PRE-INSTALLATION GUIDE for TM MODELS VS50, …

PRODUCT SPECIFICATIONSE lectricalVS50VS50 HVS80VS80 HVoltage Rating230230230230 Voltage Load Water Pressure (psi)20 - 10020 - 10020 - 10020 - 100 Flow Rate Per Pump (gal/min) w/ Rate Per Pump (gal/min) w/o Water Temperature ( F)40 - 7540 - 7540 - 7540 - 75 Power and Heat (@ 100% duty cycle)Watts per Hour1,5001,8752,4603,075 BTU per Hour5,1196,3988,39410,493 Vacuum LevelPreset at Factory (In Hg)10101010 Doctor: _____Address: _____Phone#: _____Dealer: _____Dealer Address: _____All installations must conform to local GUIDEfor MODELS VS50, VS50H, VS80 & VS80 HTMDENTAL VACUUM SYSTEMSThis VacStar Model is being installed:(AS CHECKED) VS50 VS80 VS50H VS80 HModel VS50 Model VS50 HModel VS80 Model VS80 HMEDICAL ELECTRICAL EQUIPMENTWITH RESPECT TO ELECTRICAL SHOCK, FIRE, MECHANICALAND OTHER SPECIFIED HAZARDS ONLY IN ACCORDANCE WITH UL-60601-1, CAN/CSA 66 CACLASSIFIED If voltage falls below the minimum 205 or maximum 240 during operation, a Buck/Boost Transformer must be installed.

medical electrical equipment with respect to electrical shock, fire, mechanical and other specified hazards only in accordance with ul-60601-1, can/csa c22.2 no.601.1 66ca

Tags:

  Guide, Installation, Pre installation guide

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of PRE-INSTALLATION GUIDE for TM MODELS VS50, …

1 PRODUCT SPECIFICATIONSE lectricalVS50VS50 HVS80VS80 HVoltage Rating230230230230 Voltage Load Water Pressure (psi)20 - 10020 - 10020 - 10020 - 100 Flow Rate Per Pump (gal/min) w/ Rate Per Pump (gal/min) w/o Water Temperature ( F)40 - 7540 - 7540 - 7540 - 75 Power and Heat (@ 100% duty cycle)Watts per Hour1,5001,8752,4603,075 BTU per Hour5,1196,3988,39410,493 Vacuum LevelPreset at Factory (In Hg)10101010 Doctor: _____Address: _____Phone#: _____Dealer: _____Dealer Address: _____All installations must conform to local GUIDEfor MODELS VS50, VS50H, VS80 & VS80 HTMDENTAL VACUUM SYSTEMSThis VacStar Model is being installed:(AS CHECKED) VS50 VS80 VS50H VS80 HModel VS50 Model VS50 HModel VS80 Model VS80 HMEDICAL ELECTRICAL EQUIPMENTWITH RESPECT TO ELECTRICAL SHOCK, FIRE, MECHANICALAND OTHER SPECIFIED HAZARDS ONLY IN ACCORDANCE WITH UL-60601-1, CAN/CSA 66 CACLASSIFIED If voltage falls below the minimum 205 or maximum 240 during operation, a Buck/Boost Transformer must be installed.

2 See Product Specifications. All VacStars must be wired directly from an electrical box that complies with local electrical codes to the VacStar s Electrical Connection Box. See Figure 1. Figure 1 - Interior ViewELECTRICAL CONNECTION BOXG reenBlackWhiteBlueRedPumpsPower LeadsRight(L1) Black (L2) WhiteLeft(L1) Red (L2) BlueUse 18 Gauge, 4 Conductor, Interconnect Cable Between VacStar Pump and Remote SwitchInterior Electrical Box ConnectionsConnection to 24 V Switch OnlyConnection without 24 V SwitchVacuum PumpYellow 2 Brown 4 Interconnect CableOrange 3 Remote Switch243for Future Use2 Yel4 Brn3 OrgSITE REQUIREMENTS24V CONNECTIONSE lectricalVS50VS50 HVS80VS80 HMin. Circuit Breaker Rating2 ea. 15A or 1 ea. 30A2 ea. 15A or 1 ea. 30A2 ea. 20A or 1 ea. 40A2 ea. 20A or 1 ea. 40 AWire Size AWG (Min. Gauge)2 ea. 14 or 1 ea. 102 ea. 14 or 1 ea. 102 ea. 12 or 1 ea. 82 ea. 12 or 1 ea. 8 Buck/Boost Transformer#67002#67002 #67002, 2 ea.

3 #67002, 2 Temperature(40 - 104 F 5 - 40 C)40 - 104 F (5 - 40 C)40 - 104 F (5 - 40 C)40 - 104 F (5 - 40 C)Important: Each pump should have a dedicated circuit CHARACTERISTICSM odel VS50 Model VS50 HModel VS80 Model VS80 HHeight21 inches (53 cm)Height24 inches (61 cm)Height21 inches (53 cm)Height24 inches (61 cm)Width28 inches (71 cm)Width28 inches (71 cm)Width28 inches (71 cm)Width28 inches (71 cm)Depth16 inches (41 cm)Depth16 inches (41 cm)Depth16 inches (41 cm)Depth16 inches (41 cm) WeightShipping WeightShipping WeightShipping Weight160 lbs. (73 kg)170 lbs. (77 kg)200 lbs. (91 kg)210 lbs. (95 kg)All VacStar vacuums comply with NFPA 99C level 3 requirementsUTILITY ROOM with Hydromiser UTILITY ROOMW ithout Hydromiser With or without Air/Water SeparatorVentvent to outside with 2 schedule 40 pipeAir/WaterSeparatorPower Connection Intake from Main Lineterminate with 1 FNPT fittingWater Suppy 1/2 copper tube termi-nate with 1/2 FNPT shut-off valveFloor Sink32 VACWARNING:Condensation of Water will occur in vent Accumulation of water in vent.

