Product Data SheetPS-001041, Rev. FFebruary 2010 micro motion ELITE high capacity coriolis meters offer unparalleled flow and density measurement performance in a large size meter. Best precision flow and density measurement in a meter that fits large line sizes Unique design delivers unparalleled measurement sensitivity and stability Guarantees consistent, reliable performance over the widest flow range Smart Meter Verification for quick, complete meter diagnosis without process interruptionSuperior performance in the most challenging applications Industry standard for custody transfer and critical process control Best two-phase flow capability for batching, loading, and entrained air applications Immune to fluid, process, or environmental effects for superb measurement confidenceHigh performance compact drainable coriolis meterPeak performance high capacity meterHygienic compact drainable coriolis meterStraight tube full-bore coriolis meterGeneral purpose flow-only coriolis meterF-SeriesELITE HCH-SeriesT-SeriesR-SeriesExtreme low-flow coriolis meterLF-SeriesMicro motion ELITE high capacity coriolis Flow and Density MetersPeak performance coriolis meterELITE 2 ELITE high capacity Flow and Density MetersMicro motion ELITE high capacity flow and density metersMicro motion coriolis meters from Emerson Process Management meet a vast range of application needs, ranging from extreme low-flow up to high -flow, high - capacity lines.

Product Data Sheet PS-001041, Rev. F February 2010 Micro Motion® ELITE® High Capacity Coriolis meters offer unparalleled flow and density measurement performance in a large size

Tags:

  High, Engineering, Micro, Capacity, Motion, Purdue, Coriolis, Micro motion, High capacity coriolis

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of engineering.purdue.edu

1 Product Data SheetPS-001041, Rev. FFebruary 2010 micro motion ELITE high capacity coriolis meters offer unparalleled flow and density measurement performance in a large size meter. Best precision flow and density measurement in a meter that fits large line sizes Unique design delivers unparalleled measurement sensitivity and stability Guarantees consistent, reliable performance over the widest flow range Smart Meter Verification for quick, complete meter diagnosis without process interruptionSuperior performance in the most challenging applications Industry standard for custody transfer and critical process control Best two-phase flow capability for batching, loading, and entrained air applications Immune to fluid, process, or environmental effects for superb measurement confidenceHigh performance compact drainable coriolis meterPeak performance high capacity meterHygienic compact drainable coriolis meterStraight tube full-bore coriolis meterGeneral purpose flow-only coriolis meterF-SeriesELITE HCH-SeriesT-SeriesR-SeriesExtreme low-flow coriolis meterLF-SeriesMicro motion ELITE high capacity coriolis Flow and Density MetersPeak performance coriolis meterELITE 2 ELITE high capacity Flow and Density MetersMicro motion ELITE high capacity flow and density metersMicro motion coriolis meters from Emerson Process Management meet a vast range of application needs, ranging from extreme low-flow up to high -flow, high - capacity lines.

2 Cryogenic, hygienic, high -temperature, and high -pressure micro motion meters can handle them all. micro motion meters are available with a variety of wetted parts to ensure the best material meters. coriolis meters offer dramatic benefits over traditional volumetric measurement technologies. coriolis meters: Deliver accurate and repeatable process data over a wide range of flow rates and process conditions. Provide direct inline measurement of mass flow and density, and also measure volume flow and temperature all from a single device. Have no moving parts, so maintenance costs are minimal. Have no requirements for flow conditioning or straight pipe runs, so installation is simplified and less expensive. Provide advanced diagnostic tools for both the meter and the high capacity coriolis Meters. micro motion ELITE high capacity Meters are the leading meters for precision flow and density measurement. ELITE meters offer the most accurate measurement available for virtually any process fluid, while exhibiting exceptionally low pressure with Smart Meter Verification, ELITE delivers the best in measurement and ease of use for critical applications.

