FireCR+ - 3DISC

Service Manual FireCR+ computed radiography Scanner Doc No.: TM-812 Rev: Aug 2013 Part No.: CR-FPM-01-003 3 DISC, FireCR, Quantor and the 3D Cube are trademarks of 3D Imaging & Simulations Corp, South Korea, and its affiliates. All other trademarks are held by their respective owners and are used in an editorial fashion with no intention of infringement. The data in this publication are for illustration purposes only and do not necessarily represent standards or specifications, which must be met by 3D Imaging & Simulations Corp. All information contained herein is intended for guidance purposes only, and characteristics of the products and services described in this publication can be changed at any time without notice.

Service Manual FireCR+ Computed Radiography Scanner 4 Components | FireCR+ Components Item Recommended Computer Requirement Operation System Microsoft Windows 7 CPU Core Duo / Core2 Processor Memory RAM 4GB or more Hard Disk 300GB Free Hard Disk Space Network 1Gbps Ethernet Video 32 bit Color Display Video Resolution 1280 x

Tags:

  Radiography, Computed, Computed radiography

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of FireCR+ - 3DISC

1 Service Manual FireCR+ computed radiography Scanner Doc No.: TM-812 Rev: Aug 2013 Part No.: CR-FPM-01-003 3 DISC, FireCR, Quantor and the 3D Cube are trademarks of 3D Imaging & Simulations Corp, South Korea, and its affiliates. All other trademarks are held by their respective owners and are used in an editorial fashion with no intention of infringement. The data in this publication are for illustration purposes only and do not necessarily represent standards or specifications, which must be met by 3D Imaging & Simulations Corp. All information contained herein is intended for guidance purposes only, and characteristics of the products and services described in this publication can be changed at any time without notice.

2 Products and services may not be available in your local area. Please contact your local sales representative for availability information. 3D Imaging & Simulations Corp. strives to provide as accurate information as possible, but shall not be responsible for any typographical error. Copyright 2013 3D Imaging & Simulations Corp, all rights reserved, printed, and published in South Korea by 3D Imaging & Simulations Corp. 815, Tamnip-Dong, Yuseong-Gu, Daejeon, Korea Tel : 82-42-931-2100 Fax : 82-42-931-2299 Website : E-mail : 3 DISC Americas 22560 Glenn Dr, Suite 116 Sterling, VA 20164 USA Tel : 1-703-430-6080 E-mail : 3 DISC Europe Gydevang, 39-41, 3450 Alleroed, Denmark Tel : 45-88-276-650 E-mail : Service Manual FireCR+ computed radiography Scanner FireCR+ | Contents 3 Contents Components.

3 4 Installation .. 5 Table Top .. 5 Wall Mount .. 6 Connections .. 8 Cassette and IP Visual Inspection .. 8 Calibration .. 9 Calibration Geometry .. 9 Calibration Procedure .. 10 Reader Diagnostics .. 12 Scanner Control .. 12 Diagnostics Window .. 13 Parts Removal/Replacement .. 14 Top Cover .. 14 Internal Cover .. 16 Main Board .. 17 Core Board .. 18 BLDC 19 Cassette Lock Motor Board .. 21 Optic Plate .. 22 Aligner Motor .. 23 Y-Axis Motor .. 25 Fiber Bundle .. 27 Internal Frame .. 30 RFID 32 Power Board .. 33 Troubleshooting .. 35 Image Quality .. 35 Mechanical Noise .. 39 Service Manual FireCR+ computed radiography Scanner 4 Components | FireCR+ Components Recommended Computer Requirement Operation System Microsoft Windows 7 CPU Core Duo / Core2 Processor Memory RAM 4GB or more Hard Disk 300GB Free Hard Disk Space Network 1 Gbps Ethernet Video 32 bit Color Display Video Resolution 1280 x 1024 Item Qty FireCR+ computed radiography Scanner 1 Fire Kit (Optional Product Identification Card) 1 14 x 17 Cassette 1 10 x 12 Cassette 1 14 x 17 Imaging Plate (Installed in the cassette) 1 10 x 12 Imaging Plate (Installed in the cassette)

4 1 IP Extractor 1 USB Interface Cable 1 Power Adapter 1 Power Cord 1 Transportation Case 1 FireCR+ User Manual 1 Service Manual FireCR+ computed radiography Scanner FireCR+ | Installation 5 Installation Table Top 1. Install on a flat and stable surface. 2. Adjust the feet underneath the scanner to ensure stability. 3. Allow 50cm (20 inches) of free space at the front for cassette insertion and 20cm (8 inches) in the back for cables and power switch access. Service Manual FireCR+ computed radiography Scanner 6 Installation | FireCR+ Wall Mount 1. Turn reader on its side. Note: Do not turn upside down! 2. Remove adjustable feet.

