Giraffe OmniBed - Med One Group

Giraffe OmniBed Operator s ManualUser Responsibility0086 This Product will perform in conformity with the description thereof contained in this operating manual andaccompanying labels and/or inserts, when assembled, operated, maintained and repaired in accordancewith the instructions provided. This Product must be checked periodically. A defective Product should notbe used. Parts that are broken, missing, plainly worn, distorted or contaminated should be replacedimmediately. Should such repair or replacement become necessary, Ohmeda recommends that a tele-phone or written request for service advice be made to the nearest Ohmeda Regional Service Product or any of its parts should not be repaired other than in accordance with written instructionsprovided by Ohmeda and by Ohmeda trained personnel. The Product must not be altered withoutOhmeda s prior written approval. The user of this Product shall have the sole responsibility for any mal-function which results from improper use, faulty maintenance, improper repair, damage or alteration byanyone other than Inc has declared that this product conforms with the European Council Direc-tive 93/42/EEC Medical Device Directive when it is used in accordance with the instruc-tions provided in the Operation and Maintenance alpha character indicates the year of patient probe manufactureand when the serial number was assigned; A = 1997, B = 1998, C = 1999, D = 2000, etc.

If the OmniBed is being operated as an incubator, and the doors will be open for an extended length of time, pushing the Boost Air Curtain Button increases fan speed and improves open door thermal performance. An indicator on the button lights when the Boost Air Curtain is activated. Running the fan at higher speed, however, will raise

Tags:

  Omnibed

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of Giraffe OmniBed - Med One Group

1 Giraffe OmniBed Operator s ManualUser Responsibility0086 This Product will perform in conformity with the description thereof contained in this operating manual andaccompanying labels and/or inserts, when assembled, operated, maintained and repaired in accordancewith the instructions provided. This Product must be checked periodically. A defective Product should notbe used. Parts that are broken, missing, plainly worn, distorted or contaminated should be replacedimmediately. Should such repair or replacement become necessary, Ohmeda recommends that a tele-phone or written request for service advice be made to the nearest Ohmeda Regional Service Product or any of its parts should not be repaired other than in accordance with written instructionsprovided by Ohmeda and by Ohmeda trained personnel. The Product must not be altered withoutOhmeda s prior written approval. The user of this Product shall have the sole responsibility for any mal-function which results from improper use, faulty maintenance, improper repair, damage or alteration byanyone other than Inc has declared that this product conforms with the European Council Direc-tive 93/42/EEC Medical Device Directive when it is used in accordance with the instruc-tions provided in the Operation and Maintenance alpha character indicates the year of patient probe manufactureand when the serial number was assigned; A = 1997, B = 1998, C = 1999, D = 2000, etc.

2 I and O are not A 12345 Table of Contentsii6600-0342-011 01/08/01 General InformationIncubator/Warmer Operating Modes .. 1-1 Control Modes .. 1-2 Controls and Displays .. 1-2 Incubator Operation .. 1-3 Warmer Operation .. 1-4 Operating Mode Transition .. 1-5 Accessory Options .. 1-5 Alarms .. 1-11 Options Alarms .. 1-15 Alarm Table .. 1-16 Cable Connections and Mechanical Controls .. 1-18 CheckoutMechanical Checks .. 2-1 Controller Checks ..2-2 Accessory Checks .. 2-3 Operating the OmniBedBasic Operating Procedure .. 3-1 Operation as an Incubator .. 3-3 Operation as a Warmer .. 3-4 Attaching the Skin Temperature Probe .. 3-4 Additional Operating Procedures .. 3-5 Accessing the Baby .. 3-5 Raising the Canopy .. 3-6 Raising and Lowering the Bed .. 3-6 Tilting the Bed .. 3-6 Using the X-ray Tray .. 3-7 Administering Oxygen .. 3-7 Installing wristlets .. 3-7 Using the humidifier .. 3-8 Thermalink option .. 3-9 Using the In-Bed Scale .. 3-10 Mounting Accessories.