3 ELITE meters offer the best measurement performance for mass, density, and volume, regardless of process or environmental conditions. ELITE meters provide measurement capability for two-phase flow, liquid, and gas custody transfer. ContentsLiquid flow performance .. 3 Gas flow performance .. 4 Density performance (liquid only) .. 4 Environmental effects .. 5 Temperature specifications .. 6 Power consumption .. 7 Pressure ratings .. 7 Hazardous area classifications .. 8 Materials of construction .. 9 Weight and center of gravity .. 10 Dimensions .. 11 Fitting options .. 14 Ordering information .. 16 ELITE high capacity Flow and Density Meters3 Liquid flow performance MassVo l u m e(1)(1) Specifications for volumetric flow rate are based on a process-fluid density of g/cm3 ( kg/m3). For fluids with density other than g/cm3 ( kg/m3), the volumetric flow rate equals the mass flow rate divided by the fluid s flow rateCMFHC254,00014701,470,00064711,470,0 0012,5201470 CMFHC394,00025502,550,00011,2272,550,000 21,7302550 CMFHC4120,00032663,265,87014,3793,265,87 027,8303266 Mass flow accuracy(2)(2) Stated flow accuracy includes the combined effects of repeatability, linearity, and hysteresis.

4 All specifications for liquids are based on reference conditions of water at 68 to 77 F (20 to 25 C) and 15 to 30 psig (1 to 2 bar), unless otherwise noted. of rate(3)(3) When flow rate is less than zero stability / , accuracy = [(zero stability / flow rate) 100]% of rate, and repeatability = [ (zero stability / flow rate) 100]%.Volume flow accuracy(2) of rate(3)Repeatability of rate(3)lb/minkg/hZero high capacity Flow and Density MetersGas flow performance Density performance (liquid only) When selecting sensors for gas applications, measurement accuracy is a function of fluid mass flow rate independent of operating temperature, pressure, or composition. However, pressure drop through the sensor is dependent upon operating temperature, pressure, and fluid composition. Therefore, when selecting a sensor for any particular gas application, it is highly recommended that each sensor be sized using micro motion s product selector, available at l u m e (1)(1) Standard (SCFM) reference conditions are psia and 68 F.

5 Normal (Nm3/h) reference conditions are bar-a and 0 flow rateslb/minkg/hSCFMNm3/hAir (2)(2) Typical flow rates that produce approximately 5 psid ( bar) pressure drop on air at 68 F (20 C) and 100 psi ( bar).CMFHC2121533,06716,15527,462 CMFHC3190051,81825,58743,474 CMFHC4300081,81740,40068,643 Natural gas (3)(3) Typical flow rates that produce approximately 30 psid ( bar) pressure drop on natural gas (MW ) at 68 F (20 C) and 500 psi (34 bar).CMFHC24450121,109102,797174,744 CMFHC38500231,332196,353333,780 CMFHC410,000272,155231,000392,682 Mass flow accuracy of rate (4)(4) When flow rate is less than zero stability / , accuracy equals [(zero stability / flow rate) 100]% of rate and repeatability equals [ (zero stability / flow rate) 100]% of of rate(4)g/cm3 kg/m3 Accuracy (1)(1) Accuracy includes the combined effects of repeatability, linearity, and hysteresis. Specifications are based on reference conditions of water at 68 to 77 F (20 to 25 C) and 15 to 30 psig (1 to 2 bar), unless otherwise noted.

6 To 5up to 5000 ELITE high capacity Flow and Density Meters5 Environmental effects Process temperature effectProcess temperature effect is defined as: For mass flow measurement, the worst-case zero offset due to process fluid temperature change away from the zeroing temperature. For density measurement, the maximum measurement offset due to process fluid temperature change away from the density calibration temperature effect% of maximum flow rate per Cdensity accuracy per C (1)(1) For 100 C and kg/m3 CMFHC2 effectPressure effect is defined as the change in sensor flow and density sensitivity due to process pressure change away from the calibration pressure. Pressure effect can be effect on mass flow accuracy% of rate per psi% of rate per barliquidgasliquidgasCMFHC2 effect on density accuracyg/cm3 per psikg/m3 per barCMFHC2 high capacity Flow and Density MetersTemperature specifications Accuracy 1 C of reading in CRepeatability CTemperature limits (1) (2) (3)(1) When ambient temperature is below 40 F ( 40 C), an enhanced core processor or Model 2400S transmitter must be heated to bring its local ambient temperature to between 40 F ( 40 C) and +140 F (+60 C).

