Product Data Sheet00813-0100-3499, Rev ABOctober 2019 rosemount 499 ACL-01 Free Chlorine SensorThe simplest way to reliably measure free Chlorine in waterThe rosemount 499 ACL-01 Sensor is designed to measure free Chlorine in water without samplepretreatment. These amperometric sensors are ideal for use in drinking water simple free Chlorine design Measure free Chlorine without having to deal with messyand expensive reagents. Continuous pH correction compatible when used with anauxiliary pH Sensor . Integral Pt-100 resistance temperature device (RTD) allowsfor automatic temperature compensation.

Rosemount 499ACL-01 Free Chlorine Sensor July 2017 July 2017 Rosemount 499ACL-01 Free Chlorine Sensor 4 www.Emerson.com/RosemountLiquidAnalysis www.Emerson.com ...

Tags:

  Rosemount, Sensor

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Product Data: 499ACL-01 Free Chlorine Sensor - emerson.com

1 Product Data Sheet00813-0100-3499, Rev ABOctober 2019 rosemount 499 ACL-01 Free Chlorine SensorThe simplest way to reliably measure free Chlorine in waterThe rosemount 499 ACL-01 Sensor is designed to measure free Chlorine in water without samplepretreatment. These amperometric sensors are ideal for use in drinking water simple free Chlorine design Measure free Chlorine without having to deal with messyand expensive reagents. Continuous pH correction compatible when used with anauxiliary pH Sensor . Integral Pt-100 resistance temperature device (RTD) allowsfor automatic temperature compensation.

2 Operates in samples up to 122 F (50 C) and 65 psig. Sensor measures free Chlorine consistent with EPAR egulation of maintenance and installation 1-in. male national pipe thread (MNPT) process connectionssuitable for mounting in a low flow cell installed in asidestream sample. Replacing the membrane and electrolyte solution is fast andeasy and requires no special tools or fixtures. Variopol (VP6) cable connection options, for quick cable-to- Sensor release, eliminates cable informationRosemount 499 ACL-01 Free Chlorine Sensors are available witheither an EMI protected integral cable or a Variopol (VP6)interconnecting cable (sold separately).

3 These sensors aregenerally mounted in a flow cell installed in a sidestreamsample. Three replacement membrane assemblies, three O-rings, and a 4 oz. (125 mL) bottle of electrolyte solution areprovided with each 499 ACL-01 engineering 12 rosemount 499 ACL-01 October 1: rosemount 499 ACL-01 Ordering InformationOptionDescription499 ACLC hlorine SensorSensor type01 Free chlorine02 Total Chlorine (used with rosemount TCL)03 MonochloramineTransmitter compatibility54 rosemount 1056, 1066-DO, 56, and 5081-A compatibleCable options_Integral 25-ft.

4 ( m) optimum EMI/RFI protected cableVPSensor with VP6 fittings (requires VP interconnecting cable)02 Integral 15-ft. ( m) optimum EMI/RFI protected cable03 Integral 33-ft. (10 m) optimum EMI/RFI protected cable50 Integral 50-ft. (15 m) optimum EMI/RFI protected cable06 Integral 100-ft. (30 m) optimum EMI/RFI protected cableSpecial_No selection99 Special (consult factory)Calibration and conformance certificates - optional levelCCCertificate of Calibration (no test data given)LCLoop Calibration Certificate ( Sensor and transmitter calibrated together, with testdata)ECElectronic Calibration Certificate ( Sensor calibrated against factory instrument, withtest data)SpecificationsTable 2: Sensor SpecificationsPhysical characteristicsSpecificationsRange0 to 10 ppm (mg/L) as Cl2.

5 For higher ranges, consult the to 65 psig (101 to 549 kPa abs)Temperature (operating)32 to 122 F (0 to 50 C)Process connection1-in. male national pipe thread (MNPT)Wetted partsNoryl , Viton , platinum, polyethersulfone, polyester, wood, and siliconeAccuracyAccuracy depends on the accuracy of the chemical test used to calibrate the to For samples having pH between and , consult the factory. Samples withranges below require no pH 2019 rosemount 499 ACL-01 rosemount 499 ACL-013 Table 2: Sensor Specifications (continued)Physical characteristicsSpecificationspH correctionUse continuous pH correction (requires an auxiliary pH Sensor ) if sample pH varies morethan pH (peak to peak).

