Analysis of Residual Solvents using GC/FID with Headspace ...

Analysis of Residual Solvents using GC/FIDwith Headspace and a Cyanopropylphenyl Polysiloxane PhaseA. Khan, L. Pereira, Thermo Fisher Scientific, Runcorn, Cheshire, UKOverviewA Thermo Scientific TRACE Ultra GC system with aThermo Scientific TriPlus Autosampler was used for theanalysis of Residual Solvents in accordance with USPmethod column used for the Analysis was aThermo Scientific TRACE TR-V1 This application note describes the Analysis of Residual Solvents according to the revised (effective July 2007) USPharmacopeia (USP) method. The method is used for analyzing trace levels of Solvents that are involved in theproduction of a drug, excipient or product packaging in thepharmaceutical industry. Drug products should contain lowlevels of Residual Solvents as determined by safety data. The USP 467 method involves the Analysis of 53 Solvents grouped according to their genotoxic hazards: Class 1 Solvents (known to cause unacceptable toxicities)should be avoided in the manufacturing process.

This application note describes the analysis of residual solvents according to the revised (effective July 2007) US Pharmacopeia (USP) method. The method is used for analyzing trace levels of solvents that are involved in the production of a drug, excipient or product packaging in the pharmaceutical industry. Drug products should contain low

Tags:

  Analysis, Pharmaceutical, Production

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Analysis of Residual Solvents using GC/FID with Headspace ...

1 Analysis of Residual Solvents using GC/FIDwith Headspace and a Cyanopropylphenyl Polysiloxane PhaseA. Khan, L. Pereira, Thermo Fisher Scientific, Runcorn, Cheshire, UKOverviewA Thermo Scientific TRACE Ultra GC system with aThermo Scientific TriPlus Autosampler was used for theanalysis of Residual Solvents in accordance with USPmethod column used for the Analysis was aThermo Scientific TRACE TR-V1 This application note describes the Analysis of Residual Solvents according to the revised (effective July 2007) USPharmacopeia (USP) method. The method is used for analyzing trace levels of Solvents that are involved in theproduction of a drug, excipient or product packaging in thepharmaceutical industry. Drug products should contain lowlevels of Residual Solvents as determined by safety data. The USP 467 method involves the Analysis of 53 Solvents grouped according to their genotoxic hazards: Class 1 Solvents (known to cause unacceptable toxicities)should be avoided in the manufacturing process.

2 Class 2 Solvents (associated with less severe toxicity)should be limited. Class 3 Solvents (less toxic) should be used where analytical procedures are used for identificationand quantification of the Residual Solvents : Procedure Ais used for screening and confirmation ofthe Solvents identity and use a G43 (volatiles) column. Procedure Bis used for confirmation of the solventsidentity using a G16 (wax) column. Procedure C, which uses a G43 column, is required toquantify the amount of Residual SamplesUSP 467 test mixtures solutions prepared according to USP 467 methodColumnPart NumberTRACE TR-V1 30 m mm m260V339 PGC/FID ConditionsTriPlus Headspace AutosamplerSample Volume1 mLSample Analysis Time30 minAgitator Temperature80 CIncubation Time45 minAgitator ShakeOn 10 s, Off 20 sSyringe Temperature100 CPost Injection Flush30 sTRACE GC UltraOven Program40 C (20 min), 10 C/min, 240 C (10 min)Equilibration CSplit Flow40 mL/minColumn mL/min (constant)

3 DetectorFID at 240 C Detector Air Flow350 mL/minDetector H2 Flow35 mL/minDetector N2 Makeup Flow30 mL/minConsumablesPart NumberTR-Green septa313G3211 Thermo Scientific Split FocusLiner for 50 mm needle with Quartz Wool453T1905 Liner graphite seal29033406 Graphite ferrule for mm ID mL Headspace syringe3650300620 mL clear crimp top vial60180-506 Aluminum 20 mm cap and Si/PTFE seal60180-511 Key Words TRACE TR-V1 Column Residual Solventsin PharmaceuticalProducts USP 467 methodApplication Note:ANCCSRESSOLV1010 Results and DiscussionThe acceptance criteria for USP 467,class 1 Solvents , is that the signal-to-noise ratio for 1,1,1-Trichloroethane(peak 2) is greater than 5 and that thesignal-to-noise ratio obtained fromthe other compounds in the class isgreater than 3. The chromatogramand signal to noise measurementsobtained using the method describedpreviously are shown in Figure USP 467 method specifica-tion for class 2 Solvents is that theresolution between acetonitrile anddichloromethane (peaks 2 and 3) isnot less than The chromatogramsobtained for class 2 Solvents (mixtureA and mixture B) are shown inFigures 2 and 3.

4 The method criteriafor resolution between the criticalpair is met as these are fully the other components are alsofully resolved, which demonstratesgood selectivity and efficiency of thechromatographic separation takingplace in the TRACE TR-V1 response was observed for 1,4-Dioxane, which was 1: GC/FID chromatogram for class 1 Residual solventsMeasuredSpecifiedPeak #Peak IdentificationRt (min)S/NS/N11, > 321,1, > 53 Carbon > > 351, > 3 Figure 2: GC/FID chromatogram for class 2A Residual solventsPeak #Peak IdentificationRt (min) , , , criteria set by the USP 467 method is met, thereforethe TRACE GC Ultra with Headspace configuration andTRACE TR-V1 column provides excellent capability torun the Residual solvent method, screening signal-to-noise specifications set in USP 467 forClass 1 Residual Solvents were easily exceeded, demon-strating that the set-up used provides a sensitive systemfor this resolution specifications set in USP 467 for Class 2 Residual Solvents were easily exceeded demonstratinggood selectivity and efficiency of the chromatographic separation taking place in the TRACE TR-V1 column.

5 References1. Residual Volatile Impurities: USP 467, 2008. USAIn addition to these offices, Thermo FisherScientific maintains a network of represen -tative organizations throughout the and Canada+1 800 332 3331 France+33 (0) 3 88 67 53 20 Germany+49 6103 408 0 Switzerland+41 56 618 41 11 United Kingdom+44 1509 555500 Japan+81 45 453 9220 China800-810-5118 India1800 22 8374 (toll-free)+91 22 6716 2200 All Other Enquiries+44 (0) 1928 534 050 TechnicalSupportNorth America+1 800 332 3331 Outside NorthAmerica+44 (0) 1928 534 2010 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for 1010 Part of Thermo Fisher ScientificPeak #Peak IdentificationRt (min) , 3: GC/FID chromatogram for class 2B Residual Solvents