10 K Closed Cycle Refrigerator Systems

cryo . INDUSTRIES. 10 K Closed Cycle Refrigerator Systems Sample in Vacuum Sample in Top-Loading Exchange Gas cryo Industries of America, Inc. 11124 S. Willow St. Manchester, NH 03103. Tel: (603) 621-9957 Fax: (603) 621-9960 E-mail: cryo . INDUSTRIES. cryo Industries manufactures Closed Cycle cryogenic workstations for a large variety of experimental requirements. Described below are three different standard styles. Systems can be provided with special sample T' style mounts, temperature controllers and sensors.

CRYO Industries of America, Inc. • 11124 S. Willow St. • Manchester, NH 03103 Tel: (603) 621-9957 • Fax: (603) 621-9960 • E-mail: sales@cryoindustries.com 10 K Closed Cycle Refrigerator Systems Sample in Vacuum Sample in Top-Loading Exchange Gas

Tags:

  System, Cycle, Closed, Refrigerator, Cryo, K closed cycle refrigerator systems

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of 10 K Closed Cycle Refrigerator Systems

1 cryo . INDUSTRIES. 10 K Closed Cycle Refrigerator Systems Sample in Vacuum Sample in Top-Loading Exchange Gas cryo Industries of America, Inc. 11124 S. Willow St. Manchester, NH 03103. Tel: (603) 621-9957 Fax: (603) 621-9960 E-mail: cryo . INDUSTRIES. cryo Industries manufactures Closed Cycle cryogenic workstations for a large variety of experimental requirements. Described below are three different standard styles. Systems can be provided with special sample T' style mounts, temperature controllers and sensors.

2 Custom Systems are no problem! Style 1: Sample in Vacuum Mounting Style 3: Sample in Exchange Gas Designed for cold finger' - Accepts variety of bolt in/out experimental sample mounting. Easy inserts sample access through quick - Removable Refrigerator connect vacuum shroud and - Interchangeable open Cycle liquid helium flow single thread radiation shield. Refrigerator available. A large variety of standard shrouds and T' style mounts Style 3 has all the features of Style 2 plus more! are available. The models Accepts a large variety of easy to bolt-on' inserts.

3 REF-396-C22 (Standard) and No indium or solder seals. One system can REF-516-DC2 (compact) are accept many different removable inserts, such typical optical configurations. as compact mail' for magnet gap experiments. Temperature range is from In addition, the <9 to 325 K. An optional high Refrigerator can be temperature stage extends removed for cold finger'. operation to 600 K. or other applications. Even extend the operating range to Sample 2: Sample in Exchange Gas less than 4 K with our Samples are Top Loading' and can be changed interchangeable model in seconds without shutting off or warming the RC102/110 REF liquid Refrigerator .

4 Interchange samples in seconds helium flow Refrigerator . without thermal or vacuum Avoid unreliable, cycling of the system . clogging and expensive Exchange gas cooling of J-T (Joule-Thomson). low thermal conductive third stages usually samples. Independent needed to get near 4. sample mount heater results K. For those special in fast thermal response. cases or new future Ulta-stable variable requirements you won't temperature operation. need to try to borrow a Independent temperature liquid helium cryostat control of sample mount or miss those critical data points.

5 ONE cryostat and Refrigerator cold finger. can do it all. The Model REF-469-D offers optical The Mode REF-452-D22 transmission (two windows at 180 degrees is ideal for materials apart). Other optical configurations and different characterization. The Model window materials are available. A slim body REF-399-D22 (Mossbauer) superconductor characterization system is REF- features low vibration 468-DSC. Systems are available for upgrading to a through bellows, spacer and Top Loading' cryostat using your existing Closed exchange gas isolation Cycle Refrigerator .

6 cryo Industries of America, Inc. 11124 S. Willow St. Manchester, NH 03103. Tel: (603) 621-9957 Fax: (603) 621-9960 E-mail: cryo . INDUSTRIES. cryo Industries has 25 years experience in designing and manufacturing Closed Cycle Refrigerator Systems . This has resulted in a high level of expertise and over 375 field tested and proven designs of Closed Cycle Refrigerator Systems . cryo Industries pairs innovative design with leading edge technology, by integrating cryocoolers that offer the most reliable and longest track records on the market into the Closed Cycle Refrigerator Systems .

7 cryo is able to incorporate any of the following Cryocoolers into our designs. You simply select the cooling power you need! 10 K Sumitomo (SHI) Cryocooler Cooling Power Specification Chart Model 1st Stage @ 50 Hz 2nd Stage @ 50 Hz 1st Stage at 60 Hz 2nd Stage at 60 Hz DE-202 @ 77 K @ 20 K @ 77 K @ 20 K. DE-204SL @ 80 K @ 20 K @ 80 K @ 20 K. CH-204S @ 80 K @ 20 K @ 80 K @ 20 K. CH-208R 65W @ 77 K 6W @ 20 K 80W @ 77 K @ 20 K. CH-208L 28W @ 77 K 8W @ 20 K 35W @ 77 K 10W @ 20 K. CH-210 110W @ 77 K 6W @ 20 K 110W @ 77 K 6W @ 20 K.

8 Note: DE-202 and DE-204SL use a single phase HC4E-1 water cooled compressor Note: CH-204S uses either a single or 3 Phase water cooled compressor (customer must specify). Note: CH-208 and CH-210 requires a 3 Phase (Low or High Voltage) F70 water cooled compressor 10 K CTI Cryocooler Cooling Power Specification Chart Model 1st Stage @ 50 Hz 2nd Stage @ 50 Hz 1st Stage at 60 Hz 2nd Stage at 60 Hz M-22 8W @ 77 K 1W @ 20 K. M-350 20W @ 77 K 4W @ 20 K. M-1020 34W @ 77 K 12W @ 20 K. Note: Both M-22 and M-350 use Model 8200 Single Phase Water-Cooled Compressor Note: M-1020 requires Model 9600 Three Phase Water-Cooled Compressor cryo Industries of America, Inc.

9 11124 S. Willow St. Manchester, NH 03103. Tel: (603) 621-9957 Fax: (603) 621-9960 E-mail: cryo . INDUSTRIES. Closed Cycle Refrigerator system : Sample in Vacuum Optical MODEL DE-202FF. Refrigerator . INSTRUMENTATION. HOUSING. THERMAL ANCHOR. WITH WIREWAYS..062 QUICK DISCONNECT. FOR SAMPLE ACCESS. DIA. SAMPLE MOUNT. WITH HEATER, TAPPED. MOUNTING HOLES AND. PROVISION FOR. TEMPERATURE SENSOR. COPPER SAMPLE HOLDER..44. This design features Sumitomo (SHI) Model DE-202. If higher cooling power is needed any BASE PLATE WITH.

10 SUPPORT POSTS. of the cryocooler listed on the Cooling Power Chart can be .38. integrated into this design. cryo INDUSTRIES. of America Inc. 11124 S. Willow St., Manchester 03103. PROPRIETARY INFORMATION TITLE: THE INFORMATION CONTAINED IN THIS DRAWING. IS THE SOLE PROPERTY OF cryo INDUSTRIES. OF AMERICA, INC., 11124 S WILLOW STREET. DE-202 Closed Cycle . MANCHESTER, 03103. ANY REPRODUCTION. IN PART OR WHOLE WITHOUT THE WRITTEN. Refrigerator system . PERMISSION OF cryo INDUSTRIES IS STRICTLY. PROHIBITED.