Sample Channel ZoomPPC (Spike) Channel ZoomSTORAGE CONDITIONS AND PRECAUTIONSPTS Cartridges are relatively heat stable and should be stored at 2 - 25 C. Allow the Cartridges to come to room temperature beforeopening the pouch and testing. Prolonged exposure to temperaturesabove 25 C should be avoided. To minimize contamination of thesample wells, the cartridge should be used immediately once the foilpouched seal has been opened. Cartridges are for single-test use REQUIRED BUT NOT SUPPLIEDLAL Reagent Grade Water must be used during the initial qualificationof each lot of PTS REQUIRED BUT NOT SUPPLIEDP ipettor (Endosafe PTS 400 or equivalent) and sterile , endotoxin-free glass dilution tubes or sterile, disposablepolystyrene tubes (Endosafe T300 or equivalent) for sample collection or dilution if Mixer (if necessary).EQUIPMENT REQUIRED BUT NOT SUPPLIEDE ndosafe Portable Test System (PTS) Reader:The reader is a dedicated instrument that accepts the cartridge and runs the PTS LALtest.
Inhibition is usually concentration dependent and can be overcome by dilution with LAL Reagent Water. The most common sources of inhibition are 1) conditions that interfere with the enzyme activity due to