Invacare Perfecto Oxygen Concentrator IRC5PO2, IRC5PO2W, IRC5P, IRC5PW. en HomeFill System Compatible User manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference. 2016 Invacare Corporation All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by and . All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted. Making Life's Experiences Possible is a trademark in the Pine-sol and Lestoil are registered trademarks of The Clorox Company. Dawn is a registered trademark of The Proctor and Gamble Company. Contents 7 Maintenance .. 24. Service Life .. 24. Cleaning the Cabinet .. 24. 1 General .. 4 Cleaning the Cabinet Filter .. 25. Symbols .. 4 Humidifier Cleaning and Thermic Disinfection .. 25. Intended Use .. 6 Cleaning and Disinfection Between Patients.

Invacare® Perfecto₂₂₂™Oxygen Concentrator IRC5PO2, IRC5PO2W, IRC5P, IRC5PW en HomeFill® System Compatible User Manual This manual

Tags:

  Manual, Variance

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Invacare® Perfecto₂₂₂™Oxygen Concentrator

1 Invacare Perfecto Oxygen Concentrator IRC5PO2, IRC5PO2W, IRC5P, IRC5PW. en HomeFill System Compatible User manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference. 2016 Invacare Corporation All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by and . All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted. Making Life's Experiences Possible is a trademark in the Pine-sol and Lestoil are registered trademarks of The Clorox Company. Dawn is a registered trademark of The Proctor and Gamble Company. Contents 7 Maintenance .. 24. Service Life .. 24. Cleaning the Cabinet .. 24. 1 General .. 4 Cleaning the Cabinet Filter .. 25. Symbols .. 4 Humidifier Cleaning and Thermic Disinfection .. 25. Intended Use .. 6 Cleaning and Disinfection Between Patients.

2 26. Indications For Use .. 6 Preventive Maintenance Checklist .. 27. Description .. 7 8 After Use .. 28. Contraindications .. 7 Recycling Information.. 28. 2 Safety .. 8 Wear and Tear .. 28. Label Locations .. 8 9 Troubleshooting .. 29. General Guidelines .. 9 Troubleshooting .. 29. Radio Frequency Interference .. 12. 10 Technical data .. 31. 3 Components .. 13. Specifications .. 31. Component Identification .. 13. Pneumatic Diagram .. 15 11 Warranty .. 34. 4 Accessories .. 16 North American Limited Warranty .. 34. Optional Accessories .. 16. 5 Setup .. 17. Unpacking .. 17. Inspection .. 17. Storage .. 17. 6 Usage .. 18. Introduction .. 18. Select a Location .. 18. Setting Up the Concentrator .. 19. Setting Up the Humidifier .. 19. Turning the Concentrator On .. 21. Flowrate .. 21. SensO2 Oxygen Purity Indicator .. 22. Units without SensO2 .. 23. Invacare Perfecto Oxygen Concentrator 1 General Symbols in Documentation General Warning Sign Symbols The background color inside the Signal words are used in this manual and apply to hazards or unsafe triangle is yellow on product labels.

3 Practices which could result in personal injury or property damage. See the information below for definitions of the signal words. Read manual DANGER! The color of the symbol background Danger indicates an imminently hazardous situation is blue on product labels. which, if not avoided, will result in death or serious injury. No Smoking The color of the circle with diagonal WARNING! bar is red on product labels. Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious No Open Flame injury. The color of the circle with diagonal bar is red on product labels. CAUTION! Caution indicates a potentially hazardous situation Class II, Double Insulated which, if not avoided, may result in property damage or minor injury or both. Protected against solid foreign objects of Gives useful tips, recommendations and information for mm diameter and greater. efficient, trouble-free use. Protected against vertically falling water drops. Indoor Use ONLY.

4 Keep Dry 4 1143482-H-00. General 150F. MAX. Transport and Storage Temperature O2 Indicators -20F. Symbol O2 Purity Indicator Lights (LED). MIN. O2 SYSTEM GREEN Indicator Light 95%. MAX Transport and Storage Humidity OKAY. 15%. MIN O2 over 85%. Alternating Current O2 between YELLOW Indicator Light 73% to 85%. A. YELLOW Solid Type BF equipment B. YELLOW Flashing Sensor Failure. Call a qualified technician Recycle SYSTEM RED Indicator Light FAILURE. DO NOT dispose of in household waste Refer to Troubleshooting. O2 Below 73%. Electrical Hazard LX Indicators Symbol O2 Purity Indicator Lights (LED). Manufacturer I/O SYSTEM GREEN Indicator Light OKAY. Call supplier SYSTEM RED Indicator Light FAILURE. Continuous Audible Alarm Sieve . Symbols on Product GARD . Unit running Compressor Shutdown Call a qualified technician. Unit not running 1143482-H-00 5. Invacare Perfecto Oxygen Concentrator Intended Use WARNING! The intended function and use of the Invacare Perfecto Oxygen Risk of Injury or Damage Concentrator is to provide supplemental oxygen to patients with Use of this product outside of the intended use and respiratory disorders, by separating nitrogen from room air, by way specifications has not been tested and may lead to of a molecular sieve.