7 Long-term storage of electronics at ambient temperatures below 40 F ( 40 C) is not recommended.(2) Temperature limits may be further restricted by hazardous area approvals. See page 8.(3) The extended mount option allows the sensor case to be insulated without covering the transmitter, enhanced core processor, or junction box, but does not affect temperature modelsHigh-temperature models Ambient temperature of enhanced core processor or transmitter in F ( C)Maximum process temperature in F ( C) 40 ( 40) 4 ( 20)32 (0)68 (20)104 (40)140 (60)176 (80) 400 ( 240) 292 ( 180) 184 ( 120) 76 ( 60)32 (0)140 (60)356 (180)464 (240)248 (120)113 (45)400 (204)140 (60) 40 ( 40) 4 ( 20)32 (0)68 (20)104 (40)140 (60)176 (80)662 (350) 58 ( 50)32 (0)464 (240)320 (160)176 (80)752 (400)608 (320)Maximum process temperature in F ( C)Ambient temperature of enhanced core processor or transmitter in F ( C)ELITE high capacity Flow and Density Meters7 Pressure ratings Power consumption psibarSensor rating (1) (2)(1) Process connection rating may differ from sensor rating.

8 The meter will ship from micro motion rated to the pressure rating of the sensor or the pressure rating of the process connection, whichever is lower. Please choose process connections accordingly.(2) For operating temperatures above 200 F (93 C), pressure needs to be derated as follows. Linear interpolation may be used between values. Process connection derating may differ from sensor rating. CMFHC2M1480102 CMFHC3M1480102 CMFHC4M1480102 CMFHC2A1480102 CMFHC3A1480102 PED complianceSensors comply with council directive 97/23/EC of 29 May 1997 on Pressure EquipmentMeter with enhanced core processor4 watts maximumMeter with Model 2400S transmitter7 watts maximumMeter with Model 1700/2700 transmitter11 watts maximum316L sensorsup to 200 F (up to 93 C)Noneat 300 F (at 148 C)Noneat 400 F (at 204 C) deratingat 500 F (at 260 C) deratingat 600 F (at 316 C) deratingat 650 F (at 343 C) derating8 ELITE high capacity Flow and Density MetersHazardous area classifications CSA and CSA C-USSensor with enhanced core processorAmbient temperature: 40 F to +140 F ( 40 C to +60 C)Class I, Div 1, Groups C and DClass I, Div 2, Groups A, B, C and DSensor with Model 2400S transmitterAmbient temperature: 40 F to +140 F ( 40 C to +60 C)Class I, Div.

9 2, Groups, A, B, C, and DATEX and IECExIECExSensor with enhanced core processorEx ib IIB/IIC T1 T5 Sensor with Model 2400S transmitterEx nA II T1 T5AT E XSensor with enhanced core processor0575II 2G Ex ib IIB/IIC T1 T5 (1) Gb(1) Refer to the temperature 2D Ex ib IIIC T (1) C Db IP65 Sensor with Model 2400S transmitter0575II 3G Ex nA II T1 T5 (1)II 3D Ex tD A22 IP65 T (1) CStandard models (CMFHC2M, CMFHC3M, CMFHC4M) with enhanced core processor Ambient temperature rangeTa 40 C to +60 CSensor fluid temp ( C)Max ambient temp ( C)Note 1: Use the temperature graph to determine the temperature class for a given fluid and ambient temperature. The maximum surface temperature T for dust is as follows: T5: 95 C, T4: 130 C, T3: 195 C, T2 to T1: 207 C. ELITE high capacity Flow and Density Meters9 Hazardous area classifications continued Materials of construction ATEX and IECExHigh-temperature models (CMFHC2A and CMFHC3A) with enhanced core processor Ambient temperature rangeTa 50 C to +55 CWetted parts (1)(1) General corrosion guides do not account for cyclical stress, and therefore should not be relied upon when choosing a wetted material for your micro motion sensor.

10 Please refer to the micro motion corrosion guide for proper material compatibility stainless steel Housing304L or 316L stainless steelEnhanced core processorPolyurethane-painted aluminum or 300-series stainless steel (2); NEMA 4X (IP66)(2) 316L stainless steel is 2400S transmitterPolyurethane-painted aluminum or 300-series stainless steel (2); NEMA 4X (IP66)Sensor fluid temp ( C)Max ambient temp ( C)Note 1: Use the temperature graph to determine the temperature class for a given fluid and ambient temperature. The maximum surface temperature T for dust is as follows: T5: 95 C, T4: 130 C, T3: 195 C, T2: 290 C, T1: 363 C. The minimum ambient and process fluid temperature allowed for dust is 40 high capacity Flow and Density MetersWeight and center of gravityWeight includes the weight of the sensor with CL150 weld neck raised face flanges, plus the weight of the enhanced core processor or Model 2400S of gravity is based on a sensor with integral enhanced core processor or Model 2400S transmitter, with meter empty of fluid.