Edwards XDS10, XDS5, Dry Scroll Vacuum Pump, Users Manual

A726-01-880 Issue J OriginalA72X-YY-ZZZPump TypeVariantMotor DescriptionXYYZZZ6 = XDS1001 to 99903 = Voltage set to 220 - 240 V, 50/60 Hz, single-phase4 = XDS5904 = Voltage set to 100/200 V, 50/60 Hz, single-phase905 = Voltage set to 200 - 230/380 - 460 V, 50/60 Hz, three-phase906 = Voltage set to 110 - 120 V, 50/60 Hz, single-phaseInstruction ManualXDS Dry Pump98/37/EC94/9/EC89/336/EEC73/023/EECM ay 07 Issue JXDS Dry and directive Product classification42 TECHNICAL and storage and vibration and design and the the installation: single-phase and configure the the pump to your electrical the direction of installation: three-phase and configure the the pump to your electrical the direction of and outlet test the up achieve ultimate pump condensable of ATEX materials18 CONTENTSS ectionPagedcs/7378/0507 Issue JMay 07 XDS Dry and clean the inlet and clean the gas-ballast the external fan the tip the motor pump has failed to pump has failed to achieve the required pump is pump surface temperature is above 100 pumping speed is poor256 STORAGE AND AND o

XDS Dry Pump 98/37/EC 94/9/EC 89/336/EEC 73/023/EEC. May 07 Issue J XDS Dry Pump i PAGE i ... XDS5 and XDS10 dry pumps. You must use the XDS pump as specified in this manual. Read this ... The body of the pump includes a fixed scroll and an orbiting scroll. The orbiting scroll is driven by the

Tags:

  Manual, User, Pumps, Vacuum, Scroll, Users manual, Edward, Dry pumps, Dry scroll vacuum pump, Edwards xds10, Xds10, Xds5

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Edwards XDS10, XDS5, Dry Scroll Vacuum Pump, Users Manual

1 A726-01-880 Issue J OriginalA72X-YY-ZZZPump TypeVariantMotor DescriptionXYYZZZ6 = XDS1001 to 99903 = Voltage set to 220 - 240 V, 50/60 Hz, single-phase4 = XDS5904 = Voltage set to 100/200 V, 50/60 Hz, single-phase905 = Voltage set to 200 - 230/380 - 460 V, 50/60 Hz, three-phase906 = Voltage set to 110 - 120 V, 50/60 Hz, single-phaseInstruction ManualXDS Dry Pump98/37/EC94/9/EC89/336/EEC73/023/EECM ay 07 Issue JXDS Dry and directive Product classification42 TECHNICAL and storage and vibration and design and the the installation: single-phase and configure the the pump to your electrical the direction of installation: three-phase and configure the the pump to your electrical the direction of and outlet test the up achieve ultimate pump condensable of ATEX materials18 CONTENTSS ectionPagedcs/7378/0507 Issue JMay 07 XDS Dry and clean the inlet and clean the gas-ballast the external fan the tip the motor pump has failed to pump has failed to achieve the required pump is pump surface temperature is above 100 pumping speed is poor256 STORAGE AND AND operated pipeline nozzle silencer29 INDEX 31 CONTENTS (continued)SectionPageMay 07 Issue JXDS Dry PumpiiiPAGEiii1 The single-phase XDS dry pump22 Performance characteristics63 Dimensions74 Electrical interface.

2 Single-phase pumps125 Underside of the pump146 Three-phase electrical connections: 200-230 V147 Three-phase electrical connections: 380-460 V148 Inlet strainer assembly229 Gas-ballast control assembly2310 Replacing the tip seals24 TABLES TablePage1 General characteristics52 Performance characteristics63 Dimensions (mm)74 Electrical data (single-phase pumps with item numbers -903 or -906)95 Electrical data (single-phase pumps with item numbers -904)96 Electrical data (three-phase pumps with item numbers -905)97 Wire colour codes138 Maintenance plan219 Solenoid operated pipeline valve29 ILLUSTRATIONSF igurePageIssue JMay 07 XDS Dry PumpivThis page intentionally 07 Issue JXDS Dry and definitionsThis Manual provides installation, operation andmaintenance instructions for the BOC EdwardsXDS5 and xds10 dry pumps .

