Gas Purifiers - SUPELCO

GasChromatographySUPELCO300 Order: Technical Service: Web: Purifier RecommendationsHYDROCARBONSMOISTUREOXYGE NCARRIER GAS (HELIUM, HYDROGEN OR NITROGEN)Single GC, Part time usage and low gas flowsSupelcarb-HC 120 ccMolecular Sieve 5A 200 ccSupelpure-O 120 ccGC usage and normal gas flows, up to 4 unitsOption 1: Supelcarb-HC 120 ccHigh Capacity Gas Purifier2 & OMI-21,4 High Capacity Gas Purifier2 & OMI-21,4 Option 2: Supelpure-HC 120 ccMolecular Sieve 5A 200 ccSupelpure-O 120 ccMultiple GC usage, 5 and more unitsOption 1: Supelcarb-HC 750 ccHigh Capacity Gas Purifier2 & OMI-21,4 High Capacity Gas Purifier1& OMI-21,4 Option 2: Supelcarb-HC 750 ccMolecular Sieve 5A 750 ccSupelpure-O 750 ccGC/MS UsageOption 1: SUPELCO Helium Purifier3 SUPELCO Helium Purifier3 SUPELCO Helium Purifier3 Option 2: UOP mat/sen3 UOP mat/sen3 UOP mat/sen3 AIRS ingle GC usage, up to 4 unitsSupelcarb-HC 120 ccMolecular Sieve 5A 200 ccMultiple GC usage, 5 and more unitsSupelcarb-HC 750 ccMolecular Sieve 5A 750 ccHYDROGEN (AS A FUEL GAS)GC usage and normal gas flows, up to 4 unitsSupelcarb-HC 120 ccMultiple GC Usage, 5 and more unitsSupelcarb-HC 750 cc1 One needed per GC2Do not use with hy

Gas Chromatography SUPELCO 302 Order: 1.800.325.3010 Technical Service: 1.800.359.3041 Web: www.sigma-aldrich.com/supelco Gas Delivery Gas Purifiers

Tags:

  Purifiers, Gas purifiers

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of Gas Purifiers - SUPELCO

1 GasChromatographySUPELCO300 Order: Technical Service: Web: Purifier RecommendationsHYDROCARBONSMOISTUREOXYGE NCARRIER GAS (HELIUM, HYDROGEN OR NITROGEN)Single GC, Part time usage and low gas flowsSupelcarb-HC 120 ccMolecular Sieve 5A 200 ccSupelpure-O 120 ccGC usage and normal gas flows, up to 4 unitsOption 1: Supelcarb-HC 120 ccHigh Capacity Gas Purifier2 & OMI-21,4 High Capacity Gas Purifier2 & OMI-21,4 Option 2: Supelpure-HC 120 ccMolecular Sieve 5A 200 ccSupelpure-O 120 ccMultiple GC usage, 5 and more unitsOption 1: Supelcarb-HC 750 ccHigh Capacity Gas Purifier2 & OMI-21,4 High Capacity Gas Purifier1& OMI-21,4 Option 2: Supelcarb-HC 750 ccMolecular Sieve 5A 750 ccSupelpure-O 750 ccGC/MS UsageOption 1: SUPELCO Helium Purifier3 SUPELCO Helium Purifier3 SUPELCO Helium Purifier3 Option 2: UOP mat/sen3 UOP mat/sen3 UOP mat/sen3 AIRS ingle GC usage, up to 4 unitsSupelcarb-HC 120 ccMolecular Sieve 5A 200 ccMultiple GC usage, 5 and more unitsSupelcarb-HC 750 ccMolecular Sieve 5A 750 ccHYDROGEN (AS A FUEL GAS)GC usage and normal gas flows, up to 4 unitsSupelcarb-HC 120 ccMultiple GC Usage, 5 and more unitsSupelcarb-HC 750 cc1 One needed per GC2Do not use with hydrogen3 Use with helium only4 Requires 4 ft.

2 Tubing between high-capacity gas purifier and OMI-2 How to Select PurifiersWhy use gas Purifiers ? Hydrocarbons in the carrier gas can lead to noisy baselines and irreproducible results. Moisture and oxygen canstrip phase, creating excessive bleed, damaging the GC column and shortening lifetime. Purifiers help ensure the reliable and cost-efficient performance of the GC recommendation is to use individual Purifiers to handle each of the typical contaminants. These contaminants (hydrocarbons,moisture and oxygen) are present in different impurity levels in the gases. Seldom do you find a purifier that has the correct quantitiesof each adsorbent to match the contaminants in your gases. It is better to use separate Purifiers for each contaminant and change themon a need Purifiers - These Purifiers allow a continuous monitoring of the actual gas quality that reaches the GC instrument.

