Blood BankingCentrifugationChemistryFlow CytometryHematologyHemostasisImmunoassay Information SystemsLab AutomationMolecular DiagnosticsRapid DiagnosticsBR-16544 AQuality Control (QC) Information and Troubleshooting Guide UniCel DxC Clinical Chemistry Systems UniCel DxC Integrated Systems1 Quality ControlIntroductionQuality controls are samples of known concentration that are used to monitor system performance and verify calibration. The system presents Quality Control (QC) data in both a summary format and in a chart format. It applies Westgard QC rules to the data to help determine assay and system QC samples when your laboratory protocol indicates that Control material should be analyzed. Beckman Coulter recommends a daily analysis of at least two levels of Control materials. You should also run controls with each new calibration with each new lot of reagents after specific maintenance or Troubleshooting laboratory should determine its own Control frequency based on the CLSI (formerly NCCLS) Guideline C24-A3 STATISTICAL Quality Control FOR QUANTITATIVE MEASUREMENT flag DeterminationThe DxC uses the Z-score method to standardize the scale of a normally distributed measurement variable.
QC Flag Determination ... — displays the result in red on the QC LOG and QC CHART screens. ... not complete, the system displays a warning message specifying the missing information. • If a chemistry is missing a QC File Number, you can select OK to delete the chemistry, or