4 Slope piping toward : See Optional Drain Connections If the existing drain is higher than the HydroMiser outlet, the HydroMiser must be mounted so that its outlet is above the drain. The HydroMiser can be installed up to 32 above the base of the VacStar. Order Kit PN Sink32 : See Optional Drain Connections Ventvent to outside with 2 schedule 40 pipeWARNING:Condensation of Water will occur in vent Accumulation of water in vent. Slope piping toward Connection Water Suppy 1/2 copper tube termi-nate with 1/2 FNPT shut-off valve24 VACH ydromiserHydromiserIntake from Main Lineterminate with 1 FNPT fittingTypical for both VS50H and VS80 HTypical for both VS50 and VS80 All installations must conform to local DETAILS ALL INSTALLATIONS - DRAIN OPTIONSORO DO NOT use 90 elbows except where main line terminates at vacuum Use only 45 elbows to make turns in main If piping is diverted to clear an obstruction.

5 DO NOT MAKE A TURNS45 Elbow45 ElbowCLEARING AN OBSTRUCTIONMAIN LINE Sub FLOOR installation - The sub-floor plumbing layout shown below is the recommended layout for VacStar system installations and should be used whenever installation - The overhead plumbing layout shown below is the alternate layout for VacStar system installations and should be used only when unable to use the sub-floor plumbing INSTALLATIONI ndirect connection (Air gap) with a DRAIN PIPE1-1/2 InchP-TrapCLOSED VENTED DRAIND irect connection to vented drain. No traps before Notes:1. See Optional Drain Connections shown 8-Foot Maximum Height from Main Line to Consult Dental Unit Manufacturer's Guidelines for correct reduced size and height of termination of vacuum line inside junction Limit branches. Orient main line under junction box or When main line is 1-1/2" or larger, use 45 Y's and elbows Long radius 90 elbows can be used as alternates to 45 LINESee Plumbing Requirements for Main Line Diameters needed for specific units.

6 See Notes 3, 4 & Diameter RiserMinumum Slope: 1/4 inch per 10 FeetFloor SinkSee Note 1 Junction BoxRiser - See Note 2 RISER to MAIN LINE DETAIL45 Elbow45 YTo PumpMain LineFrom Dental Unit1/2 Inch RiserCeilingInterior WallNote: Consult Dental Unit Manufacturer's Guidelines for correct reduced size and height of termination of vacuum line inside junction : See Optional Drain Connections MAIN LINESee Plumbing Requirements for Main Line Diameters needed for specific Slope: 1/4 inch per 10 FeetCeilingFloor SinkJunction BoxInterior Wall1/2-Inch Diameter Riser10-Foot Maximum Height from Riser Trap to Main Line2-Inch Exhaust/VentBlack Flex Tubing2-Inch Exhaust/VentBlack Flex Tubing1/2-Inch RiserTo Main LineDental Unit1/2-InchRISER TRAP DETAIL (Using 45 Elbows)1/2 Inch RiserRISER to MAIN LINE DETAIL45 Elbow45 Elbow45 Y1/2 Inch DiameterTo STS Pump1" MinMain LinePLUMBING SITE REQUIREMENTSP lumbingVS50 & VS50 HVS80 & VS80 HMin.

7 CFM @ 0" Hg 3244 Air Exhaust with Hydromiser 2" schedule 40 pipe2" schedule 40 pipeOverhead PlumbingMain Line Dia. ID in inches1 to 1 "1 to 2"End Fitting1" FNPT1" FNPTO verhead Main Line1/2" ID1/2" IDFloor Plumbing Main Line Dia. ID in inches1 to 1 "1 to 2"End Fitting1" FNPT1" FNPTB ranch Line Dia. ID in inches 1 to 1 "1 to 1 "Note: Suction piping must slope at least a for each 10 feet of run towards the pump. Use PVC Schedule 40 or Copper Type sure to use the required pipe type for associated systemEquipment Room TemperatureThe VacStar equipment must be used in a controlled temperature environment. Maintain equipment room temperature between 40 and 105 degrees Fahrenheit. Adequate forced ventilation must be provided across the unit by placing an appropriate exhaust fan opposite an equivalent air intake vent. The fan should be higher than the associated intake vent.

8 Exhaust Vent the exhaust piping is venting to the outside of the building, precautions must be taken to protect the equipment room from weather elements and animal intrusion. This can be accomplished by using one of the three methods shown REQUIREMENTSWall-Mounted Outside Vent ProtectionShroud & ScreenRoof-Mounted Outside Vent ProtectionScreenShroud & ScreenVacStar and Hydromiser are trademarks of Air Techniques, Inc. 2008 Air Techniques, Inc. Part Number 55349, Rev. JYou can obtain more information about Air Techniques products by visiting our web site at installation information is available on our Authorized Air Techniques Dealer web site at YOUR PRACTICEAir Techniques Walt Whitman Road, Melville, NY 11747-3062 Website: Coast: 1-800-AIR-TECH (1-800-247-8324) West Coast: 1-800-822-2899