5 3. Attach brackets to the scanner. Note: Washers go between the brackets and the bolts. Service Manual FireCR+ computed radiography Scanner FireCR+ | Installation 7 4. Secure wall mount bracket to wall. Note: Make sure that the wall bracket is mounted level. 5. Carefully attach the scanner to the wall bracket. Note: Make sure all four bracket hooks are securely attached to the wall mount bracket. Service Manual FireCR+ computed radiography Scanner 8 Installation | FireCR+ Connections 1. Connect power supply and USB cable. Cassette and IP Visual Inspection 1. Remove the IPs from the cassettes by hooking and pulling the plate using the IP Extractor.

6 2. Visually inspect for defects (air bubbles, scratches, dust, etc.) 3. Inspect cassette housing for physical damage. Service Manual FireCR+ computed radiography Scanner FireCR+ | Calibration 9 Calibration Calibration Geometry Note: Exposure must always cover the entire cassette. Service Manual FireCR+ computed radiography Scanner 10 Calibration | FireCR+ Calibration Procedure This screen will appear: Step 1: Auto alignment Step 2: Erase Run an erase cycle to remove any residual radiation that may be left on the phosphor. Step 3: Scan Blank Without exposing the cassette, insert it into the reader and press the Scan Blank button.

7 Note: Use the same cassette for the entire calibration. Step 4, 5, & 6: Scan Low, Medium, & High Dose Expose the entire cassette at the recommended values and if the value is out of range adjust mAs until the numbers are green. Note: mAs and calibration values are directly proportional, so a 5% increase in mAs will equal a 5% increase in calibration value, etc. Step 7: Calibration! Press and wait until the software confirms that the calibration was successful. Note: Cancelling the calibration before completion will force you to start over. Service Manual FireCR+ computed radiography Scanner FireCR+ | Calibration 11 Understanding the Cal-files: Calibration files can be found in the software folder.

8 Windows Photo Viewer can view the Cal-files. We recommend that you open the High Dose calibration of each cassette size to verify an artifact free calibration. 14 x 17 Cassette Calibration 10 x 12 Cassette Calibration File Description File Description Scan Blank Scan Blank Scan Low Dose Scan Low Dose Scan Mid Dose Scan Mid Dose Scan High Dose Scan High Dose Cal0 - Cal3 compressed together Cal4 - Cal7 compressed together Good Calibration Bad Calibration (Defective IP) Service Manual FireCR+ computed radiography Scanner 12 Reader Diagnostics | FireCR+ Reader Diagnostics Scanner Control 1. Software on/off 2.

9 Run manual erasing 3. Run scan 4. Stop scan 5. Select the resolution of scan 6. Laser on/off 7. Eraser on/off 8. Stage control *(current should be 40 & 40 ) 9. Beam aligner control 10. PMT(Photo Multiplier Tube) gain 11. ADC(Analog to Digital Converter) offset 12. Diagnostics window on/off 13. Save values highlighted in orange 14. Send a report to technical support 15. Connect to remote support. Service Manual FireCR+ computed radiography Scanner FireCR+ | Reader Diagnostics 13 Diagnostics Window Gauge 1: Rotating mirror RPM (Has to be 1800 RPM +/- 10) Gauge 2: Right PSD peak Gauge 3: Right PSD balance Gauge 4: Left PSD peak Gauge 5: Left PSD balance Gauge 6: Internal frame temperature Gauge 7: Mainboard temperature Gauge 8: Eraser temperature Gauge 9: Laser module temperature Service Manual FireCR+ computed radiography Scanner 14 Parts Removal/Replacement | FireCR+ Parts Removal/Replacement Top Cover 1.

10 Turn the power off. 2. Unplug the USB cable and power cord. 3. Remove 4 front screws and 3 rear screws from the base plate ( hex). 4. Pull back cover. Service Manual FireCR+ computed radiography Scanner FireCR+ | Parts Removal/Replacement 15 5. Slowly lift cover from the back. Note: Before removing the cover, disconnect the LED connector underneath. 6. Refit in reverse order. Service Manual FireCR+ computed radiography Scanner 16 Parts Removal/Replacement | FireCR+ Internal Cover 1. Remove Top Cover. (Pg. 14) 2. Remove the 6 screws on top of the Internal Cover (#2 Phillips). Note: Some scanners have 10 screws. 3. Disconnect the cassette lock cables from the cassette lock board and feed them through the hole in the middle of the Internal Cover.