Example: tourism industry

Pfeiffer Turbopump HiPace 80 Operating Instructions

Translation of the original instructionsOperating InstructionsENPT 0208 BEN/D (1008) HiPace 80 Turbopump2 Table of contentsTable of contents1 About this manual.. 32 Safety .. Safety precautions .. Protective equipment .. Proper use.. Improper use.. 73 Transport and storage .. Transport .. Storage .. 84 Product description .. Product identification .. Function .. Range of application ..105 Installation .. Set-up .. Preparatory work.. Assembly .. Connections to the Turbopump ..156 Operation.

Translation of the original instructions Operating Instructions PT 0208 BEN/D (1008) EN HiPace 80 Turbopump

Tags:

  Operating, Instructions, Operating instructions, Instructions operating instructions

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Pfeiffer Turbopump HiPace 80 Operating Instructions

1 Translation of the original instructionsOperating InstructionsENPT 0208 BEN/D (1008) HiPace 80 Turbopump2 Table of contentsTable of contents1 About this manual.. 32 Safety .. Safety precautions .. Protective equipment .. Proper use.. Improper use.. 73 Transport and storage .. Transport .. Storage .. 84 Product description .. Product identification .. Function .. Range of application ..105 Installation .. Set-up .. Preparatory work.. Assembly .. Connections to the Turbopump ..156 Operation.

2 Commissioning .. Operation modes .. Function description .. Monitoring of the operation conditions .. Switching off and venting .. 247 Maintenance / replacement .. intervals and responsibilities .. the Operating fluid reservoir .. the electronic drive unit .. 278 Decommissioning .. Shutting down for longer periods .. Re-starting.. Disposal .. 289 Malfunctions .. Rectifying malfunctions .. 2910 Service .. 3011 Spare parts HiPace 80 .. 3112 Accessories.. 3213 Technical data and dimensions.

3 General .. 3413 Technical data and dimensions .. Dimensions .. 36 Declaration of conformity .. 373 About this manual1 About this ValidityThis Operating manual is for customers of Pfeiffer Vacuum. It describes the func-tioning of the designated product and provides the most important information for safe use of the unit. The description follows applicable EU guidelines. All informa-tion provided in this Operating manual refer to the current state of the product's de-velopment. The documentation remains valid as long as the customer does not make any changes to the Operating Instructions can also be downloaded from Applicable docu-ments*also available via information about other certifications, if applicable, please see the signet on the product or.

4 TUVdotCOM-ID ConventionsSafety instructionsThe safety Instructions in Pfeiffer Vacuum Operating manuals are the result of risk evaluations and hazard analyses and are oriented on international certification standards as specified by UL, CSA, ANSI Z-535, SEMI S1, ISO 3864 and DIN 4844. In this document, the following hazard levels and information are considered: HiPace 80 Operating instructionsOperating Instructions "Electronic drive unit TC 110"PT 0204 BN*Declaration of conformityPart of this documentDANGERI mmediate dangerDeath or very severe injuries can NIN GPossible dangerInjuries or severe property damages can dangerInjuries or property damages can or noteCommand to perform an action or information about properties, the disregarding of which may result in damage to the this manualPictograph definitionsInstructions in the text Work instruction: here you have to do something.

5 Abbreviations usedDCU:Display and Operating unitHPU:Handheld programming unitTC:Electronic drive unit for turbopumpTPS:Mains packSymbols usedThe following symbols are used consistently throughout the diagrams: High vacuum flange Fore-vacuum flange Electric connection Cooling water connection Venting connection Sealing gas connectionProhibition of an action or activity in connection with a source of danger, the disregarding of which may result in serious of a displayed source of danger in connection with operation of the unit or to perform an action or task associated with a source of danger, the disregarding of which may result in serious Safety precautions Do not expose any body parts to the vacuum.

6 Observe all safety and accident prevention regulations. Regularly check the proper observance off all safety measures. Power supply: The Turbopump power supply must apply to the requirements of double insulation between mains input voltage and Operating voltage according to the regulations of IEC 61010 and IEC 60950. Therefore Pfeiffer Vacuum recom-mends to use exclusively original-power packs and -accessories. Only in this case Pfeiffer Vacuum is able to guarantee the compliance of the European and North American guidelines. A safe connection to the protective earthing conductor (PE) is recommended (protection class III).

7 Do not loosen any plug connection during operations. Wait for the rotor to reach standstill before peforming work on the high vacuum to informEach person involved in the installation, operation or maintenance of the vacuum pump must read and observe the safety-related parts of these Operating Instructions . The operator is obligated to make Operating personnel aware of dangers originating from the vacuum pump, the pumped medium and the entire and operation of accessoriesPfeiffer Vacuum pumps can be equipped with a series of adapted accessories. The in-stallation, operation and maintenance of connected devices are described in detail in the Operating Instructions of the individual components.

8 For information on order numbers of components, see "Accessories". Use original accessory parts the safety system against excess rotation speedTo provide the functioning of the integrated safety system for avoiding excess rotation speed, the pump must run-up from the standstill at least once a year. Switch off the pump and await the complete standstill (rotation speed = 0 Hz). Run-up the pump according to this Operating NIN GDanger - Electrical installationSafe operation after installation is the responsibility of the operator. Do not independently modify or change the pump and electrical equipment.

9 Make sure that the system is integrated in an emergency off safety circuit. Consult Pfeiffer Vacuum for special NIN GDanger of electric shockIn case of defect, the parts connected to the power supply are under voltage. Always keep the mains connection freely accessible so you can disconnect it at any Keep leads and cables well away from hot surfaces (> 70 C). Never fill or operate Turbopump with cleaning agent. Do not operate the Turbopump with open high vacuum flange. Do not carry out any unauthorized modifications or conversions to the pump. When returning the Turbopump observe the shipping Protective equipmentDetermined situations concerning the handling of vacuum pumps require wearing of personal protective equipment.

10 The owner, respectively the employer are obli-gated to provide adequate equipment to any Operating Proper use The vacuum pump may only be used to generate a vacuum. Only operate the Turbopump with an approved backing to health by hazardous substances during maintenance or installationDepending on the process vacuum pumps, components or Operating fluids can be con-taminated by toxic, reatcive or radioactive substances. Wear adequate protective equipment during maintenance and repairs or in case of NIN GRisk of injury through falling objectsWhen transporting vacuum pumps by hand, there is a danger through loads slipping and falling down.


Related search queries