6 If pH variability is less or seasonal, the pH Sensor is generally (typical)Sample conductivity> 50 S/cmInterferencesPeroxides, permanganate, and chloraminesResponse time22 sec to 95% of final reading at 77 F (25 C)Electrolyte oz. (25 mL), approximatelyElectrolyte life3 months (approximately); for best results, replace electrolyte lengthSee Table 1 for standard cable length (maximum)300 ft. (91 m), up to 100 ft. ( m) is flowFlow through: 2 to 5 gpm ( to 19 L/min)Open channel: 1 ( m/sec)Low flow cell (PN 24091-00): 8 to 15 gph (30 to 57 L/hr)Low flow cell (PN 24091-01): 2 to 5 gph ( to 19 L/hr)Weight/shipping weight1 lb.

7 ( kg)Table 3: Other SpecificationsTypePNWetted materialsProcessconnectionMaximumtempera tureMaximumpressure1 -in. tee23567-00 CPVC and Buna N;body is schedule 80 CPVC1 -in. socket122 F (50 C)65 psig (549 kPaabs)2-in. tee915240-03 PVC and Buna N;body is schedule 80 PVC -in. NFPT120 F (49 C)60 psig (515 kPaabs)915240-041-in. NFPT915240-051 -in. NFPTLow flow cell (1)24091-00 and24091-01: containsa bubble sweepingnozzle to keepbubbles fromaccumulatingagainst , 316stainless steel, andsiliconeCompression fittingfor -in. or nationalpipe thread (FNPT)158 F (70 C)90 psig(722 kPa abs)Valved rotameter9390004 for usewith low flow cellPN24091-01 Flow: to 5 gph( to 19 L/hr)Acrylic, 316stainless steel, andViton -in.

8 NFPT (316stainless steel)150 F (65 C)100 psig (858 kPaabs) rosemount 499 ACL-01 October 3: Other Specifications (continued)TypePNWetted materialsProcessconnectionMaximumtempera tureMaximumpressureValved rotameter196-898754 Flow: 2 to 10 gph( to 76 L/hr)Polycarbonate, 315stainless steel,brass, and Buna N1/8 in. NFPT (brass)54 C (130 F)100 psig (858 kPaabs)(1)Temperature and pressure specifications for the low flow cell exceed the temperature and pressure specifications for the 2019 rosemount 499 ACL-01 rosemount 499 ACL-015 Dimensional drawingsFigure 1: Standard Sensor with Integral CableA.

9 Sensor cableB. 1-in. national pipe thread (NPT), two placesC. 1-3 openingD. Pressure equalizing port, two placesE. -in. NPT, electolyte fill plugF. Membrane retainerRosemount 499 ACL-01 October 2: Sensor with Variopol Cable ConnectorA. Variopol receptacle (female connector)B. Variopol plug (male connector)C. 1-in. NPT, two placesD. 1-3/16-in. wrench openingE. Pressure equalizing port, two placesF. -in. NPT electrolyte fill plugG. Membrane retainerLength of assembled Sensor is in. (236 mm).October 2019 rosemount 499 ACL-01 rosemount 499 ACL-017 Figure 3: Low Flow Cell (PN 24091-00)A.

10 OutletB. InletThe bubble-sweeping nozzle in the inlet is not 499 ACL-01 October 4: Flow-Through Tee (1 -in.) (PN 23567-00)Allow at least in. (90 mm) head room for Sensor 2019 rosemount 499 ACL-01 rosemount 499 ACL-019 Figure 5: Flow-Through Tee (2 in.) (PN 915240-03, -04, -05)Allow at least 6 in. (150 mm) head room for Sensor 499 ACL-01 October 4: Connector Cable (Required for All First Time Installations)Part numberDescription23747-02 Cable, VP 6, conductivity/amperometric, 10-ft. ( m)23747-03 Cable, VP 6, conductivity/amperometric, 50-ft.