Transcription of DAQSTATION CX2000 User's Manual - Insatech - …
1 user sManualYokogawa Electric CorporationModel CX2000 /CX2010/CX2020/CX2210/CX2200/CX241 0/CX2420/CX2610/CX2620 DAQSTATION CX2000IM 04L31A01-01E7th EditioniIM 04L31A01-01E7th Edition : May 2007 (YK)All Rights Reserved, Copyright 2001 Yokogawa Electric CorporationThank you for purchasing the CX2000 . This Manual describes the functions (excludingthe communications functions), installation and wiring procedures, operating procedures,and handling precautions of the CX2000 . To ensure correct use, please read thismanual thoroughly before beginning operation. The following six manuals are alsoprovided in addition to this Manual . Read them along with this Manuals Provided on the Accompanying CD-ROMM anual TitleManual CX1000/IM 04L31A01-17 EDescribes the communications functions of theCX2000 CommunicationsCX1000/ CX2000 using the Ethernet/serialInterface user s for CXIM 04L31A01-61 EDescribes the functions and operatingUser s Manualprocedure of the software DAQSTANDARDfor CX that comes with the ManualsManual TitleManual Opration GuideIM 04L31A01-02EA guide providing simple explanations ofcontrol-related operations for the CX2000 (includes a chart of setting values).
2 CX2000 Installation andIM 04L31A01-71 EDescribes concisely the installationConnection Guideprocedures and wiring procedures of on the Use ofIM 04L31A01-72 ECautions regarding the use of the CX1000/ CX2000 The same information is written on pages ii andiii of this user s Control of IM 04L31A01-91 CGives a description of pollution Caused bythe ProductNotes This Manual describes the CX2000 , style number S3. The contents of this Manual are subject to change without prior notice as a result ofcontinuing improvements to the instrument s performance and functions. Every effort has been made in the preparation of this Manual to ensure the accuracy ofits contents. However, should you have any questions or find any errors, please contactyour nearest YOKOGAWA dealer. Copying or reproducing all or any part of the contents of this Manual without thepermission of Yokogawa Electric Corporation is strictly prohibited.
3 The TCP/IP software of this product and the document concerning the TCP/IPsoftware have been developed/created by YOKOGAWA based on the BSDN etworking Software, Release 1 that has been licensed from the Regents of theUniversity of vigilantplant, DAQSTATION , and DAQSTATION are registered trademarks of YokogawaElectric Corporation. Microsoft and Windows are registered trademarks or trademarks of MicrosoftCorporation in the United States and/or other countries. Adobe and Acrobat are registered trademarks or trademarks of Adobe SystemsIncorporated. Company and product names that appear in this Manual are registered trademarks ortrademarks of their respective holders. The company and product names used in this Manual are not accompanied byregistered trademark or trademark symbols ( and ).
4 Revisions1st EditionDecember 20015th EditionJuly 20032nd EditionMarch 20026th EditionNovember 20033rd EditionApril 20027th EditionMay 20074th EditionJune 2003iiIM 04L31A01-01 ESafety PrecautionsAbout This Manual This Manual should be read by the end user . Read this Manual thoroughly and have a clear understanding of the product before operation. This Manual explains the functions of the product. YOKOGAWA does not guarantee that the product will suita particular purpose of the user . Under absolutely no circumstances may the contents of this Manual be transcribed or copied, in part or inwhole, without permission. The contents of this Manual are subject to change without prior notice. Every effort has been made in the preparation of this Manual to ensure the accuracy of its , should you have any questions or find any errors or omissions, please contact your nearestYOKOGAWA Related to the Protection, Safety, and Alteration of the Product The following safety symbols are used on the product and in this Manual .
5 Handle with care. (To avoid injury, death of personnel or damage to the instrument, the operatormust refer to the explanation in the Manual .)Functional ground terminal. (Do not use this terminal as a protective ground terminal.)Protective grounding terminalAlternating current Direct current For the protection and safe use of the product and the system controlled by it, be sure to follow theinstructions and precautions on safety that are stated in this Manual whenever you handle the special note that if you handle the product in a manner that violate these instructions, the protectionfunctionality of the product may be damaged or impaired. In such cases, YOKOGAWA does notguarantee the quality, performance, function, and safety of the product. When installing protection and/or safety circuits such as lightning protection devices and equipment forthe product and control system or designing or installing separate protection and/or safety circuits forfool-proof design and fail-safe design of the processes and lines that use the product and the controlsystem, the user should implement these using additional devices and equipment.
