Pressure-Volume Curves of the Respiratory System

Pressure-Volume Curves of the Respiratory SystemR Scott Harris MDIntroductionEquation of MotionStatic Versus Dynamic Pressure-Volume CurvesStatic Compliance, Chord Compliance, Specific Compliance, and thePressure- volume CurveMeasurementSupersyringe MethodConstant-Flow MethodMultiple-Occlusion MethodPhysiologic MeaningLung Versus Chest WallSupine PostureHysteresisAlterations in Disease StatesObesityAcute Respiratory Distress SyndromeCongestive Heart FailureEmphysemaAsthmaInterstitial Lung DiseaseIntra-Abdominal HypertensionSummaryThe quasi-static Pressure-Volume (P-V) curve of the Respiratory System describes the mechanicalbehavior of the lungs and chest wall during inflation and deflation. To eliminate resistive andconvective acceleration effects, the measurement of volume and pressure must be performed duringshort periods of apnea or during very slow flow. There are 3 main techniques for acquiringquasi-static P-V Curves : the supersyringe method, the constant flow method, and the multiple-occlusion (or ventilator) method.

Pressure-Volume Curves of the Respiratory System R Scott Harris MD Introduction Equation of Motion Static Versus Dynamic Pressure-Volume Curves Static Compliance, Chord Compliance, Specific Compliance, and the

Tags:

  System, Volume, Respiratory, Curves, Volume curves of the respiratory system

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Pressure-Volume Curves of the Respiratory System

1 Pressure-Volume Curves of the Respiratory SystemR Scott Harris MDIntroductionEquation of MotionStatic Versus Dynamic Pressure-Volume CurvesStatic Compliance, Chord Compliance, Specific Compliance, and thePressure- volume CurveMeasurementSupersyringe MethodConstant-Flow MethodMultiple-Occlusion MethodPhysiologic MeaningLung Versus Chest WallSupine PostureHysteresisAlterations in Disease StatesObesityAcute Respiratory Distress SyndromeCongestive Heart FailureEmphysemaAsthmaInterstitial Lung DiseaseIntra-Abdominal HypertensionSummaryThe quasi-static Pressure-Volume (P-V) curve of the Respiratory System describes the mechanicalbehavior of the lungs and chest wall during inflation and deflation. To eliminate resistive andconvective acceleration effects, the measurement of volume and pressure must be performed duringshort periods of apnea or during very slow flow. There are 3 main techniques for acquiringquasi-static P-V Curves : the supersyringe method, the constant flow method, and the multiple-occlusion (or ventilator) method.

2 For the information to be interpreted correctly, one must under-stand the interaction between the lungs and the chest wall, the effects of the supine position, and themeaning of hysteresis. The P-V curve has been studied in many disease states, but it has beenapplied most extensively to patients with acute Respiratory distress syndrome, in hopes that it mightallow clinicians to customize ventilator settings according to a patient s individual respiratorymechanics and thus protect the patient from ventilator-induced lung injury. However, lack ofstandardization of the procedure used to acquire P-V Curves , difficulties in measuring absolute lungvolume, lack of knowledge regarding how to use the information, and a paucity of data showing abenefit in morbidity and mortality with the use of P-V Curves have tempered early enthusiasmregarding the clinical usefulness of the quasi-static P-V words: lung mechanics, compli-ance, lung recruitment, Pressure-Volume curve, mechanical ventilation; acute Respiratory distress syn-drome; waveforms.

3 [Respir Care 2005;50(1):78 98. 2005 Daedalus Enterprises]78 RESPIRATORYCARE JANUARY2005 VOL50 NO1 IntroductionIn spontaneously breathing subjects the diaphragm andchest wall together form a mechanical pump that movesair in and out of the lungs to exchange oxygen and carbondioxide to and from the blood. When the Respiratory sys-tem fails and mechanical ventilation is necessary, the re-spiratory System s mechanical behavior can change in char-acteristic ways that can be assessed by the mechanicalventilator or with simple equipment available in respira-tory care departments. The quasi-static Pressure-Volume (P-V) relationship is one aspect of mechanical behaviorthat has been used to gain information about the way thelungs deform during breathing in health and has long been hoped that with mechanically ventilatedpatients the P-V curve would allow the clinician to diag-nose lung disease, customize ventilator settings, follow thecourse of disease, and make prognoses.

4 Surprisingly, de-spite over a half-century of research on P-V Curves , westill have a limited understanding of the meaning of theP-V relationship. So, though some of those goals havebeen realized, much research still needs to be done beforeP-V Curves can be widely used of MotionThe equation of motion describes the pressure change atthe airway opening during breathing. The equation as-sumes one degree of freedom, meaning that the lung ex-pands equally in all directions (isotropic expansion). It iswritten as:PAO VC V R V I Pmus(1)in which PAOis the pressure at the airway opening (mouthor endotracheal tube), Pmusis the pressure generated by therespiratory muscles, V is lung volume , C is Respiratory - System compliance, V is gas flow, R is airway resistance,V is convective gas acceleration, and I is impedance. Whatis evident from that equation is that a static P-V curveeliminates the resistive and impedance effects on pressure,such that only the compliance is assessed.

