feasa LED ANALYSERThe Innovative Solution for Testing LEDsThe feasa ANALYSER is an innovative solution for testing multiple LEDs simultaneously for Color and Brightness. There are two Models feasa I(ICT) and feasa F(Functional). These can beordered in 3, 5, 10 and 20 Channel choosing which Model is most suitable for your application there are a number of issuesto consider. In this regard the choice of Interface is offers a simple interface to the LED ANALYSER with no requirement for an additionalpower supply. Baud rates up to 460800 baud are RS232 Serial Port is easy to use with a max baud rate of 115200.

FEASA™ LED ANALYSER The Innovative Solution for Testing LEDs The Feasa Analyser is an innovative solution for testing multiple LEDs simultaneously for Color

Tags:

  Feasa led analyser, Feasa, Analyser, Feasa analyser

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of FEASA™ LED ANALYSER

1 feasa LED ANALYSERThe Innovative Solution for Testing LEDsThe feasa ANALYSER is an innovative solution for testing multiple LEDs simultaneously for Color and Brightness. There are two Models feasa I(ICT) and feasa F(Functional). These can beordered in 3, 5, 10 and 20 Channel choosing which Model is most suitable for your application there are a number of issuesto consider. In this regard the choice of Interface is offers a simple interface to the LED ANALYSER with no requirement for an additionalpower supply. Baud rates up to 460800 baud are RS232 Serial Port is easy to use with a max baud rate of 115200.

2 It requires the use of an external power 20pin ICT Port can be used in either Frequency Out or Synchronous Serial OutThe Frequency Out protocol can be used where access to an RS232 Serial Port is not available. Three frequencies are used to represent the Color and Intensity of the Serial PortThe Synchronous Serial protocol is suitable when tester resources are limited or no other options are ChainMultiple LED Analysers can be connected together using the Daisy Chain Connectors. Onlyone RS232 Serial Port or USB Port is required to connect up to 30 LED Trigger InputThe feasa I provides an External Trigger Input which can be used to synchronise LED measurements with an external event such as an LED switching Enterprises Limerick Ireland Telephone: + 353 61 330333 - Fax : + 353 61 330452 - Website: Office: feasa Enterprises Limited, Holland Road, National Technology Park, Castletroy, , in Ireland, No.

3 106933. Copyright 2011 feasa Enterprises Limited. All rights 1 of 4 Rev 1st December, 2015 feasa IFeasa FUSBRS23220 Pin Port Frequency Out20 Pin Port Synchronous SerialDaisy ChainExternal Trigger InputNOYESYESYESNOYESYESYESNONOYESNOFEAS A LED ANALYSERThe Innovative Solution for Testing LEDsFeasa Enterprises Limerick Ireland Telephone: + 353 61 330333 - Fax : + 353 61 330452 - Website: Office: feasa Enterprises Limited, Holland Road, National Technology Park, Castletroy, , in Ireland, No. 106933. Copyright 2011 feasa Enterprises Limited. All rights 2 of 4 Rev 1st December, 2015 TEST TIMEThe speed of the test is dependent on the intensity of the LEDs being tested, Bright LEDshave a shorter Test Time, Dimmer LEDs have a longer Test capture (measurement) of up to 20 LEDs is done in parallel and can be achieved in timesas fast as 102ms depending on the Intensity (Brightness).

4 The data is read back from each fiber sequentially and takes approximately 5ms per fiber, forexample:Ultra High Brigh LEDs1 LED -Capture Time is 2ms and Read Back is 5ms, Total 7ms20 LEDs - Capture Time is 2ms and Read Back is 100ms, Total 102msDim LEDs 1 LED -Capture Time is 650ms and Read Back is 5ms, Total 655ms20 LEDs -Capture Time is 650ms and Read Back is 100ms, Total 750msUSB / RS232 SERIAL PORT TEST CAPTURE TIMESThe Read Back Time per fiber is always approximately ICT the Capture Times are the same as USB/RS232 Serial Port. However, the Read Back Times are dependent on the frequencies being measured.

