PDF4PRO ⚡AMP

Modern search engine that looking for books and documents around the web

Example: air traffic controller

Retractable pH/ORP Sensor - Emerson

Model 396 RInstruction SheetPN 51A-396 2011 SPECIFICATIONSMODEL 396 RMeasurements and RangespH: 0-14; ORP: -1500 to 1500 mV*Available pH ACCUGLASST ypesGPLR hemi bulb or GPLR flat bulbWetted MaterialsPolypropylene, EPDM, titanium, glass, (platinum: ORP only)Process ConnectionNone, use 1 inch process connector or ball valve kit (1-1/2 inch or 1-1/4 inch sold separately)Temperature Range0 to 100 C (32 to 212 F)Pressure Range - Hemi bulb100 to 1136 kPa (abs) (0 to 150 psig)Pressure Range - Flat bulb100 to 790 kPa (abs) (0 to 100 psig)Maximum Pressure at Retraction Code 21: 542 kPa (abs) (64 psig) or InsertionCode 25: 343 kPa (abs) (35 psig) Minimum Conductivity100 S/cm, nominalPreamplifier OptionsRemotePERCENT LINEARITY FOR MODEL 396 RpH rangeGPLR Hemi bulbGPLR Flat bulb0-294%93%2-1299%98%12-1397%95%13-149 2% Retractable SENSORSR etractable sensors must notbe inserted nor retractedwhen process pressures are in excess of 64 psig(542kPa) for option 21 or 35 psig (343 kPa) for o

iNSTALLATION For sensor dimensions, see Figures 1 and 2. For sensor orientation and installation, see Figures 3 - 5. For wiring, see Figures 6 - 20.

Loading..

Tags:

  Sensor, Ph orp sensor

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Spam in document Broken preview Other abuse

Transcription of Retractable pH/ORP Sensor - Emerson

Related search queries