3 3-11 Gas Cylinder Holder .. 3-12 Tubing Management Arm .. 3-12 Oscillator Port Cover .. 3-12 Maintaining the OmniBedRepair Policy .. 4-1 Maintenance schedule .. 4-2 Disassembling the OmniBed for complete cleaning .. 4-3 Cleaning and disinfecting individual components .. 4-5 AccessoryAppendixSpecifications .. A-1 Performance .. A-2 Table of Contentsiiii6600-0342-011 01/08/01wWarningsBefore using the OmniBed , read through this entire manual. As with all medical equip-ment, attempting to use this device without a thorough understanding of its operation mayresult in patient or user injury. This device should only be operated by personnel trained inits operation under the direction of qualified medical personnel familiar with the risks andbenefits of this type of device. Additional precautions specific to certain proceduresare found in the text of this the Checkout procedures in this manual before putting the unit into operation. Ifthe unit fails any portion of the checkout procedure it must be removed from use not use the OmniBed in the presence of flammable anesthetics; an explosion hazardexists under these disconnect the power before performing service or maintenance proceduresdetailed in this manual.

4 Apply power only if you are specifically instructed to do so as partof the air dry the OmniBed after cleaning it with flammable agents. Small amounts offlammable agents, such as ether, alcohol or similar cleaning solvents left in the incubatorcan cause a competent individuals trained in the repair of this equipment should attempt toservice it as detailed in the Service Manual (6600-0343-000).Detailed information for more extensive repairs is included in the service manual solely forthe convenience of users having proper knowledge, tools and test equipment, and forservice representatives trained by Ohmeda : A Note provides additional information to clarify a point in the : A Caution statement is used when the possibility of damage to the : A Warning statement is used when the possibility of injury to the patient orthe operator ! Consult the manual for more B Earth TerminalxProtection Earth TerminalAlarm Silence Alternating CurrentHot surface, do not touchMAXM aximum European Union RepresentativeDefinitions1-1 6600-0342-011 08/08/01 1-1 General InformationIncubator/Warmer operating modesThe GiraffeTMOmniBedTM combines the features of an incuba-tor and a radiant warmer.

5 When used as an incubator, a fanand heater beneath the bed circulate warm air through theclosed infant compartment. The baby can be accessedthrough portholes or doors. When you require even greateraccess to the baby, the canopy can be raised and the com-partment walls can be lowered. As the canopy raises, doorsopen to expose a focused radiant heating element inside thecanopy that maintains the baby s temperature like a OmniBed allows the full range of clinical procedureswithout the disruptive intervention of transferring the baby fromone care bed to 6600-0342-011 08/08/01 1-2 General InformationControl ModesThe OmniBed has two control modes when used as either anincubator or a warmer. With the compartment closed, theOmniBed controls the inside temperature comparing airtemperature readings from the probe on the compartment wallto a setting you enter using the control panel (air mode) orfrom a probe attached to the baby s skin (baby mode).

6 Withthe canopy raised, the OmniBed controls radiant heater outputfrom a heater power percentage setting that you enter usingthe control panel (manual mode) or by using temperaturereadings from a probe attached to the baby s skin (babymode).Controls and DisplaysThe controls and displays, located at the top of the uprights atthe head of the bed, can be divided into two areas: on the leftare the buttons and numeric temperature displays that controlthe temperature regulation of the infant, on the right are thegraphics display screen and control knob that control all theaccessory can be shown in Celsius orFahrenheit degrees: the default setting is RegulationAt start up, the OmniBed prompts you to select one of thecontrol modes. If the unit was used in the last 2 hours, it willask if you would like to retain the settings you used you are using the OmniBed as an incubator (canopy closed)-you select baby or air mode. If you are using the OmniBed asa warmer (canopy open)- you select baby or manual you start with the canopy open, or open the bed during thefirst minutes after start up, and you do not select a controlmode, the graphics screen will show the Warm Up Mode message and the radiant heater will automatically run at 100%power.

