29 THE INTERNATIONAL JOURNAL OF METROLOGYJULY AUGUST 2000It is suggested that an Accurate calibration be definedas a calibration within 1%-2% tolerance. This tolerancewas chosen because many instrument manufacturers tolerances fall into this range. Furthermore, the UnitedStates Pharmacopeia guidelines [1] under the USP 24 Monographs require a cell calibration within a 2%tolerance. [2] Key industries affected by such issues arepharmaceutical, biotechnology, semiconductor, andpower generation. This paper contains fresh experimentaldata, determines the accuracy of some low-conductivitystandards, and evaluates the feasibility of their launched this study in response to anincredible amount of confusion and controversy on thepart of end-users concerning
29 THE INTERNATIONAL JOURNAL OF METROLOGY JULY • AUGUST 2000 It is suggested that an accurate calibration be defined as a calibration within 1%-2% tolerance. This tolerance
Domain:
Source:
Link to this page:
Please notify us if you found a problem with this document:
{{id}} {{{paragraph}}}
Tape: Linearization of Roughness Measurements, Sensor, National Instruments, Accurate, PH & Conductivity Measurements for Scrubbers, Thermocouples, Contact surface charge/voltage measurements, Measurements, Wavetronix SmartSensor HD, Delta P Sensor and Barometric Sensor, NFC Sensor Tag Application Example, Texas Instruments, PH Measurement To Determine Acidification of