5 Using an Agilent i3070 theRead Back Times are 400ms to 700ms approximately. Range Capture Time C (Auto Capture) 350msC1 (Low Intensity) 650msC2 (Medium Intensity) 200msC3 (High Intensity) 22msC4 (Super High Intensity) 4msC5 (Ultra High Intensity) 2ms feasa LED ANALYSERThe Innovative Solution for Testing LEDsFeasa Enterprises Limerick Ireland Telephone: + 353 61 330333 - Fax : + 353 61 330452 - Website: Office.

6 feasa Enterprises Limited, Holland Road, National Technology Park, Castletroy, , in Ireland, No. 106933. Copyright 2011 feasa Enterprises Limited. All rights 3 of 4 Rev 1st December, 2015 USB / RS232 Frequency OutSynchronous Serial- Red, Green, Blue (RGB)- Hue, Saturation, Intensity (HSI)- Dominant Wavelength- CCT- CIE xy- CIE u'v'- Hue, Saturation, Intensity (HSI)- Red, Green, Blue (RGB)- Hue, Saturation, Intensity (HSI)- CCT- CIE xyDRIVERS/SOFTWAREF easa provides a comprehensive suite of Drivers and Software for ease of IFeasa FTest Models for Agilent i3070 Test Code for TeradyneDLL used for TestingProgramming examples in Labview, C++YESYESYESYESNONOYESYESIn addition, feasa also provides a number of programmes to allow for the most efficient andappropriate use of the LEDsAutomotive-RJ45 Connectors-Daytime Running Lights-Display Panels-Brake Lights-Emergency Signals-Centre High Mount Stop Lights-Traffic Lights-Side Turn Signals-Railway Signals-Emergency Stop SignalInterior Lights (Automotive & Avionics)

7 LCD Backlighting-Dashboard-TV-Map Lights-Notebook/PC-Mood Lights-Cell Phones/Smart PhonesAviation Lighting-Landing LightsOUTPUTSFEASA LED ANALYSERThe Innovative Solution for Testing LEDsFeasa Enterprises Limerick Ireland Telephone: + 353 61 330333 - Fax : + 353 61 330452 - Website: Office: feasa Enterprises Limited, Holland Road, National Technology Park, Castletroy, , in Ireland, No. 106933. Copyright 2011 feasa Enterprises Limited. All rights 4 of 4 Rev 1st December, 2015 SPECIFICATIONSORDERING INFORMATIONF easa LED AnalyserFeasa IFeasa F3 Channel5 Channel10 Channel20 ChannelPart No.

8 : feasa 3 IPart No.: feasa 5 IPart No.: feasa 10 IPart No.: feasa 20 IPart No.: feasa 3 FPart No.: feasa 5 FPart No.: feasa 10 FPart No.: feasa 20 FFeasa IFeasa FOPTICALT otal Operating Wavelength RangeACCURACYD ominant WavelengthCorrelated Color TemperatureChromaticity(with OH3 Optical Head)REPEATABILITYD ominant WavelengthCorrelated Color TemperatureChromaticity xyHueSaturationIntensityELECTRICALS upply VoltageSupply CurrentPHYSICALD imensions of 3, 5, 10 ChannelDimensions of 20 ChannelFiber LengthFiber DiameterMinimum Bend Radius of FiberOperating Temperature Range450nm to 650nm+ 2nm @ 590nm+ 200K @ 2856K+ @ x= , y= + 1nm+ 50K @ 2856K+ < 1< 1%< 1% 100mm x 29mm x 50mm140mm x 29mm x , incl.

9 Cladding15mm0 C to +50 C450nm to 650nm+ 2nm @ 590nm+ 200K @ 2856K+ @ x= , y= + 1nm+ 50K @ 2856K+ < 1< 1%< 1% x 57mm x 55mm145mm x 57mm x , incl. cladding15mm0 C to +50 C