6 If you are replacing parts or consumable items of the product, make sure to use parts specified byYOKOGAWA. This product is not designed or manufactured to be used in critical applications that directly affect orthreaten human lives. Such applications include nuclear power equipment, devices using radioactivity,railway facilities, aviation equipment, air navigation facilities, aviation facilities, and medical equipment. Ifso used, it is the user s responsibility to include in the system additional equipment and devices thatensure personnel safety. Do not modify this product. The CX is designed for indoor 04L31A01-01 EWARNINGP ower SupplyEnsure that the source voltage matches the voltage of the power supply before turning ON GroundingMake sure to connect the protective grounding to prevent electric shock before turning ON of Protective GroundingNever cut off the internal or external protective earth wire or disconnect the wiring of the protectiveearth terminal.
7 Doing so invalidates the protective functions of the instrument and poses a potentialshock of Protective GroundingDo not operate the instrument if the protective earth or fuse might be defective. Make sure tocheck them before Not Operate in an Explosive AtmosphereDo not operate the instrument in the presence of flammable liquids or vapors. Operation in suchenvironments constitutes a safety Not Remove CoversThe cover should be removed by YOKOGAWA s qualified personnel only. Opening the cover isdangerous, because some areas inside the instrument have high ConnectionConnect the protective grounding before connecting to the item under measurement or to anexternal control to the Protective StructureOperating the CX2000 in a manner not described in this Manual may damage its from Responsibility YOKOGAWA makes no warranties regarding the product except those stated in the WARRANTY that isprovided separately.
8 YOKOGAWA assumes no liability to any party for any loss or damage, direct or indirect, caused by theuser or any unpredictable defect of the Precautions of the Software YOKOGAWA makes no warranties regarding the software accompanying this product except thosestated in the WARRANTY that is provided separately. Use the software on a single PC. You must purchase another copy of the software, if you are to use the software on another PC. Copying the software for any purposes other than backup is strictly prohibited. Please store the original media containing the software in a safe place. Reverse engineering, such as decompiling of the software, is strictly prohibited. No portion of the software supplied by YOKOGAWA may be transferred, exchanged, sublet, or leased foruse by any third party without prior permission by PrecautionsivIM 04L31A01-01 EChecking the Contents of the PackageUnpack the box and check the contents before operating the instrument.
9 If some of thecontents are not correct or missing or if there is physical damage, contact the dealerfrom which you purchased you open the operation cover on the front panel, a name plate is located on theback side of the cover. Check that the model name and suffix code given on the nameplate on the rear panel match those on the the operationcoverNOSUFFIXMODELWARNINGDo not remove the by trained personnel and SUFFIX/A6/A6R/A4F/A4FR/CST1/D5/M1/N2/P1/ TPS4/PG1/PG2/BT1 1 2 3 5 0 1 2 1 2CX2000CX2010CX2020CX2200CX2210 CX2220 CX2410CX2420 CX2610 CX2620 ExternalstoragemediumModelSuffix CodeOptional CodeDescriptionCommunicationinterfaceDis played languageOptionsNumber of internal control loops: 0, number of inputs for measurement: 10 chNumber of internal control loops: 0, number of inputs for measurement: 0 chNumber of internal control loops: 0, number of inputs for measurement: 20 chNumber of internal control loops: 2, number of inputs for measurement.
10 10 chNumber of internal control loops: 2, number of inputs for measurement: 0 chNumber of internal control loops: 2, number of inputs for measurement: 20 chNumber of internal control loops: 4, number of inputs for measurement: 10 chNumber of internal control loops: 4, number of inputs for measurement: 20 chNumber of internal control loops: 6, number of inputs for measurement: 10 chNumber of internal control loops: 6, number of inputs for measurement: 20 chFloppy disk100MB Zip diskATA flash memory card (Compact flash and adapter)250MB Zip diskEthernetEthernet + RS-232 serial interface portEthernet + RS-422A/485 serial interface portJapaneseEnglish6 measurement alarm outputs*16 measurement alarm output, 8 remote inputs*14 measurement alarm outputs, 1 FAIL output, 1 memory end output*14 measurement alarm outputs, 1 FAIL output, 1 memory end output,8 remote inputs*1 Control extension DIO (12 DIs, 12 DOs)*1*2 VGA outputComputation functionThree-wire isolated RTD (input for measurement)24-VDC/AC power supply driven24-VDC transmitter output (4 loops)*1 Batch headerProgram control (number of program patterns: 4)*3 Program control (number of program patterns: 30)*3*1 Only one can be specified at once.