Transcription of ENDOSCOPE CLEANING GUIDE - Professional Education
1 Do not hit the distal end of the endo-scope againsthard surfaces!Do not forget to attach water-resistant capsand plugs! ENDOSCOPE CLEANING GUIDEA ttach leakage tester. Turn on pump. Immerse ENDOSCOPE in clean water. Perform leakage test (angulate tip). Remove from water. Detach tester from the air source. After the tip has deflated, detach the tester from the water-resistant water into the down the insertiontube with a detergent-soaked AW channel CLEANING adapter. Flush water, then air, into elevator-wire the universal cordportion of the suctionchannel. Repeat until alldebris is the insertion tubeportion of the suctionchannel. Repeat until alldebris is in freshly pre-pared detergent all external ENDOSCOPE distal tip(and forceps-elevator, ifapplicable).
2 234511109121617 Rinse the entire instrumentin clean clean water through all channels (includingelevator-wire channel, if applicable). Flush air through all the suctioncylinder and instru-ment channel until alldebris is in detergent solutionfor the time recommendedby the detergentsolution into elevator-wire injection tube & channel plug (or all-channel irrigator). Flush detergent solution into air, water, and balloon channels. Brush forceps-elevator,if Tube & Channel Plug(or All-Channel Irrigator)5cc SyringeInjection Tube & Channel Plug(or All-Channel Irrigator)Washing TubeFor endoscopes withelevators only:1418212019 Attach suction valve. Aspirate detergent, thenair into suction and balloon balloon if the balloon by lifting up the rear band of the bal-loon after slightly Channel- CLEANING Adapter(MAJ-629)78 ABC13 Brush the balloonchannel.
3 Repeat untilall debris is Tube5ccSyringe30cc SyringeFlush detergent, water, then air into elevator- wire endoscopes withelevators only:6 Washing Tube(MH-974)5cc SyringeUse a soft, lint-free clothto remove excess moisturefrom the ENDOSCOPE and CLEANING accessories in preparation for Feedingthrough balloonchannelSuction throughballoon channelSuction122122-Step ValvesAspirate detergent,then air, through the suction and balloon all detach-able parts and necessary water-resistant caps and plugsto the endoscopeconnector and ultrasonic : This GUIDE is only a summary of the steps necessary to properly clean your ENDOSCOPE . Be sure to follow the detailed steps outlined in the INSTRUCTION MANUAL that was included with your ENDOSCOPE to companion ENDOSCOPE DISINFECTION GUIDE for disinfection information 2016 Olympus America Inc.
4 / Printed in USA OAIGI0516 PWC18269 PRECLEANING Immediately After Use LEAKAGE TESTINGMANUAL CLEANINGRINSING Wear all appropriate personal protective equipment. Meticulous CLEANING is essential for effective disinfection. Be sure to reprocess all removable parts ( , valves) according to the instruction manual. Visually inspect the equipment after CLEANING . If debris remains, repeat the procedure. IMPORTANTCAUTIONS uction CLEANING AdapterDistal tip If a leak is detected, the ENDOSCOPE must be repaired. Follow the manufacturer's GastrovideoscopesSuction ValveWhile immersed:While immersed:While immersed:While immersed.