Example: bachelor of science

FortéBio Bio-layer Interferometry Kinetic Analysis Tutorial

Selvakumar Dakshnamurthy, PhD Richard Yip, PhDField Applications ScientistsFort BioBio-layer Interferometry Kinetic Analysis TutorialCONFIDENTIAL2 Introduction to Biolayer Interferometry (s. 3-12) Kinetic Analysis Basics (s. 13-18) Basic kinetics Data Acquisition (s. 19-35) Data Analysis HT (s. 36-63 )OutlineCONFIDENTIAL3 Full life-cycle offering for biomolecular interaction Analysis Label-free assays based on Bio-layer Interferometry (BLI) and Surface plasmon Resonance (SPR) platforms Instruments, consumables, software, post-sale services (one-on-one training)ForteBio is a Market Leader in Label-Free Biomolecular AnalysisOctet & BLItzPioneerBio-Layer Interferometry (BLI)SPRCONFIDENTIAL4 Octet RED96eOctet K2 Octet QKeOctet RED384 Octet HTXM olecular Weight Range> 150 Da> 150 Da> 5000 Da> 150 Da> 150 Da# Spectrometers8211616# Channels per Read828161 -96 MicroplatePositions11122 Biosensor RerackingYesYesYesYesYesRobot CompatibleNoNoNoYesYesSample Vessel Formats969696, 96 HA96 / 96HA 384 / 384TW96 / 96HA 384 / 384 TWMinimum Sample Volum

• Label-free assays based on Bio-Layer Interferometry (BLI) and Surface Plasmon Resonance (SPR) platforms • Instruments, consumables, software, post-sale services (one-on-one training) ForteBio is a Market Leader in Label-Free Biomolecular Analysis Octet & BLItz Pioneer

Tags:

  Analysis, Early, Tutorials, Kinetics, Fort, Plasmon, Interferometry, 233 bio bio layer interferometry kinetic analysis tutorial

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of FortéBio Bio-layer Interferometry Kinetic Analysis Tutorial

1 Selvakumar Dakshnamurthy, PhD Richard Yip, PhDField Applications ScientistsFort BioBio-layer Interferometry Kinetic Analysis TutorialCONFIDENTIAL2 Introduction to Biolayer Interferometry (s. 3-12) Kinetic Analysis Basics (s. 13-18) Basic kinetics Data Acquisition (s. 19-35) Data Analysis HT (s. 36-63 )OutlineCONFIDENTIAL3 Full life-cycle offering for biomolecular interaction Analysis Label-free assays based on Bio-layer Interferometry (BLI) and Surface plasmon Resonance (SPR) platforms Instruments, consumables, software, post-sale services (one-on-one training)ForteBio is a Market Leader in Label-Free Biomolecular AnalysisOctet & BLItzPioneerBio-Layer Interferometry (BLI)SPRCONFIDENTIAL4 Octet RED96eOctet K2 Octet QKeOctet RED384 Octet HTXM olecular Weight Range> 150 Da> 150 Da> 5000 Da> 150 Da> 150 Da# Spectrometers8211616# Channels per Read828161 -96 MicroplatePositions11122 Biosensor RerackingYesYesYesYesYesRobot CompatibleNoNoNoYesYesSample Vessel Formats969696, 96 HA96 / 96HA 384 / 384TW96 / 96HA 384 / 384 TWMinimum Sample Volume180 L per well180 L per well180 L per well40 L per well40 L per wellAffinity range (approximate)

2 1 mMto10 pM1 mMto10 mMto10 pM1 mMto10 pM1 mMto10 pMSample UsageNon-destructive and recoverableTemperature Control15 40 C4 C above ambient to 40 CAnalysis time per sampleUp to 12 hrswith evaporation coverUp to 4 hrs21 CFR Part 11 ComplianceAvailable as option for all systemsOctet ModelsCONFIDENTIAL5 Biosensor-based TechnologyRED96/QKeBlitzRED384/QK384 HTXUse of biosensorsis core to BLI technologyCONFIDENTIAL6 Optics box moves the biosensors to samples One biosensor tray One 96-well sample/reagent plate Upto8 interactions simultaneously in one experiment for Octet RED96eOctet 96/QKeOctet 384/HTXThe Octet Design Features Microplateformatforsamplesallowsforalarg enumberofinteractionstobestudiedinoneexp eriment.

