Evita 4. Intensive Care Ventilator Instructions for Use Software 1-178-95. Working with these Instructions for Use Working with these Instructions for Use Header line the of the main chapter Operation Calibrating The title of the specific sub-section is printed underneath the main header to help you find your way quickly from subject to subject. Calibrating the CO2 sensor The CO2 sensor must be calibrated: if the check values are not met on checking calibration with filter or calibration gas. Page as part of the half-yearly inspection of Evita 4. the Instructions for Use in combined text/illustrations. The information is Switch on Evita 4. Wait about for 3 minutes for the machine to complete its warm-up phase. expressed in the form of practical actions, giving the user Press the Calibration key.
5 For Your Safety and that of Your Patients Strictly follow the Instructions for Use Any use of the apparatus requires full understanding and strict observation of these instructions.