AUTOCLAVE TECHNICAL SPECIFICATION 44 and 55 Compact ...

AUTOCLAVE TECHNICALSPECIFICATION44 and 55 Compact Sterilizer Sterilization & Infection Control PartnersPage1of10 FEATURE SUMMARY Chamber volume from 120 to 310 liters Design pressure meets ASME and PED requirements Temperaturerange 105 C (221 F)to 138 C (280 F) 18 kW/ 27 kWintegral electrical steam generator orexternal steam source User friendly control system with touch screen display 30 Programs: 8factory set programs, 2 test programs, 20programmable cycle programs Built-in printer Ethernet connection port for PC access via network USB port to download cycle data to memory device Pressure gauges on front panel 316L stainless steel chamber and door Chamber has a mirror-like finish Stainless steel piping Conforms to Medical Device Directive 93/42 EEC andPED 97/23 EC, FDA Clearance Conforms to standards: ASME, AAMI/ANSI-ST8, EN 285,UL Company approvedfor 21 CFR 820, ISO 9001:2008 andISO 13485:2003

AUTOCLAVE TECHNICAL SPECIFICATION 44 and 55 Compact Sterilizer Series Rev. 23-01-2014 Page 3 of 10 CHAMBER CONSTRUCTION Materials The sterilizer chamber is constructed from solid, high quality materials. Chamber and door thickness is no less than 5 mm. Standard Configuration Materials: Chamber + Door: 316L stainless steel Jacket: 304L stainless ...

Tags:

  Specification, Construction, Technical, Technical specifications

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of AUTOCLAVE TECHNICAL SPECIFICATION 44 and 55 Compact ...

1 AUTOCLAVE TECHNICALSPECIFICATION44 and 55 Compact Sterilizer Sterilization & Infection Control PartnersPage1of10 FEATURE SUMMARY Chamber volume from 120 to 310 liters Design pressure meets ASME and PED requirements Temperaturerange 105 C (221 F)to 138 C (280 F) 18 kW/ 27 kWintegral electrical steam generator orexternal steam source User friendly control system with touch screen display 30 Programs: 8factory set programs, 2 test programs, 20programmable cycle programs Built-in printer Ethernet connection port for PC access via network USB port to download cycle data to memory device Pressure gauges on front panel 316L stainless steel chamber and door Chamber has a mirror-like finish Stainless steel piping Conforms to Medical Device Directive 93/42 EEC andPED 97/23 EC, FDA Clearance Conforms to standards: ASME, AAMI/ANSI-ST8, EN 285,UL Company approvedfor 21 CFR 820, ISO 9001:2008 andISO 13485.

2 2003 (Medical Devices) Optional configurations availablePRODUCT DESCRIPTIONThis AUTOCLAVE series is designed to cover a large field ofapplications for hospitals, CSSD and medical AUTOCLAVE series has a chamber volume range from 120liters to 310 AUTOCLAVE operates with saturated steam as the sterilizingagent with a temperaturerange from 105 C (221 F) to 138 C(280 F)anda workingpressurethat meetsASMEand AUTOCLAVE is equipped withan18kWor27kWintegralelectrical steam generator which supplies thenecessary steam for the sterilization sterilizermay also be connected to an external steam SterilizingSupply Department (CSSD)

3 Operating TheaterOut Patient ClinicDEVICE PICTURE44 Compact Serieswith Hinged Door44 Compact Series with Vertical Sliding DoorInformation furnished by Tuttnauer is believed to be accurate and , no responsibility is assumed by Tuttnauer for its SPECIFICATION is subject to change without SPECIFICATION44 and 55 CompactSterilizer SPECIFICATIONCHAMBER VOLUME & SIZEMODELCHAMBER DIMENSIONS(mm)VOLUME(liter)WidthHeightDe pth4472408408730120448084514044969701605 596508508970250551201210310 DOOR SELECTION Single door Double door (pass through) Manual right hinged Manual left hinged Automatic vertical sliding (door on left,controlon right) Automatic vertical sliding (door on right,controlon left)All automatic doors have a robust hydraulic mechanism foroperating ACCESSFrom loading side and left or right SELECTIONS Installed in a wall[Single Door] Cabinet enclosure[Single Door] Cabinet enclosure through one wall[Double Door] Installed through two walls[Double Door]ELECTRICAL COMPONENTSThe following parts are located in theelectric boxes: The Solid State Relays for the command of the vacuumpump, and the water pump.

4 The Circuit Breakers for protection of all power circuits. The power supplies for powering of electronic circuitry andsolenoid valves. An overloadswitch is used to protect the device from highoverload currents. The connection elements (sockets, plugs, etc) andcomponents have a degree of protection conforming to theIPX4 protection level and conform to the EN 60529:1991standardVOLTAGE SUPPLYT hree phaseor one phaseelectrical powersupply is : 415V 3-Phase with Neutral 380V / 400V 3-Phase with Neutral 230V 3-Phase with Neutral 208V 3-Phase with or without Neutral 220V / 230V 1-Phase (not 55120 model) 110V / 115V 1-Phase ** For 18kW with building steam supply(optionalvacuum by ejector)Frequency.