3 Compatible with 2-96-well or 384-well sample plates. Upto16 interactionssimultaneously in one experiment for Octet RED384 Upto96 interactionssimultaneously in one experimentfor Octet HTXCONFIDENTIAL7 The Octet Dip and Read Biosensor consists of a fiber optic embedded into a polypropylene hub with a sensor-specific chemistry at the tip|----600 mm ------|Dip and Read Biosensors Two-dimensional binding surface Biocompatible Matrix(minimizes non-specific binding) Uniform Non-denaturingCONFIDENTIAL8 In BLI, light is directed down an optical fiber (the sensor) toward two interfaces separated by a thin layer at the end of the fiber The two reflected beams interfere constructively or destructively at the spectrometer CCD detector array1 2 RefTestReflections R & T are in phaseConstructive interference Strongsignal at the spectrometerReflections R & T out of phaseDestructive interference Weak signal at the spectrometerR TR TBio-Layer InterferometryCONFIDENTIAL9 Monitoring nm-shift Against Time Relative IntensityWavelength (nm)100%0nm shiftTime (sec)

4 CONFIDENTIAL10 Quantitation Direct, 1-step Sandwich ELISA mg/mL pg/mLKinetics ka, kd, KD Proteins, Abs Peptdes, oligos Small molecules FragmentsDiversity Function testing Epitope binning Rank ordering IsotypingVersatile Applications On ForteBio Label-Free SystemscBLI Cell capture based assays Dynamic Mass Redistribution (DMR) Toxicity assaysCONFIDENTIAL11 Size Range & Octet Versatility in Interaction AnalysisCellsBacteriaVirusIgMIgAIgG, IgD, IgEAntibody FragmentsProteinsPeptidesNucleotides (DNA)Small MoleculesAtoms200 nm1000 nm11,000100,0001,000,00075 nm150MW7 nm1 nmBacteriaVirusAntibody -AntigenReceptor -LigandDNA -DNA DNA -ProteinAntibody Fragment -AntigenAntigen Fusion ProteinAntibody -PeptideMultiple Antibody PairingsAntibody -Small MoleculeProtein -Small MoleculeRED96e, RED384, K2, HTXQKe,QK384 CONFIDENTIAL12 The Art of Biosensor RegenerationKinetic Characterization on the OctetCONFIDENTIAL14A+BABkonoffknm shiftDissociationAssociationThe Ideal Binding BehaviorIn a simple 1.

5 1 binding model, the association and dissociation phases are described by a single exponential functionCONFIDENTIAL15 BiosensorApplicationAntibody-Specific Capture Anti-Human IgG Fc Capture (AHC)Human IgG Fc region, Kinetic Analysis Anti-Human IgG Fc Capture (AHQ)Human IgG Fc region, quantitation Anti-Mouse Fc Capture (AMC)Mouse IgG1, 2a& 2b Fc regions, Kinetic Analysis Anti-Mouse Fc Capture (AMQ)Mouse IgG1, 2a& 2b Fc regions, quantitation Anti-Human Fab-CHI (FAB)Fab-CH1 domains of human IgG Protein A (ProA)Quantitation of various species IgG Protein G (ProG)Quantitation of variousspecies IgG ProteinL (ProL)Quantitation of IgG via kappa light chainAffinity Tag Capture Streptavidin (SA)Biotinylated ligands High Precision Streptavidin (SAX)Biotinylatedligands(4% CV loaded SA) Super Streptavidin (SSAB iotinylated ligands (high-density surface) Anti-GST (GST)GST-tagged recombinant proteins Anti-Penta HIS (HIS1)HIS-tagged recombinant proteins Anti-Penta HIS 2ndGen (HIS2)HIS-tagged recombinant proteins Ni-NTA (NTA)HIS-tagged recombinant proteinsImmobilization Amine Reactive 2nd Gen (AR2G)Covalent coupling to reactive amine groups Aminopropylsilane (APS))

