Troubleshooting and DiagnosticsTips and Tricks 2013 Waters Corporation1 Cerdanyola del Vall s, Enero 2014 tips and Tricks Troubleshooting and diagnosticsTroubleshooting and Diagnostics Troubleshooting System Pressure Problems Incorrect Retention time Loss of precision Carryover/Contamination 2013 Waters Corporation2 Carryover/Contamination Split and Distorted Peaks Baseline NoiseTroubleshooting and diagnosticsTroubleshooting and Diagnostics Troubleshooting System Pressure Problems Incorrect Retention time Loss of precision Carryover/Contamination 2013 Waters Corporation3 Carryover/Contamination Split and Distorted Peaks Baseline
System Pressure problems Overpressure –Check if something has changed (column, mobile phase, temperature) –If nothing has changed, remove the column and replace it with a union to check if the system pressure is the usual. –If system pressure is high loosen fittings beginning with the last connection in line and working backward to the pump