1 College of American Pathologists (CAP) gh2 survey Data: (updated 5/12) The American Diabetes Association (ADA) recommends that laboratories use only HbA1c assay methods that have been NGSP certified and report results as %HbA1c . The ADA also recommends that all laboratories performing HbA1c testing participate in the College of American Pathologists (CAP) fresh sample proficiency testing survey (see ADA Recommendations section on this website for more details). CAP GH2 data for the first survey of 2012 are summarized below. The NGSP target or reference values are based on replicate analyses using seven NGSP certified secondary reference methods. 2012 GH2-A (fresh pooled samples) GH2-01 GH2-02 GH2-03 NGSP Reference Value (%HbA1c)t

1 College of American Pathologists (CAP) GH2 Survey Data: (updated 5/12) The American Diabetes Association (ADA) recommends that laboratories use only HbA1c assay methods that have been

Tags:

  American, Survey, College, Pathologist, College of american pathologists, Gh2 survey

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of College of American Pathologists (CAP) GH2 …

1 1 College of American Pathologists (CAP) gh2 survey Data: (updated 5/12) The American Diabetes Association (ADA) recommends that laboratories use only HbA1c assay methods that have been NGSP certified and report results as %HbA1c . The ADA also recommends that all laboratories performing HbA1c testing participate in the College of American Pathologists (CAP) fresh sample proficiency testing survey (see ADA Recommendations section on this website for more details). CAP GH2 data for the first survey of 2012 are summarized below. The NGSP target or reference values are based on replicate analyses using seven NGSP certified secondary reference methods. 2012 GH2-A (fresh pooled samples) GH2-01 GH2-02 GH2-03 NGSP Reference Value (%HbA1c)t no.

2 Labs Mean %HbA1c Mean bias % CVMean %HbA1c Mean bias % CV Mean %HbA1c Mean bias % CV * Abbott Architect c 74 * Axis-Shield Afinion 28 * Bayer (Metrika) A1cNOW# 28 * Beckman AU systems 35 * Beckman Synchron LX Systems 21 * Beckman UniCel DxC Synchron 300 * Bio-Rad D-10 244 * Bio-Rad in2it 10 * Bio-Rad Variant II 111 * Bio-Rad Variant II Turbo 183 * Bio-Rad Variant II Turbo 47 * Roche Cobas c311 12 * Roche Cobas c500/700 248 * Roche Cobas Integra 400 53 * Roche Cobas Integra 800 144 * Roche/Hitachi Modular P 14 * Siemens Advia New Reagent 41 * Siemens Advia Original Reagent

3 20 * Siemens DCA 2000/2000+ 71 * Siemens DCA Vantage 253 * Siemens Dimension ExL new reagent 62 * Siemens Dimension ExL orig reagent 29 * Siemens Dimension RxL new reagent 114 * Siemens Dimension RxL orig reagent 76 * Siemens Dimension Vista new reagent 181 * Siemens Dimension Xpand new reagent 65 * Siemens Dimension Xpand orig reagent 37 * Tosoh G7 Auto HPLC 184 * Tosoh G8 Auto HPLC 260 * Trinity Biotech HPLC (Affinity) 31 * (Ortho Clin Diag) Vitros 5,1 FS Chem System 210 * = NGSP certified at the time of the survey t Assigned as the mean of 3 replicate analyses per day for two days per method using 7 NGSP certified secondary reference methods.

4 #EDTA in the CAP sample has been shown by the manufacturer of A1 CNow+ to cause artificially low results by this method. Routine samples for this method are from fingerstick and do not include EDTA. The manufacturer recommends the use of heparin anticoagulant instead of EDTA when testing venous samples Gray shading indicates bias > HbA1c or CV > 5% (except Bayer A1cNow bias) Commentary by R. Little, , NGSP Network Coordinator for the NGSP Steering Committee In 2012, based on data from the GH2-A survey : Bias from the NGSP target and variability ( 2SD) are shown in the table above and in figure 1 for each method. The shaded rectangle (fig 1) reflects the current CAP acceptance limit of 7%. In addition to the Bayer A1cNow# (see footnote above), the method-specific biases were over for 8 2methods for one or more levels: Axis-Shield Afinion, Bio-Rad D-10, Bio-Rad in2it, Bio-Rad VII Turbo , Siemens Advia original and new reagent, and Tosoh G7 and G8.

5 Method-specific, between-laboratory CV s ranged from to All but 4 methods (in2it, Advia new and original, Cobas c311) had CVs below 5% for all three levels. Approximately 97% of laboratories were using methods that had between-lab CVs < at all three HbA1c levels; only about 20% of laboratories are using methods with CVs <3% at all three HbA1c levels. The current pass limit for the gh2 survey is 7%. The overall pass rate for this survey was , and of labs passing for the low, mid and high samples, respectively. For individual methods, the lowest pass rate was and the highest was 100% (Sacks, Chemistry Resource Committee, CAP GH2-A 2012). Methods with small bias and low CVs will have the highest pass rates and, conversely, methods with large bias and/or high CVs will have the lowest pass rates.

6 The overall CVs for the last five surveys are shown in Table 1. This 2012A survey s CVs were still above at two levels; our goal is at or below (Clin Chem 57:793-8, 2011). There continues to be a few methods with either high CVs or high bias or both. But there are also many methods that show consistent good performance. NOTE: The NGSP certification evaluates agreement of each method at the manufacturing site using one lot of reagents and calibrators, one instrument, and one application under optimal conditions. CAP precision reflects between-laboratory reproducibility, often with more than one lot of reagents and calibrators, and sometimes with different instruments ( Cobas Integra 400 & Cobas Integra 800) and/or different applications ( Cobas Integra hemolysate or whole blood application).

7 In addition, if changes were made in the method just prior to NGSP certification, it is possible that not all participating laboratories in the field would have made the change at the time of the CAP survey . For these reasons, it is important that laboratories review not only the certification status of HbA1c methods but also their performance in the CAP survey over time (a good indication of field performance) when selecting or evaluating HbA1c assay methods. 3 Figure 1: Bias and Variability from the NGSP Target GH2-A 2012 low level (mean 2SD)NGSP Target+/- 7%# EDTA interference# GH2-A 2012 mid level (mean 2SD)NGSP Target 7%# EDTA interference# . 011 . GH2-A 2012 Hi level (mean 2SD)NGSP Target 7%## EDTA interference 4 Table 1.

8 Overall Variability for 2010-2012 for all GH2 participants Mailing Sample# # of labs TargetAll method mean A-2010 01 2573 02 2566 03 2581 B-2010 04 2693 05 2691 06 2685 A-2011 01 2652 02 2645 03 2649 B-2011 04 2877 05 2872 06 2871 A 2012 01 3298 02 3316 03 3301