3 Theygive an indication of a needed purifier change before an analytical problem is caused by insufficient gas purity. We strongly recommendthese indicators to ensure a reliable delivery of pure carrier indicators (blue beads) - Most moisture Purifiers that contain blue beads will show a color change at moisture levels above50ppm. They can be regenerated under certain indicators - Purifiers such as the OMI line of Purifiers indicate the presence of oxygen in your carrier gas. Oxygen indicators aredesigned for use primarily as an indicator. They are not designed for use as the main oxygen trap when purifying gas of low Purpose Purifiers - These Purifiers are designed for specific Purifiers - Purifiers such as the SUPELCO helium purifier and the UOP mat/sen are designed to produce high purity Helium carriergas for use with GC/MS instruments.

4 This results in greater sensitivity of the Dioxide Purifier - This purifier is used with techniques involving FTIR, IR, Total Organic Carbon (TOC) and other detectorssensitive to Carbon Dioxide (CO2). We offer a purifier that removes CO2 and shows a color change. The reaction of CO2 with theabsorbent forms moisture. We recommend the use of a moisture-removing purifier downstream of the CO2 -removing information is available in SUPELCO Bulletin 918 (T197918) on Selecting Purifiers for Gas DeliveryGas PurifiersGasChromatographySUPELCO301 Order: Technical Service: Web: Purifiers for Hydrocarbon RemovalMAXIMUM FLOW &CAPACITYSTATEDNAME OFCARRIER GASESCONTAMINANTSINLET PRESSURE(GRAMS HCLIFEPURIFERPURIFEDREMOVEDPRODUCT DESCRIPTION2(CC/MIN) (PSIG/BAR)X 75%)3(CYLINDERS)1 Supelcarb HCHe, H2, N2, Ar/CH4,C3 and higher120cc tube2000250 " 244481/4" 24449750cc tucbe10,000250 " 245641/2" 24565 Supelpure HCHe, H2, N2, Ar/CH4,C4 and higher120cc tube2000250 " 22445-U 1/4" 22446750cc tube10,000250 " 245181/2" 24519 UOPHeO2, H2O, CO2, CO364cc tube500250 Gashalocarbons,1/8" 22680-U 1/4" 22681 Purifierhydrocarbons,sulfursGas DeliveryGas PurifiersGas Purifiers for Moisture RemovalMAXIMUM FLOW &CAPACITYSTATEDNAME OFCARRIER GASESCONTAMINANTSINLET PRESSURE(GRAMS H2 OLIFEPURIFERPURIFEDREMOVEDPRODUCT DESCRIPTION2(CC/MIN) (PSIG/BAR)X 75%)(CYLINDERS)

5 1 Molecular Sieve He, H2, N2,H2O200cc tube2000250 Trapmany others1/8" 206191/4" 20618750cc tube10,000250 " 239911/2" 23992 High CapacityHe, N2, Ar/CH4O2, H2O, CO2, CO32cc tube41100150 Purifier not compatible1/8" 223961/4" 22398with air, O2, H2 OMIHe, H2, N2, Ar/CH4O2, H2O, CO2, CO,OMI-2, 15cc tube51000150 compatibleNH3, alkynes,1/8" 23906with air or O2halocarbons,OMI-4, 90cc tube51000150 ,1/8" 23909hydrogen halidesGateKeeperHe, H2, Ni2, ArO2, H2O, H2, CO2, CO,35cc cartridge1000200 compatible(removes H2as a24970with air or O2contaminant, also can500cc cartridge60,000200 H2 carrier gas)24971 Supelpure-OHe, H2, N2, Ar/CH4O2, H2O, CO2, CO,120cc tube250100 compatibleNH3, alcohols,1/8" 224491/4" 22450-Uwith air or O2alkanes, alkenes,750cc tube10,000250 , aromatics,1/4"5030881/2" 503096diethylether, halogens, sulfursUOPHeO2, H2O, CO2, CO,364cc tube500250 Gashalocarbons,1/8" 22680-U 1/4" 22681 Purifierhydrocarbons,sulfurs1 Based on 9 X 51 cylinders (218 cubic feet) of carrier gas containing 50ppm of contaminants supplying 1 chromatograph with 2 capillary columns.

6 Ask oursupplier for specifications for the cylinders you use. To correct for other volumes of gas per cylinder, multiply Stated Life (cylinders) by the number of cubic feet inyour cylinders, then divide by of adsorbent material dimensions of fitting catalog as medium weight (C8 C20) required (tube must be installed in a specially designed heater).5 Holder a flow rate of 1000 : Technical Service: Web: DeliveryGas PurifiersGas Purifiers for Oxygen RemovalMAXIMUM FLOW &CAPACITYSTATEDNAME OFCARRIER GASESCONTAMINANTSINLET PRESSURE(GRAMS O2 LIFEPURIFERPURIFEDREMOVEDPRODUCT DESCRIPTION2(CC/MIN) (PSIG/BAR)X 75%)(CYLINDERS)1 OMIHe, H2, N2, Ar/CH4O2, H2O, CO2, CO,OMI-2, 15cc tube31000150 compatibleNH3, alkynes,1/8" 23906with air or O2halocarbons, halogens, OMI-4, 90cc tube31000150 halides1/8" 23909 High CapacityHe, N2, Ar/CH4O2, H2O, CO2, CO32cc tube41100150 Purifier not compatible1/8" 223961/4" 22398with air, O2, H2 Supelpure-OHe, H2, N2, Ar/CH4O2, H2O, CO2, CO,120cc tube250100 compatibleNH3, alcohols,1/8" 224491/4" 22450-Uwith air or O2alkanes, alkenes,750cc tube10,000250 , aromatics,1/4"5030881/2" 503096diethylether,halogens, sulfursGateKeeperHe, H2, N2, ArO2, H2O, H2, CO2, CO35cc cartridge1000200 compatible(removes H2 as a24970with air or O2contaminant, also can500cc cartridge60,000200 H2 carrier gas)

