Transcription of Technical Information Nivotester FTC325 - …
{{id}} {{{paragraph}}}
TI380F/00/ InformationNivotester FTC325 Level limit switchWith Intrinsically Safe Signal Circuit for Connectionto a Capacitance Sensor Applications Level limit detection in tanks containing liquids and silos containing bulk solids. For capacitance level probes, which may also be applied in hazardous areas of category ATEX II (1) GD Overspill protection for tanks containing flammable or non-flammable fluids hazardous to water Dry running protection for pumps Two-point control ( s with 3- wire )Your benefits Intrinsically safe signal circuit [EEx ia] IIC for using sensors in hazardous areas Calibration using operating keys High functional safety thanks to.
Nivotester FTC325 6 Endress+Hauser Probe construction Level limit detection with FTC325 PFM L00-FTC625xx-05-06-xx-en-000 Nivotester FTC325 PFM Nivotester FTC325 3-WIRE
Domain:
Source:
Link to this page:
Please notify us if you found a problem with this document:
{{id}} {{{paragraph}}}