5 50Hz or 60 HzMODELSTEAM GENERATORPOWER (kW)4472184480449655965512027 Customized electrical specifications are available (additionalcost).LANGUAGEThe operator display is available in 26 languages including:English, Spanish, German, Dutch, Russian, Chinese, AND CODEST uttnauer products meet the following provisions andstandards:Europe:EN285: 2006+A2:2010 for Large AutoclavesUSA:ANSI/AAMI-ST 8: 2008 Hospital SterilizersDirectives & Guidelines:97/23/EC-Pressure Equipment Directive2002/95/EC-RoHS Directives2004/108/EC-Electromagnetic compatibility2006/95/EC-Electrical equipment2006/42/EC-Machinery DirectiveSafety and EMC Standards:EN 61010-1: 2001EN 61010-2-040: 2005EN 61326-1: 2006 EMC Directive 89/336EN 60529.

6 1991-Degrees of Enclosure Protection (IPX4)Pressure Vessel and Steam Generator ConstructionStandards:ASME Code, Section VIII,Division 1, unfired pressure vesselsASME Code, Section I, for boilersGood Practice Standards:ISO 17665-1: 2006 ANSI/AAMI ST 8: 2008 Quality System Standards:ISO 9001:2008 (Quality Systems)ISO 13485:2003 (Quality Systems for Medical Devices)Compliance with 21 CFR 820 AUTOCLAVETECHNICAL SPECIFICATION44 and 55 CompactSterilizer CONSTRUCTIONM aterialsThe sterilizer chamber is constructed from solid, high qualitymaterials. Chamber and door thickness is no less than 5 Configuration Materials:Chamber + Door:316L stainless steelJacket:304L stainless steelOptional Materials (Stainless Steel):Chamber + Door:316 TiJacket:316L or 316 TiChamber DesignThe chamber isfully stainless steel mesh strainerprotects the drain port from blockage by debris.

7 The chamberis mounted on a steel framework which is height hinged door gasket: a silicone gasket is permanentlyfixed in the sliding door gasket: a silicone gasket is fixed in agroove in the door frame and the gasket is sealed against thedoor by means of steam TreatmentA passivation layer is applied to the internal surface through anelectro-chemical treatment which results in a smooth and shinystainless steel. The resultant surface is polished to less m which is highly protected against corrosion. Theinternal corners are rounded to aid HeatingThe fully jacketed chamber is pre-heated by steam in sterilizer jacket and door are completely insulated with a 1"chloride free glass wool thereby keeping the AUTOCLAVE cool onthe outside.

8 Customized thicker insulation is optional. Theinsulation is completely enclosed within a stainless steel insulation reduces the energy consumption by keeping theheat inside the jacket and FEATURESDoor Safety Systems: A pneumatic safety component (pressure switch) isinstalled in the AUTOCLAVE which prevents opening of thedoors until pressure in thechamber reaches roompressure. Door chamber cannot be opened when chamber ispressurized Steam is not allowed into the chamber when the door isopen A cycle cannot start if the door is open or not properlylocked The sliding door movement will stop immediately if anobject is detected in front of the closing door Double door safety is implemented through interlocks whichprevent both doors from being opened simultaneouslySafety Valves:Both chamber and jacket are equipped withpressure relief safety valves.

9 Ifthe pressure exceeds theallowed limit then the safety valves will Steam Generator Safety:A water level controlsystem maintains a constant water level that ensures safeoperation of the heaters. Furthermore, a sensor ensures thatthe heater does not operate without a minimum water Shut-Off:Easily accessible emergency switchesfor immediate cycle Independent Monitoring:The digital monitoringsystem provides a cross-reference and guarantees accurateresults. The operator has two means of monitoringtemperature and RING VACUUM PUMPThe vacuum pump effectively removes air from the pump is mounted on vibration absorbers to vacuum action the flow path of the media is through thecondenser (heat exchanger) and the vacuum pump/ejector todrain.

10 This protects the vacuum system from & COMPONENTSThe piping system of the AUTOCLAVE consists of air-operated ballvalves, which control the condensate and steam flow in and outof the chamber, operates the vacuum, and the air inlet configuration:Piping and fittings:304 threaded stainless steelComponents*:BrassOptional configuration (stainless steel):Primary piping & fittings**:316 + triclamp fittingsPrimary components:316 triclamps or 316 threaded* Components Manual valve, non-return valve,pressure regulator, pneumatic valves, steam trap, etc.** Primary piping system that connects to the chamberup to and including the first GAUGESThe single door configuration has four pressure gaugesthat indicate the pressure in the chamber, jacket, doorgasket, and the steam pressure gauges are located on the AUTOCLAVE 's frontpanel (loading side).