Page 1 EverFloTM SERVICE & TECHNICAL INFORMATION 1039055, VER. 06 EverFloTM Oxygen Concentrator EverFloTM Q Oxygen Concentrator Service & Technical Reference Manual Page 2 EverFloTM SERVICE & TECHNICAL INFORMATION 1039055, VER. 06 REVISION HISTORY Revision Description Author 06 Updated to include Field Communications Reformatted K. Carter Page 3 EverFloTM SERVICE & TECHNICAL INFORMATION 1039055, VER. 06 Chapter 1. INTRODUCTION .. 5 EVERFLO Oxygen Concentrator OVERVIEW .. 5 SERVICE NOTICE.
EverFlo is functioning properly. A more accurate Oxygen Analyzer may also be used for a more precise measurement of the oxygen purity. As an example, a unit could be measured in the factory with the +/- 1% Oxygen Analyzer to measure 91.1%. If this same unit is then measured with a +/- 2% Oxygen Analyzer, the reading could display a result of 89.1%.