Transcription of BD BodyGuard T Syringe Pump
1 BD BodyGuard T Syringe Pump . Evolution from CME T34 Ambulatory Syringe driver The intent of this document is to explain evolutions from CME T34 , CME T34 3rd Edition to the New BD BodyGuard T Default Syringe Brands Configured for Use BD BodyGuard T key changes versus CME T34TM include: Syringe Sizes (ml) T34 2nd Edition Syringe Sizes (ml) T34 3rd Edition Manufacturer and BD BodyGuard T. Pump name to BD BodyGuard T Removal of the patient guide BD Plastipak - 3 5 10 20 30 50 - 3 5 10 20 30 50. BD Branded labeling Battery life enhancement Reference number: 999-103 BDEN for English Extra level of security for Bolus enabling function Braun Omnifix 2 - 5 10 20 30 50 2 - 5 10 20 30 50.
2 999-103 BDSE for Swedish Near End of infusion alarm activated 15 min Monoject - - 6 12 20 35 50 - 3 6 12 20 35 - New packaging that include pump, DFU, QRG prior the end of infusion Removal of KVO function Codan/Once - 5 10 20 30 50 - - - 10 20 30 50. Alarms Terumo - - 5 10 20 30 50 - - 5 10 20 30 50. BD BodyGuard T Syringe pump - alarm tones and priorities Nipro - - - - - - - - - 5 10 20 30 50. Description Alarm Type Audio Signal as per 60601-1-8 Visual Signal as per 60601-1-8. Down occlusion It is possible to disable default syringes from memory. This procedure is not detailed in the DFU as it should only End battery be undertaken by trained, certified service centres or biomedical engineers.
3 Please consult a biomed engineer or End of infusion your local T34 /BD BodyGuard T Syringe pump supplier should this need arise. High Priority Alarm High Priority RED flashing visual Syringe displaced during infusion Requires immediate user response Volume min. 45 dBA approx Operation LED flashes RED. Category 3rd Edition versus 2nd edition BD BodyGuard T versus CME T34 . Restart pump Switch off and on Battery life Battery life update: Battery life update: ERROR XX 25h @ 1ml/h 50h @ 1ml/h Pump paused too long 20h @ 5ml/h 35h @ 5ml/h Low Priority Low Priority Alarm YELLOW solid visual Always check battery power prior to each Always check battery power prior to each Low battery 2 or 3 pulses Requires user awareness Operation LED solid YELLOW.
4 Near end Volume min. 45 dBA approx infusion infusion Bolus started/completed Time to alarm before end of infusion Fixed to 3 min before end of infusion Fixed to 15 min before end of infusion Keypad lock/unlock Low battery alarm before end of Activated hours before end of battery alarm Enhanced to at least 30 min before end of Syringe plunger at the battery (depending on battery brand) battery alarm (depending on battery brand). limit of travel Information Signal Epidural indication - intended use Intended use removed Intended use removed Syringe loaded Provides information that may Purge started/end 1 or 2 pulses No visual Immunoglobulins, biosimilar - Intended use updated to include Intended use updated to include or may not require actions from Power on/off clinicians intended use infusion started/resumed/ Minimum flow rate Range updated to Range updated to stopped by user Flow rate increments Range updated to: Range updated to.
5 Service interval alert 10 ml/h in ml/h increments 10 ml/h in ml/h increments Clinical updates 9V battery power failure Backup alarm Buzzer No visual 10 ml/h in ml/h increments 10 ml/h in ml/h increments RED flashing visual 30 ml/h in ml/h increments 30 ml/h in ml/h increments Alarm output volum failure Backup alarm Buzzer 50 299 ml/h in 1 ml/h increments 50 299 ml/h in 1 ml/h increments Operation LED flashes RED. 300 650 ml/h in 5 ml/h increments. 300 650 ml/h in 5 ml/h increments. Main visual changes Bolus function Updated to include new function Adding 1 level of security by adding a enable/disable option in technician menu T34 2nd Edition T34 3rd Edition Minimum occlusion pressure Updated from 100mmHg to 200mmHg 100mmHg to 200mmHg Front label has pump name Fluid ingress - classification IP22 IP22.
6 Weight Updated from 210g to 230g 230g New connectors - battery Update to rear assembly Update to rear assembly Cover labels and branding Update to CME labels Update to BD BodyGuard T labels Key pad has text May have battery sticker on rear case Key pad has Universal Symbols Fluid drain hole Rear label Updated to comply with standards Updated to comply with standards KVO Rate Operation Update to range from ml/h KVO option removed BD BodyGuard T to ml/h Rear drainage New design feature to avoid fluid New design feature to avoid fluid accumulation accumulation GTIN barcode labels Updated to comply to GTIN standard Updated to comply to GTIN standard Packaging New box contains pump, direction for use Key pad has Universal Symbols Fluid drain hole and quick reference guide Continues overleaf Category 3rd Edition versus 2nd edition BD BodyGuard T versus CME T34.
7 Software Update to software Update to software Add primary sound self-test into Update to software Update to software boot-up process Update BD BodyComm software for BD BodyComm & later BD BodyComm & later 3rd edition pump software Service interval alert added New default at 12 months - New default at 12 months - (1-year period) can be switched off in Tech. mode can be switched off in Tech. mode Reduce safe stop current threshold New limits New limits and use this threshold during travel calibration ( Pinch Hazard ). Date input order changed Updated to date: year, month, day Updated to date: year, month, day RTC battery changed Updated from disposable battery to Updated from disposable battery to rechargeable 3V battery rechargeable 3V battery Technical updates RTC battery failure Need to update time and date under Can update date and time when turning Change Set Up', then to reset service on the pump due date under Tech mode.
8 Add microphone test and threshold New alarm test during startup New alarm test during startup Add backup alarm test into Tech. New test function in 3rd edition New test function in 3rd edition mode Add LEDs test into Tech. mode New test function in 3rd edition New test function in 3rd edition Add microphone level calibration into New feature in 3rd edition New feature in 3rd edition Tech. mode Keypad LED yellow colour LED colour update from red and green to red, LED colour update from red and green to red, yellow and green yellow and green Part differences - hardware Update included in TSM Update included in TSM.
9 Battery indication (vs voltage) was Changes to battery measurement method Battery gauge accuracy enhancement changed in 3rd Edition Pressing on/off or stop key disabled Update 3rd edition pump alarms if 9V battery Update 3rd edition pump alarms if 9V battery the alarm is removed is removed Bolus option In Clinican menu Adding 1 level of security by adding a enable/disable option in technician menu Pump master N/A Allows software update via the pump serial port BD, the BD Logo, BodyComm, BodyGuard , T34 and Plastipak are trademarks of Becton, Dickinson and Company or its affiliates.
10 All other trademarks are the property of their respective owners. 2020 BD. All rights reserved. 0000CF05107 Issue 1.