Transcription of Humidity and Dew-Point Instruments Easidew I.S.
1 Humidity and Dew-Point Transmitter(For Hazardous Area Applications)The Easidew (intrinsically safe) transmitter is a Dew-Point transmitter designed and certifi ed for use in hazardous area applications (fl ammable or explosive gases) in all regions with certifi cation from: IECEx, CSA, FM and Easidew is available with a choice of G1/2" BSP, 3/4" or 5/8" UNF process connections. It is designed for ease of use, incorporating all of the features needed to make installation and operation into your air or gas management system as simple as 's Advanced Ceramic Moisture Sensor technology based transmitter is calibrated to international standards and is delivered with a traceable calibration certifi IECEx, CSA, FM, ATEX certifi ed transmitter for use in hazardous areas G1/2" BSP, 3/4" or 5/8" UNF process connection Dew-Point or ppmV moisture content 2-wire loop powered connection Rugged 316 stainless steel IP66 construction Measurement range -100 to +20 C Accuracy 2 Cdp Calibration Certifi cate (NPL, NIST)Technical Specifi cationsPerformanceMeasurement range-100 to +20 C dew pointAccuracy 2 C dew pointResponse time5 mins to T95 (dry to wet) C dew pointCalibrationTraceable 13 point calibrationElectrical Specifi cationsOutput signal4-20 mA (2-wire connection, current source)
2 User confi gurable over rangeOutputDew point or moisture content for ppmVAnalog output scaled rangeDew point: -100 to +20 CMoisture content in gas: 0-3000 ppmVNon-standard available upon requestSupply voltage12 to 28 V DCLoad resistanceMax 250 @ 12 V (500 @ 24 V)Current consumption20 mA maxCE markedCertifi edOperating Specifi cationsOperating temperature-40 to +70 COperating pressure45 MPa (450 barg) maximumOverpressure ratingx2 operating pressure 90 MPa (900 barg)Flow rate1 to 5 Nl/min mounted in standard sampling block; 0 to 10m/sec direct insertionTemperature coeffi cientTemperature compensated across operating temperature rangeMechanical Specifi cationsIngress protectionIP66 in accordance with standard BS EN 60529:1992,NEMA 4 in protection accordance with standard NEMA 250-2003 Hazardous area certifi catesATEX - II 1 G Ex ia IIC T4 (-20 C Ta +70 C)FM - IS / I / I / ABCD / T4 Ta = +70 CCSA - IS Class I Div I Groups ABCD T4 IECEx - Ex ia IIC T4 (-20 C Ta +70 C)Housing material316 stainless steelDimensionsL=132mm x 45mm (with connector)Filter (sensor protection)Standard: HDPE Guard <10 mOptional.
3 316 stainless steel sintered guard <80 mProcess connection and materialG1/2 BSP, 3/4 - 16 UNF, 5/8 - 18 UNF316 stainless steelWeight150gInterchangeabilityFully interchangeable transmitterElectrical connectionsHirschmann GDS series (DIN 4350-C)Diagnostic conditions(factory programmed)ConditionSensor faultUnder-range dew pointOver-range dew pointOutput23 mA4 mA20 mAApproved galvanic Michell Instruments 2013 Humidity and Dew-Point InstrumentsPlease note: Michell Instruments adopts a continuous development program which sometimes necessitates specifi cation changes without notice. Issue No: Easidew IS_97168_V4_UK_0113 Easidew Codes To construct the order code, select the relevant feature from the tables below, starting with the base model, which is {Feature A} and then add on options to create a string: {Feature A}+{Feature B}+{Feature C}+{Feature D}Base Model {Feature A}Base unit. Easidew Hazardous Area Transmitter 2-wire 4-20 mAEA2-IS-TXAccessories and Spare PartsProcess Connection SizeG1/2 3/4 5/8 Sensor ProtectorsPack of 10 replacement HDPE guards for protection against fi ne particulate <10 mEA2-HDPEEA2-HDPEEA2-HDPES tainless steel sintered guard replacement for protection against fi ne particulate <80 mSSGSSGSSGPack of 10 particulate fi lter cartridges for protection against fi ne particulate <25 m (for Sampler kits)SSF-PF-10 PKSSF-PF-10 PKConnection AdaptorsG 1/2" BSP to 5/8" UNF adaptorAPT-GEI1/2" NPT to 5/8" UNF adaptorAPT-TSO3/4" NPT to 5/8" UNF adaptorAPT-PANS ampling BlocksG1/2 316 stainless steel sample block (with 1/8" NPT inlet & outlet ports)SB-12G1/2 316 stainless steel sample block with fi lter (with 1/8 BSP & 1/4 BSP ports)A0003513/4" 316 stainless steel sample block with (1/8 NPT inlet & outlet ports)
4 SB-345/8 316 stainless steel sample block (with 1/8 NPT inlet & outlet ports)CSB5/8 Sampler kit For atmospheric or pressure dp measurements (145 psig/10 barg max)EA2-SAM 5/8 HP sampler kit For atmospheric or pressure dp measurements(2900 psig/200 barg max) galvanic isolator (for Easidew - Pepperl & Fuchs)GI-PF-01 GI-PF-01GI-PF-01 Other PartsPTFE sample tube 6mm outside diameter, thick-walled (for sampler kit) (if more than the standard is required)PTFE-xxPTFE-xxCable with mini-DIN mating socket (for NON sensors)EA2-CAB-08EA2-CAB-08EA2-CAB-08 Additional cable total length (if more than the standard is required)EA2-CAB-xxEA2-CAB-xxEA2-CAB-xxA dditional/replacement transmitter labels - 2 labelsGEN-LABELGEN-LABELGEN-LABELO-rings /Bonded seals for process connection 5 seals1/2-BS3/4-OR5/8-BSSensor connectorEA2-SCEA2-SCEA2-SCEasidew communications kit for EA2EA2-CKEA2-CKEA2-CKElectrical Connections 4-20 mA connections 2-wirePin 14-20 mA Pin 3 POWER Connection {Feature B}G1/2 BSP process connectionB13/4 UNF process connectionB25/8 UNF process connectionB3 Protection {Feature D}Standard HDPE guard (with protection against fi ne particulate (<10 m))D1 Stainless steel sintered guard D2 Range {Feature C}-100 to +20 Cdp output rangeC1 Non-standard measurement range: v = zerovalue, w = full scale value, x = unit (C = Cdp,F= Fdp, P=ppmV) y = Pressure for ppmV conversion in bar/psi, Z = pressure unit (PG=psig, PA=psia, BG=barg, BA=bara)blank = 0 barg/0 psig)CX(.
5 132mm( )46mm( )10mm( )5/8 UNF10mm( )hhhggg 27mm( )g5/8 - 18 UNF Bonded Seal x 2mm( 1 x )27mm( ) A/F45mm( )gggg132mm( )46mm( )10mm( )G1/2 BSP10mm( )hhhggg 27mm( )g27mm( ) A/F45mm( )ggggG1/2 BSP Bonded Seal x ( x )132mm( )46mm( )10mm( )10mm( )hhhg27mm( ) A/F45mm( )gggggg 27mm( )3/4 UNFg3/4 - 16 UNF O-Ring x ( x )Order example: EA2-IS-TX+B1+C1+D2 Easidew Hazardous Area, 2 wire, 4-20 mA output, G1/2" BSP process connection, -100 to +20 Cdp range, stainless steel sintered guard5/8" Process ConnectionG 1/2" Process Connection3/4" Process ConnectionDimensions EA2-IS-TXB1C1 D2