METHOD 6010C INDUCTIVELY COUPLED PLASMA …

6010C - 1 Revision 3 November 2000 METHOD 6010 CINDUCTIVELY COUPLED PLASMA - atomic EMISSION SCOPE AND APPLICATION COUPLED PLASMA - atomic emission spectrometry (ICP-AES) may be usedto determine trace elements in solution. The METHOD is applicable to all of the elements listedbelow. With the exception of groundwater samples, all aqueous and solid matrices require aciddigestion prior to analysis. Groundwater samples that have been prefiltered and acidified will notneed acid digestion. Samples which are not digested require either an internal standard or shouldbe matrix-matched with the standards. If either option is used, instrument software should beprogrammed to correct for intensity differences of the internal standard between samples andstandards.

6010C - 1 Revision 3 November 2000 METHOD 6010C INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY 1.0 SCOPE AND APPLICATION 1.1 Inductively coupled plasma-atomic emission spectrometry (ICP-AES) may be used