7 When ten minutes has elapsed and you still have notselected a control mode, the graphics display will show the Preheat Zone message and the radiant heater will then run atthe preheat level; the default preheat level setting is 25% Alarm Light2. Alarm Silence Button3. Options Control Knob4. Graphics Screen5. Options Displays and Controls6. Temperature Regulation Controls7. ControlModeButtons8. Temperature/Power controls9. Numeric Displays10. Power Failure Indicator25%50%75%100%37c? Greater than 37 C Indicator2. Greater than 37 C Button3. Baby Mode Indicator4. Baby Mode Button5. Manual Mode Indicator6. Manual Mode Button7. Air Mode Indicator8. Air Mode Button9. Boost Air Curtain Button10. Boost Air Curtain Indicator11. Decrease Temperature/Power12. Increase Temperature/Power37c2345678910111211-3 6600-0342-011 08/08/01 1-3 General InformationIncubator operationWhen operating as an incubator, select the control mode bypushing the Air Mode Button at the bottom or the Baby ModeButton at the top of the temperature regulation Control ModeThe Air Mode Indicator lights when air mode is selected.

8 Twotemperature values are shown in the lower third of the display:the value shown in the larger numerals at the top is the airtemperature measured by the compartment probe, the valueshown below in smaller numerals is the air set temperature. Theset temperature can be adjusted using the temperature/powerbuttons located at the left of the mode buttons. Temperaturesettings from 20 to 37 C can be entered in C from 37 to 39 C can only be entered afterpressing the >37 C : When operating the OmniBed as an incubator in the airmode with a probe attached to the infant s skin, the infant stemperature will also be shown in the baby temperaturesection at the top of the Control ModeThe Baby Mode Indicator lights when the baby mode is temperature values are shown in the upper third of thedisplay: the value shown in larger numerals at the top is thetemperature measured by the probe on the infant s skin, thevalue shown below in smaller numerals is the baby set tempera-ture, which can be adjusted by using the temperature/powerbuttons located at the left of the mode buttons.

9 Temperaturesettings from 35 to 37 C can be entered in C settings from 37 to C can only be enteredafter pressing the >37 C : When operating the OmniBed as an incubator in the babymode, the compartment temperature will also be shown in the airtemperature display section at the bottom of the Air CurtainIf the OmniBed is being operated as an incubator, and thedoors will be open for an extended length of time, pushing theBoost Air Curtain Button increases fan speed and improvesopen door thermal indicator on the button lights when the Boost Air Curtain isactivated. Running the fan at higher speed, however, will raisethe noise level in the infant compartment. To deactivate theBoost Air Curtain, push the button a second time to resumeWhisper Quiet mode. The Boost Air Curtain will automaticallyshut off after twenty minutes and the indicator light will go button must be pressed again to continue for another twentyminutes at high : The fan speed may increase automatically due to otherrequirements without pushing the Boost Air Curtain button, andin these cases the button indicator will not Air Set Temperature2.

10 Air Temperature3. Air Mode Button25%50%75%100% Baby Mode Button2. Baby Set Temperature3. Baby 6600-0342-011 08/08/01 1-4 General InformationWarmer OperationWhen operating as a radiant warmer, select the control modeby pushing the Manual Mode Button in the middle or the BabyMode Button at the top of the Control ModeThe Manual Mode Indicator lights when the manual mode isselected. The power setting is indicated by the bar graph LEDin the middle third of the display. The percentage of full powercan adjusted in 5% increments using the temperature/powerbuttons located to the left of the mode : when operating the OmniBed as a warmer in themanual mode with a probe attached to the infant s skin, theinfant s temperature will also be shown in baby temperaturesection at the top of the Control ModeThe Baby Mode Indicator lights when baby mode is temperature values are shown in the upper third of thedisplay: the value shown in larger numerals at the top is thetemperature measured by the probe on the infant s skin, thevalue shown below in smaller numerals is the baby set tem-perature.