7 24971 OxisorbHe, H2, N2, Ar/CH4O2, H2O, CO2, CO52cc cartridge3500090 , H2, N2, Ar/CH4O2, H2O, CO2, CO180cc cartridge3000250 " 229921/4" 22993 UOPHeO2, H2O, CO2, CO,364cc tube500250 Gashalocarbons,1/8" 22680-U 1/4" 22681 Purifierhydrocarbons, sulfurs1 Based on 9" X 51" cylinders (218 cubic feet) of carrier gas containing 50ppm of contaminants supplying 1 chromatograph with 2 capillary columns. Ask yoursupplier for specifications for the cylinders you use. To correct for other volumes of gas per cylinder, multiply Stated Life (cylinders) by the number of cubic feet inyour cylinders, then divide by of adsorbent material - dimensions of fitting - catalog required (tube must be installed in a specially designed heater).5 Data for determining oxygen capacity was not available from the manufacturer.

8 Stated life in cylinders in the manufacturer's a flow rate of 1000 : Technical Service: Web: Indicating Purifier Removesas Much Oxygen as MostNonindicating Capacity (grams)Purifier APurifier BOMIP urifier CPurifier DTrace R-S, R-ClTrace Water & Hydrocarbons713-0687-223906 OMI Indicating Purifiers Simultaneously remove O2, water vapor, CO, CO2, mostsulfur compounds, most halogen compounds, alcohols, phenolsto less than 10ppb Purify helium, hydrogen, nitrogen, argon-methane Color change indicates purifier exhaustion Glass body does not diffuse air or off-gas Ideal for Hall, ECD, GC/MS detection systems OMI-4 purifier protects multiple instruments (three times thecapacity of OMI-2 tubes)Irreversibly remove contaminants from carrier gas. Install anOMI purifier downstream from your primary gas purifying device,and tell at a glance whether or not oxygen and water vapor arebeing effectively eliminated from your system.

9 The OMI purifierwill provide point-of-use gas polishing and final visual assuranceof gas quality before the gas enters the GC. OMI purifier tubescontain Nanochem resin, developed for the demanding gas purityneeds of the semiconductor manufacturing industry. As little as1ppm of oxygen or water will change the indicating resin fromblack to tubes are easily replaced. Simply unscrew the end assemblyfrom the tube holder and replace the tube. The design preventsair from entering the new tube during : For optimum performance, we do not recommend storingOMI tubes for longer than 6 Your Column from Many Carrier Gas ContaminantsNON-INDICATINGOMIINDICATINGO XYGENCONTAMINANTPURIFIERDEVICESTRAPSO xygenYesYes1 Yes1 WaterYesNoMaybeCarbon monoxideYesNoMaybeCarbon dioxideYesNoNoAlcohols/PhenolsYesNoNoSul fur-containing compoundsYesNoNo2 Halogen-containing compoundsYesNoNo21If incoming oxygen level does not exceed compounds may poison some of these Purifier Tube323906 OMI-2 Tube Holder, 1/8" fittings323921 Seal Kit for OMI-2 Tube Holder(includes 2 Teflon seals and tool)23917 OMI-4 Purifier Tube323909 OMI-4 Tube Holder, 1/8" fittings323926 OMI-1 Replacement Tube4(includes 2 ferrules)

10 23900-U 3/8" Ferrules (pk. of 10)22311 1/4" to 1/8" Swagelok SS Reducer21517 3 First time users must order both purifier tube and corresponding is not fit OMI-2 tube holder use with OMI-1 installation kit only (kit nolonger available).DIMENSIONS OF OMI PURIFIERSOMI-2 Tube: 6" x 5/8" ODTube Holder: 10" x 1 1/2" ODEndfittings: 2 1/2" : 12" x 1" ODTube Holder: 16" x 1 1/2" ODEndfittings: 2 1/2" DeliveryGas PurifiersGasChromatographySUPELCO304 Order: Technical Service: Web: Capacity Gas PurifierNo other purifier removes both oxygen and water in such largequantities. To reliably protect your GC columns and detectorsfrom oxygen and water vapor damage, you should use a gaspurifier specifically designed to ensure maximum gas purity.