5 It is not intended to sustain or support life. product damage, loss of product function, or personal injury. DANGER! DO NOT use this product in any way other than Risk of Injury or Death described in the Specifications and Intended Use This product is to be used as an oxygen supplement and sections of this manual . is not intended to be life-supporting or life-sustaining. ONLY use this product if the patient is capable of DANGER! spontaneous breath, able to inhale and exhale without Risk of Injury or Death the use of a machine. While Invacare strives to produce the best oxygen DO NOT use in parallel or series with other oxygen Concentrator in the market today, this oxygen concentrators or oxygen therapy devices. Concentrator can fail to produce oxygen due to power failure or device malfunction. DANGER! ALWAYS have a backup source of oxygen readily Risk of Injury or Death available. Depending on their medical condition, patients on flow In the event the Concentrator fails to produce oxygen, rates greater than 5 L/min may be at an increased risk the Concentrator will briefly alarm signaling the patient for serious injury or death in the event of failure.

6 To switch to their backup source of oxygen. Refer to ALWAYS discuss this increased risk with your health Troubleshooting for more detail. care supplier BEFORE using this product if you are prescribed a flow rate greater than 5 L/min. Indications For Use The intended function and use of the Invacare Perfecto Oxygen Concentrator is to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve. It is not intended to sustain or support life. 6 1143482-H-00. General Description The Invacare Perfecto2 Concentrator is used by patients with respiratory disorders who require supplemental oxygen. The device is not intended to sustain or support life. The oxygen concentration level of the output gas ranges from 87%. to The oxygen is delivered to the patient through the use of a nasal cannula. The Invacare Perfecto2 Concentrator uses a molecular sieve and pressure swing adsorption methodology to produce the oxygen gas output.

7 Ambient air enters the device, is filtered and then compressed. This compressed air is then directed toward one of two nitrogen adsorbing sieve beds. Concentrated oxygen exits the opposite end of the active sieve bed and is directed into an oxygen reservoir where it is delivered to the patient. The Invacare Perfecto2 Concentrator is capable of operation by the patient in a home environment or in an institutional environment. Device operates at a nominal 120 VAC/60 Hertz supply. Service information will be available upon request to qualified technical personnel only. Contraindications There are no known contraindications. 1143482-H-00 7. Invacare Perfecto Oxygen Concentrator 2 Safety Label Locations A. Serial number label is located on the resonator B. intake assembly. Specification Label is located on the back of the C. Concentrator at the base. D. 8 1143482-H-00. Safety General Guidelines DANGER! In order to ensure the safe installation, assembly and operation of the Risk Of Death, Injury, Or Damage From Fire Concentrator these instructions MUST be followed.

8 Textiles, oil or petroleum substances, grease, greasy substances and other combustibles are easily ignited DANGER! and burn with great intensity in oxygen enriched air and Risk Of Death, Injury, Or Damage when in contact with oxygen under pressure. To avoid Improper use of the product may cause death, injury or fire, death, injury or damage: damage. This section contains important information for DO NOT SMOKE while using this device. the safe operation and use of this product. DO NOT use near OPEN FLAME or IGNITION. DO NOT use this product or any available optional SOURCES. equipment without first completely reading and DO NOT use any lubricants on Concentrator unless understanding these instructions and any additional recommended by Invacare. instructional material such as user manuals, service NO SMOKING signs should be prominently displayed. manuals or instruction sheets supplied with this Avoid creation of any spark near oxygen equipment. product or optional equipment. This includes sparks from static electricity created by If you are unable to understand the warnings, cautions any type of friction.

9 Or instructions, contact a healthcare professional, Keep all matches, lighted cigarettes, electronic dealer or technical personnel before attempting to cigarettes or other sources of ignition out of the use this equipment. room in which this Concentrator is located and away Check ALL external components and carton for from where oxygen is being delivered. damage. In case of damage, or if the product is not Keep the oxygen tubing, cord, and Concentrator out working correctly, contact a technician or Invacare from under such items as blankets, bed coverings, for repair. chair cushions, clothing, and away from heated or hot THE INFORMATION IN THIS DOCUMENT IS surfaces including space heaters, stoves, and similar SUBJECT TO CHANGE WITHOUT NOTICE. electrical appliances. CAUTION! Federal (statutory) law restricts this device sale to or on the order of a medical practitioner licensed by a governmental agency where he/she practices. ONLY a licensed medical practitioner may order the purchase or use of this device.

10 1143482-H-00 9. Invacare Perfecto Oxygen Concentrator DANGER! DANGER! Risk of Death, Injury, from Electric Shock Risk of Injury or Death To reduce the risk of burns, electrocution, death or To avoid choking or ingestion of chemicals from airway injury to persons: contamination: DO NOT disassemble. Refer servicing to qualified DO NOT use the Concentrator in the presence of service personnel. There are no user serviceable pollutants, smoke, fumes, flammable anesthetics, parts. cleaning agents, or chemical vapors. Avoid using while bathing. If continuous usage is required by the physician's prescription, the WARNING! Concentrator must be located in another room at Risk of Injury or Death least 7ft ( ) from the bath. To prevent injury or death from product misuse: DO NOT come in contact with the Concentrator Closely supervise when this Concentrator is used by while wet. or near children or impaired individuals. DO NOT place or store Concentrator where it can Monitor patients using this device who are unable to drop into water or other liquid.