5 Thus, static P-Vcurves are also called compliance Curves . Also evident isthe Respiratory muscle contribution to pressure. If a subjectis not sufficiently sedated (or paralyzed), the measuredP-V curve may not be representative of the lung compli-ance properties alone and may include effects of the re-spiratory Versus Dynamic Pressure-Volume CurvesWhenever one measures pressure when airflow hasstopped, there is always a continual change in pressure lowering in an exponential fashion. Because we must al-low the subject to breathe, the measurement must be in-terrupted after a few seconds. Therefore, the System neverreaches truly static conditions, so we obtain what has beencalled a quasi-static P-V curve. This means that theresultant curve depends somewhat on how long the clini-cian waits for static conditions. It is possible to use a slowflow rate to approximate static conditions, but the flowrate must be 9 L/min to largely eliminate the pressurechange from resistive elements of the Respiratory ,2 The confusing term dynamic compliance has beenused to mean various things, but it originally referred tocompliance calculated during a tidal breath at the points ofzero flow on the dynamic P-V loop.

6 Some investigatorshave made arguments that perhaps P-V measurementsshould be made during breathing rather than in quasi-staticconditions, since that might be more representative of themechanical behavior that the lung experiences duringbreathing. Ranieri et al3used the shape of the inspiratorypressure-time curve during constant flow to calculate a stress index, which, they argue, can detect overdisten-tion or recruitment. Lichtwarck-Aschoff et al4advocateusing the slice method to estimate static compliancefrom a dynamic P-V loop, using a linear resistance-com-pliance model. Karason et al5developed a method to cal-culate the alveolar pressure during dynamic/therapeuticconditions, which they call the dynostatic pressure. Theyargue that this method has the advantage that it provides abreath-by-breath analysis of the P-V relationship, withoutrequiring ventilator et al6studied the effect of increasing constantflows on the shape of the inspiratory P-V loop.

7 They mea-sured a quasi-static P-V curve, using the supersyringe tech-nique, and compared it with inspiratory P-V Curves mea-sured with 10, 30, and 50 L/min constant inspiratory flowsin dogs with oleic-acid-induced lung injury. They foundthat the curve consistently shifted to the right at all in-spiratory flows tested (Fig. 1). They also found that duringthe measurement of the dynamic P-V Curves there wasincreased volume at the beginning of the dynamic inspi-rations while on positive end-expiratory pressure (PEEP),which, they concluded, was recruitment from tidal venti-R Scott Harris MD is affiliated with the Pulmonary and Critical CareUnit, Department of Medicine, Massachusetts General Hospital, HarvardMedical School, Boston, Scott Harris MD presented a version of this article at the 34th RESPI-RATORYCAREJ ournal Conference, Applied Respiratory Physiology: Useof Ventilator Waveforms and Mechanics in the Management of CriticallyIll Patients, held April 16 19, 2004, in Cancu n, : R Scott Harris MD, Pulmonary and Critical Care Unit,Bulfinch 148, Massachusetts General Hospital, 55 Fruit Street, BostonMA 02114.

8 E-mail: OF THERESPIRATORYSYSTEMRESPIRATORYCARE JANUARY2005 VOL50 NO179lation independent of PEEP. From that study it is clear thatmeasuring P-V Curves with tidal ventilation at high flow( 10 L/min) still leaves a large pressure change fromresistive components, and, furthermore, that there is aPEEP-independent volume -recruitment effect. This lastpoint will be discussed below, in the section that describesthe multiple-occlusion method for measuring P-V Compliance, Chord Compliance, SpecificCompliance, and the Pressure-Volume CurveIt is important when discussing P-V Curves to under-stand the difference between the termsstatic compliance,chord compliance, andspecific compliance. Static com-pliance usually implies compliance calculated from 2 vol-ume points during quasi-static conditions. This is usuallydone by performing an end-inspiratory hold on the venti-lator and measuring the pressure after a few seconds.

9 Thetidal volume (VT) is divided by the pressure change. Thiscompliance measurement assumes a linear relationship be-tween volume and pressure, but the P-V relationship is notlinear, so care must be taken in interpreting changes instatic compliance measured that way. For example, anincrease in static compliance after a PEEP increase on aventilator could mean simply a shift in location on the P-Vcurve, rather than a change in intrinsic properties of thelung (Fig. 2). In plane geometry, a chord is a line seg-ment that connects 2 points on a curve. Therefore, staticcompliance calculated this way is actually chord compli-ance. It would be more precise to say static chord com-pliance. Specific compliance is compliance that is nor-malized by a lung volume , usually total lung capacity(TLC) or functional residual capacity (FRC). Thus, giventhe same volume excursion, a child will have lower chordcompliance than an adult, but they will have the samespecific compliance.

10 Specific compliance is often used tobetter assess the intrinsic elastic properties of the lung, byeliminating the effects of variations in lung are 3 main techniques for constructing a quasi-static P-V curve: the supersyringe method, the constant-flow method, and the multiple-occlusion MethodThe supersyringe method consists of connecting a su-persyringe to the end of the endotracheal tube after allow-ing the Respiratory System to reach relaxation lung volume ,and then measuring airway pressure and insufflated vol-ume. The syringe s plunger is moved in regular steps (usu-ally about 100 mL each), with 2 3-second pauses to allowfor quasi-static conditions (Fig. 3), and then a Pressure-Volume plot is drawn (Fig. 4). Usually, when airway pres-sure reaches 40 cm H2O, inflation is stopped and deflationis performed in the same way. This method is simple andallows construction of both inflation and deflation also has disadvantages: it requires additional equipment,disconnecting the patient from the ventilator, and the pa-tient has to be sedated and paralyzed.