6 Adsorption to hydrophobic moietiesKinetic Biosensors are HighlightedBiosensors for Kinetic AnalysisCONFIDENTIAL16 kinetics Biosensors Minimal Baseline Drift Higher Coefficient of Variation (CV)Quant Biosensors Precise CV must be within a certain range Not checked for baseline drift Short assays, high signalDifferent Manufacturing and QC CriteriaDifferences Between Kinetic and Quant BiosensorsCONFIDENTIAL17 LIGAND-Immobilized moleculeon biocompatible surface ( Biotin-tagged antibody) ANALYTE -binding partner in buffer ( protein, SM, DNA, etc.)SAMPLE WELL -bound/unbound molecules in solution(96-/384-well plate)Label-free TerminologyCONFIDENTIAL18 Binding (nm)TimeBaselineLoadingBaselineAssociati onDissociation 1 to 96 samples can be analysed in parallel Measure on rates and off rates, multiple binding models Data is displayed in real-time Experimental protocols can be customizedAutomated Kinetic Work Flow on the OctetBaselineOctet BiosensorsLigandAnalyteCONFIDENTIAL19 Kinetic Analysis Workflow with Data Acquisition V11 Double click the Data Acquisition software icon to start the programCONFIDENTIAL20 Option 2: TemplatesOption 1.

7 Wizard Use either Experiment Templates or the Experiment Wizard for assay design and optimizations such as pH scouting, regeneration scouting, up kinetics AssaysCONFIDENTIAL21 Select Blank ExperimentSetting up kinetics AssaysClick to StartSelect Basic KineticsCONFIDENTIAL22 Acquisition Steps for Basic KineticsWork from left to right, from Tab 1 to Tab Definition In this tab, all the information about the sample plate and its wells will be entered Definition In this tab, specific experimental steps are established Assignment In this tab, sensors are assigned to Experiment In this tab, you can review the steps that make up the Experiment In this tab, you can select where you would like your data saved and name the data file.

8 Run settings may also be changedCONFIDENTIAL23 Tab 1: Plate DefinitionEnter analyte concentrations in Molar terms and include at least 1 well of buffer (0 nMAnalyte Sample) for referencingSetting up kinetics MethodsRight click on highlighted wells and select well type from the dropdown dialog box CONFIDENTIAL241. To set up a dilution series, highlight the concentrations of interest, right click, and select Set Well Data3. Select a starting value, series operator, series operand, and the dilution orientation, and click OK*Tab 1: Plate Definition2. Check the box by dilution seriesCONFIDENTIAL25 Set Shake speed to 1000 rpm. Stacking the Baseline, Association, and Dissociation steps consecutively is important!

9 To assign each microplate column to the step list: Move black arrow ( )to the desired step Use mouse cursor to click on the corresponding column (number)Use Add to generate a Step ListSelect and enter Time, RPM, and Type for each step, change step name as neededTab 2: Assay Definition Setting up kinetics MethodsCONFIDENTIAL268/16 channels of simultaneous detectionTab 3: Sensor Assignment Always input biosensor lot # for troubleshoot if error happened Ensure the replace biosensor after use is checked if biosensors are being used for next experiment Click on the Fill Plate button to start from the A1 position from the Sensor Tray If need to move biosensor location, highlight the corresponding columns follow by clicking on Remove Setting up kinetics MethodsCONFIDENTIAL27 Slider Use this Black Arrow to review all the steps Make necessary changes if necessaryTab 4: Review of Experiment CONFIDENTIAL28a2.

10 Change settings of interest, including plate temperature. The Red96e has a temperature range between 15 to 40 C. Other instruments have heating up to 40 C but not cooling. For these instruments, the recommendation is to set the temperature from 2 C above ambient up to 40 C, to allow the instrument to consistently heat to the recommended , ,nametheexperimentintheExperimentrunname (subdirectory).bTab 5: Run ExperimentCONFIDENTIAL291. After confirming the assay setup, placing the biosensors in assay buffer in the pre-hyrdationplate, and pipetting the sample plate as defined in the plate definition, you are ready to hit GO hitting Go, a box will appear as a reminder to pre-hydrate your biosensors.


Related search queries