply andprocessing- Power generation and heat distribution FM Approved Sensor Assemblies- Sensors approved for Intrinsic SafeClass 1, Div 1- Sensors approved for
Resistance Temperature Detectors (RTD) Nominal resistance/Standard/Tolerance Resistance elements with platinum measurement windings are used. In accordance with DIN EN 60 751 the nominal resistance is defined as follows: • 100 ohms at 0 °C • Temperature coefficient 3.85 x 10-3 (K -1) - Averaged between 0 °C and 100 °C.