3 You must use theXDS pump as specified in this Manual . Read thismanual before you install, operate and maintain theXDS pump. Important safety information is highlighted asWARNING and CAUTION instructions; you mustobey these instructions. The use of WARNINGSand CAUTIONS is defined this Manual , page, figure and tablenumbers are units used throughout this Manual conform tothe SI international system of units of are stated as absolute pressuresthroughout this Manual , unless otherwise following IEC warning labels appear on thepump:Warning - refer to - risk of electric are given where failure to observe the instruction could result in injury or death to are given where failure to observe the instruction could result in damage to the equipment, associated equipment and - hot directive implications This equipment is designed to meet therequirements of Group II Category 3equipment in accordance with Directive 94/9/EC of the European Parliament and theCouncil of 23rd March 1994 on theapproximation of the laws of the MemberStates concerning equipment and protectivesystems intended for use in potentiallyexplosive atmospheres.

4 (The ATEX Directive)The ATEX Category 3 applies in respect ofpotential ignition sources internal to theequipment. An ATEX Category has not beenassigned in respect of potential ignitionsources on the outside of the equipment asthe equipment has not been designed for usewhere there is an external potentiallyexplosive is no potential source of ignition withinthe pump during normal operation but theremay be potential sources of ignition underconditions of predicted and rare malfunctionas defined in the Directive. Accordingly,although the pump is designed to pumpflammable materials and mixtures, operatingprocedures should ensure that under allnormal and reasonably predicted conditions,these materials and mixtures are not withinexplosive limits.

5 Category 3 is consideredappropriate for the avoidance of ignition in thecase of a rare malfunction which allowsflammable materials or mixtures to passthrough the pump while within their JMay 07 XDS Dry PumpINTRODUCTION2 PAGE2 Figure 1 - The single-phase XDS dry pump1. NW25 inlet port2. Gas-ballast control3. Cooling fan4. DN25NW/3/4 BSP exhaust5. Baseplate6. Voltage indicator7. On / off switch8. Hour counterMay 07 Issue JXDS Dry PumpINTRODUCTION3 PAGE3 When flammable materials are present withinthe equipment you must: Not allow air to enter the equipment. Ensure the system is leak tightFor further information, please refer to the backpage of this instruction Manual for the contact de-tails of your nearest BOC XDS pump is shown in Figure 1.

6 Refer toFigure 1 for item numbers in brackets in thefollowing XDS pump is a compact, reliable Vacuum pumpwhich is suitable for use on vapour handlingprocesses. The XDS pump may be used for somepumping applications involving corrosive substancesand particulates; for information on pumpingflammable gases, please refer to Section andcontact BOC Edwards for any further body of the pump includes a fixed Scroll and anorbiting Scroll . The orbiting Scroll is driven by theelectric motor through an eccentric cam on themotor drive shaft. The movement of the orbitingscroll, meshed with the fixed Scroll , formssuccessive crescent shaped volumes in the which enters the pump through the inlet iscompressed by the movement of the orbiting scrolland swept towards the centre of the fixed compressed gas enters the exhaust port nearthe centre of the stationary Scroll and is exhaustedfrom the pump through the XDS pump is a dry Vacuum pump, as all thebearings, with their hydrocarbon lubricant, areisolated from the Vacuum inlet of the pump is fitted with a removable inletstrainer which prevents the entry of debris into pump mechanism is driven directly by a single-phase or three-phase electric motor.

7 Single-phasemotors are fitted with an on/off switch (7) and athermal overload device. When the motor is toohot, the thermal overload device switches off thepump. The thermal overload device has anautomatic reset; when the motor cools down, thedevice resets and (unless you have incorporatedsuitable control equipment which must be manuallyreset), the motor will restart. The pump is air-cooled by a fan mounted at theopposite end to the pump is mounted on a baseplate on rubber of suitable vibration isolators and otheraccessories are provided in Section controlTo pump high vapour loads, gas-ballast is deliveredinto the pump to prevent condensation of thevapour carried by the pumped can be introduced to the low Vacuum stagesthrough the gas-ballast valve.

