Transcription of Project No: INSPECTION & TEST RECORD …
1 Project No: Project Name:Document No:1 Tag Number:System:23 Manufacturer:Data Sheet No:45 Model Number:P&ID No:67 Range:Eng Unit:Layout Drawing No:89101112 INSPECTION CHECKLIST13 Check Instrument tagplate details against data sheet14 Check Instrument has no physical damage15 Check no water ingress/condensation1617 CALIBRATION1819202122232425262728 ReadingInputN/APUNCHITEMINSPECTION & TEST RECORDINSTRUMENT calibration ( transmitter )Raise input signal and RECORD corresponding output signal at readings of 0, 25, 50, 75, 100% of rangeLower input signal and RECORD corresponding output signal at readings of 100, 75, 50, 25, 0% of rangeCheck above results are within manufacturers stated accuracy, otherwise adjust and re-do items aboveITEMSOK2829303132333435363738394041 42434445464748 Completed ByWitnessed ByAccepted ByNameSignDate% SpanActual1007550250 Error %% SpanActualActual% SpanError %FallingRising