Transcription of USER GUIDE AND SPECIFICATIONS USB-6008/6009
{{id}} {{{paragraph}}}
user GUIDE AND SPECIFICATIONSUSB-6008/6009 This GUIDE describes how to use the National Instruments USB-6008/6009 data acquisition (DAQ) devices and lists NI USB-6008/6009 provides connection to eight analog input (AI) channels, two analog output (AO) channels, 12 digital input/output (DIO) channels, and a 32-bit counter with a full-speed USB manual revision updates naming conventions to reflect the conventions used in NI-DAQmx. Table 1 notes the correlation between the old and updated 1. Digital Output Driver Type Naming ConventionsHardware FunctionalityNI-DAQmx TerminologyOpen-drainOpen collectorPush-pullActive driveTable 2. Differences Between the USB-6008 and USB-6009 FeatureUSB-6008 USB-6009AI Resolution12 bits differential,11 bits single-ended14 bits differential,13 bits single-endedMaximum AI Sample Rate, Single Channel*10 kS/s48 kS/sMaximum AI Sample Rate, Multiple Channels (Aggregate)*10 kS/s42 kS/sDIO ConfigurationOpen collectorOpen collector or active drive* Might be system user GUIDE and 1.
USER GUIDE AND SPECIFICATIONS USB-6008/6009 This guide describes how to use the National Instruments USB-6008/6009 data acquisition (DAQ) devices and lists specifications.