PACEMAKER System Speciflcation - SQRL - Home

PACEMAKER System SpecificationCopyrightc 2007 Boston ScientificJanuary 3, 2007 DISCLAIMER: This document is provided for academic purposes only,specifically demonstration of formal methodologies and engineering Scientific disclaims any use of information contained herein for anyproduct whatsoever, disclaims safety, disclaims efficacy, and disclaims fitness forpurpose. Any person or entity who uses this document for anything assumescomplete responsibility for its to copy is granted, provided the copyright declaration and thisdisclaimer are direct any comments or questions System Specification2 Contents1 Scope.

PACEMAKER System Speciflcation 10 not respond to electrical stimulation. In addition, the presence of clinically sig-niflcant conduction disturbances may contraindicate the use of atrial pacing.

Tags:

  System, Pacemaker, Pacemaker system speciflcation, Speciflcation

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of PACEMAKER System Speciflcation - SQRL - Home

1 PACEMAKER System SpecificationCopyrightc 2007 Boston ScientificJanuary 3, 2007 DISCLAIMER: This document is provided for academic purposes only,specifically demonstration of formal methodologies and engineering Scientific disclaims any use of information contained herein for anyproduct whatsoever, disclaims safety, disclaims efficacy, and disclaims fitness forpurpose. Any person or entity who uses this document for anything assumescomplete responsibility for its to copy is granted, provided the copyright declaration and thisdisclaimer are direct any comments or questions System Specification2 Contents1 Scope.

2 Acronyms .. 52 System System Overview .. System Components .. Device Overview .. Device Controller-Monitor (DCM) Overview .. Lead System Overview .. Indications and Contraindications .. User Characteristics and Operational Environment .. Hospitals/Physician Users .. Device Usage with Other Equipment .. System Operational Life Phases .. Pre-Implant Phase .. Implant .. Predischarge Follow-up .. Routine Follow-up .. Ambulatory .. Explant .. Disposal .. 123 System Model Type.

3 Device Controller-Monitor (DCM) .. DCM User Languages .. DCM User Interface .. DCM Utility Functions .. Printed Reports .. Strip Chart Recording Support .. DCM-PG Telemetry .. Lead Support .. Pacing Pulse .. Pulse Amplitude .. Pulse Width .. Rate Sensing .. Sensitivity Adjustment .. Bradycardia Operating Modes .. No Response To Sensing (O) .. Triggered Response To Sensing (T) .. Inhibited Response To Sensing (I) .. Tracked Response To Sensing (D) .. Bradycardia States.

4 17 PACEMAKER System Permanent State .. Temporary Bradycardia Pacing .. Pace-Now State .. Magnet State .. Power-On Reset (POR) State .. Magnet Test .. Implant Data .. 194 Measured Data .. P and R Wave Measurements .. Lead Impedance Measurement .. Battery Status .. Threshold Test .. Bradycardia History .. Rate Histograms .. Rate Trending .. Recording Duration and Time Stamp .. Sensor Trending .. Real-time Electrograms .. Electrogram Viewing .. Real-time Electrogram Event Marker Annotations.

5 Atrial Markers: AS AP AT TN .. Ventricular Markers: VS VP PVC TN .. Marker Modifiers: () -Hy -Sr .. Augmentation Markers: ATR-Dur ATR-FB ATR-End PVP Faults and Error Handling .. Faults .. Errors .. 275 Bradycardia Lower Rate Limit (LRL) .. Upper Rate Limit (URL) .. Atrial-Ventricular (AV) Delay .. Paced AV Delay .. Sensed AV Delay .. Dynamic AV Delay .. Sensed AV Delay Offset .. Refractory Periods .. Ventricular Refractory Period (VRP) .. Atrial Refractory Period (ARP).

6 Post Ventricular Atrial Refractory Period (PVARP) .. Extended PVARP .. Refractory During AV Interval .. Noise Response .. Atrial Tachycardia Response (ATR) .. 31 PACEMAKER System Atrial Tachycardia Detection .. ATR Duration .. ATR Fallback .. Rate-Adaptive Pacing .. Maximum Sensor Rate (MSR) .. Activity Threshold .. Response Factor .. Reaction Time .. Recovery Time .. Hysteresis Pacing .. Rate Smoothing .. 33A Programmable Parameters34B Measured Parameters35 List of Tables1 Model Type and Lead Port.

7 132 Bradycardia Operating Modes .. 173 Battery Status and Therapy Availability .. 224 Intervals Associated with Histogrammed Events .. 235 Event Marker Annotations .. 256 Programmable Parameters for Bradycardia Therapy Modes .. 287 Programmable Parameters .. 348 Measured Parameters .. 35 PACEMAKER System Specification51 IntroductionThis System Specification for PACEMAKER defines the functions and operat-ing characteristics, identifies the System environmental performance parameters,and characterizes anticipated uses of the ScopeThis document identifies the functions that the System must perform and pro-vides a description of these functions and their primary AcronymsAPAtrial PaceASAtrial SenseARP Atrial Refractory PeriodATR Atrial Tachycardia ResponseAVAtrial-to-VentricularBOL Beginning Of (battery)

8 LifeBPM Beats Per MinuteccCardiac Cycle(s)CCIC ardiac Cycle IntervalDCM Device Controller-MonitorECG Electrocardiogram, external heart signalsEGM Electrogram, internal heart signalsEOL End Of (battery) LifeEPElectrophysiology, electrophysiologistERN Elective Replacement NearERT Elective Replacement TimeHRL Hysteresis Rate LimitICD Implantable Cardio-DefibrillatorIS-1 Industry Standard lead type 1 LRL Lower Rate LimitMSR Maximum Sensor RateNSR Normal Sinus RhythmOROperating RoomPGPulse GeneratorPOR Power-On ResetPMT PACEMAKER -Mediated Tachycardiappm Pulses Per MinutePVARP Post-Ventricular Atrial Refractory PeriodPVC Premature Ventricular ContractionSIRS ensor Indicated RateSRD Sustained Rate DurationURL Upper Rate

9 LimitVPVentricular PacePACEMAKER System Specification6 VSVentricular SenseVRP Ventricular Refractory PeriodPACEMAKER System Specification72 System System OverviewThis PACEMAKER System Specification describes the PACEMAKER -specificprogramming application and pulse generator (PG). The PACEMAKER systemsupports the following needs of patients that require bradycardia pacing support: Implantation Ambulatory Follow-up ExplantationThe PACEMAKER System : Provides dual chamber, rate adaptive bradycardia pacing support Provides historical data on device performance Provides user diagnostics through brady analysis functionsThe bradycardia analysis functions permit the following pacing measurementsand tests to be performed.

10 Lead impedance Pacing threshold P and R wave measurement Battery status Temporary brady pacing Motion sensor System ComponentsThe PACEMAKER System consists of three major components: Device (also called the pulse generator or PG) Device Controller-Monitor (DCM) and associated software LeadsA standard cardiac donut magnet is a minor System System Device OverviewThe device monitors and regulates a patient s heart device detects and provides therapy for bradycardia device provides programmable, single- and dual-chamber, rate-adaptivepacing, both permanent and adaptive rate modes.