8 Alternatively, an inertgas such as nitrogen can be supplied through asuitable external gas-ballast control has three-positions: Closed (position '0') Low flow (position 'I') High flow (position 'II') pump scrolls are made from anodisedaluminium and the motor housing made fromaluminium only. All surfaces of the pump which areexposed to the pumped gases are free from copper,zinc and JMay 07 XDS Dry Product classificationType of protection against Class 1electric shockSupply connection110-120 V, 50/60 Hz or 220-240 V, 50/60 HzMode of operation:ContinuousProtection against Not protectedhazards of explosionProtection against Not protectedingress of liquids(IPXO)Installation categoryIIPollution degree2 Degree of mobilityPortable(benchmounted)May 07 Issue JXDS Dry PumpTECHNICAL DATA5 PAGE52 TECHNICAL and storage conditionsAmbient temperature.

9 Operating range10 C to 40 CStorage range-30 C to 70 CSurface temperature of pump above ambient (whenoperating at ultimate) typically 45 C Maximum humidity (operation)90%Altitudeup to 2000 characteristics of the pumps are defined inTable 1 below. Characteristics shown aredisplacement, pumping speed and maximumpermitted interface on pump switch off typically 30 mbar ls-1(leak tightness 1x10-6 mbar ls-1). characteristicsThe position of the gas-ballast control defines theperformance characteristics of the pump. Theseperformance characteristics are listed fully inTable curves shown in Figure 2 show the relationshipbetween inlet pressure and pumping speed for 10 XDS 5 Maximum displacement (m3/hr)50 Hz electrical supply60 Hz electrical pumping speed (m3/hr)Pneurop 6602 (1979)50 Hz electrical supply60 Hz electrical permitted inlet pressure and gas-ballast inlet pressureMaximum permitted outlet bar gauge1 bar bar gauge1 bar gaugeTable 1 - General characteristicsIssue JMay 07 XDS Dry PumpTECHNICAL DATA6 PAGE6 Figure 2 - Performance characteristicsParameterUnitsXDS 10 XDS 5 Gas-ballast control closed (position '0')typical ultimate Vacuum (total pressure) 50 Hztypical ultimate Vacuum (total pressure)

10 60 HzmbarPambarPa< 7 x 10-2< 7< 6 x 10-2< 6< 7 x 10-2< 7< 6 x 10-2< 6 Gas-ballast control closed (position '1')typical ultimate Vacuum (total pressure) 50 Hztypical ultimate Vacuum (total pressure) 60 HzGas-ballast flowwater vapour pumping capacitymbarPambarPal / ming / x x x x 10-112370 Gas-ballast control closed (position '2')typical ultimate Vacuum (total pressure) 50 Hztypical ultimate Vacuum (total pressure) 60 HzGas-ballast flowwater vapour pumping capacitywater vapour tolerancembarPambarPal / ming / x 10-1353 x 10-13015> 200355 x x 10-1353 x 10-13015> x 103 Table 2 - Performance characteristics1. 60 Hz performance2. 50 Hz performanceMay 07 Issue JXDS Dry PumpTECHNICAL DATA7 PAGE7 Figure 3 - DimensionsDimension Reference(Refer to Figure 3) xds5 XDS10A =B =C =D =E =F =G =H =J =K =L =M =N =P =551351472902881332202492104272971202302 7555135147290288133220249210427315120230 275 Table 3 - Dimensions (mm)Issue JMay 07 XDS Dry PumpTECHNICAL tilt angle10 Motor rotational speed(50 Hz supply)1440 rpmMotor rotational speed(60 Hz supply)1730 rpmMass (kg) xds5 =23 xds10 = / outlet connectionsDN25 NWOverall dimensions (mm) and vibration dataSound pressure measured atultimate Vacuum pump toISO 1120155 dB(A) at 1 mVibration: measured at theinlet port to ISO 10816-1.