Example: stock market

Schneider Electric Foxboxo Measurement & Instrumentation ...

& InstrumentationProductCatalogContact Swanson Flo | | 800-288-7926 Specifying Your Instrument or Accessory You can easily specify many instruments and accessories described in this catalog. Sections covering our most popular items include all the technical data you need to know for most applications. To specify the appropriate item, simply follow the step-by-step procedure at the end of each description. Your Foxboro representative can help Example:Choose the product the information needed to complete the order, just contact your Foxboro representative, or call, in North America:1-866-746-6477, or International: Give us the ordering information using this catalog. Then tell us where you want the order shipped and when you need it delivered. We ll do the Help? If you have questions, let us know.

This transmitter measures absolute pressure and transmits a 4-20 mA, 1 to 5 Vdc, or digital output signal over a pair of wires. For complete specifications, refer to Product Specification Sheets PSS 2A-1C13

Tags:

  Pressure, Transmitter, Absolute, Absolute pressure

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Schneider Electric Foxboxo Measurement & Instrumentation ...

1 & InstrumentationProductCatalogContact Swanson Flo | | 800-288-7926 Specifying Your Instrument or Accessory You can easily specify many instruments and accessories described in this catalog. Sections covering our most popular items include all the technical data you need to know for most applications. To specify the appropriate item, simply follow the step-by-step procedure at the end of each description. Your Foxboro representative can help Example:Choose the product the information needed to complete the order, just contact your Foxboro representative, or call, in North America:1-866-746-6477, or International: Give us the ordering information using this catalog. Then tell us where you want the order shipped and when you need it delivered. We ll do the Help? If you have questions, let us know.

2 We re ready and eager to assist you. Applications are a Foxboro specialty. We can help you figure out the best instrument for your catalog available ask your Foxboro representative for catalog CFT51 How to Order Specify model number CFT51 followed by order code for each selectionMass Flowtube InterfaceFoxboro CFS10 and CFS20 Series MountingRemote Mounted ..ENominal Supply Voltage120 to 240 V ac, 50 or 60 to 36 V dc ..BOrderingiContact Swanson Flo | | 800-288-7926iiTable of Contents 32145678910111213 pressure and Differential pressure Flow DP Primary Elements (Integral and Compact Orifice) Flow In-Line Flow Meters (Coriolis, Vortex, Mag Flow) Temperature and Humidity Configurator/Communicator Analytical Positioners Liquid Level, Density, or Interface Recorders Controllers Indicators Accessories and Supplies General Information Contact Swanson Flo | | 800-288-79261-1 pressure The following chapters contain Product Specifications of the Instruments.

3 IAP10 absolute pressure Transmitters Direct MountIAP20 absolute pressure Transmitters Bracket MountIGP10 Gauge pressure Transmitters Direct MountIGP20 Gauge pressure Transmitters Bracket MountIGP25 Gauge pressure Transmitters Multirange for Wide Adjustment (400:1 Turndown)IGP50 Gauge pressure Transmitters Premium PerformanceIGP60 Gauge pressure Transmitters Premium PerformanceIDP10 Differential PressureIDP15 Differential pressure - Draft RangeIDP25 Differential pressure Multirange for Wide Adjustment Capability (400:1 Turndown)IDP31 Differential pressure 100 msec Response TimeIDP32 Differential pressure High Static pressure IDP50 Differential pressure Premium PerformanceIMV25 Multivariable transmitter AP, DP, TIMV30 Multivariable transmitter AP, DP, T, with Flow CalculationsIMV31 Multivariable transmitter AP, DP, T, with Level CalculationsIPI10 P to I (Pneumatic-to-Current) ConverterPressure Seals Remote and Direct-Connect Seals, including Flange Mount for Level, and Both Sanitary and Pulp & Paper Industry ConnectionsContact Swanson Flo | | 800-288-7926 Pressure1 Maximum Overrange ( absolute )C MPa 45 psi bar or kg/cm2 D MPa 450 psi bar orkg/cm2E 31 MPa 4500 psi 315 bar or kg/cm2 Range Limits ( absolute )

4 C 0 & MPa 0 & 30 psi 0 & bar or kg/cm2 D 0 & MPa 0 & 300 psi 0 & 21 bar orkg/cm2E 0 & 21 MPa 0 & 3000 psi 0 & 210 bar or kg/cm21-2 IAP10 IAP10 I/A Series Intelligent absolute pressure Transmittersg Choice of Mounting Stylesa IAP10 for compact light weight and direct-to-process mounting (bracket optionally available) g Rugged & Dependablea Field-proven silicon strain gauge technologya Corrosion-resistant epoxy finishg Superior Performancea Accuracy to of spana Ambient temperature effects to ( URL+ ) span per 28 C (50 F)g Choice of Electronics Modulesa Intelligent HART, FoundationFieldbus, FoxCom, and4-20 mA versionsa Economical 4-20 mA and 1 to 5 Vdc versions g LCD Indicator/Pushbutton Configuratora Optional on Foundation Fieldbus,FoxCom/4-20 mA, andHART/4-20 versions; Standard on4-20 mA and 1 to 5 Vdc versionsg Standard Warranty 5 YearsFunctional SpecificationsSensor Temperature Limits: DC200: -46 & +121 C (-50 + 250 F) FC77: -29 & +85 C (-20 & +185 F)Ambient Temperature Limits: DC200: -40 +85 C (-40 & +185 F) FC77: -29 & +85 C (-20 & +185 F)Electrical Classification: Various agency certifications for Zone and Division hazardous locations.

5 Refer to Product Specification sheets for complete transmitter measures absolute pressure and transmits a 4-20 mA, 1 to 5 Vdc, or digital output signal over a pair of complete specifications, refer to Product Specification Sheets PSS 2A-1C13 A, B, C, D, E, J, K, and Limits Code Span LimitsC & MPa 1 & 30 psi & bar or kg/cm2D & MPa 10 & 300 psi & 21 bar or kg/cm2 E & 21 MPa 100 & 3000 psi & 210 bar or kg/cm2 Span, range and overrange limits:Direct Connected absolute Press. IAP10 Version Output Choices Configure From -Da FoxCom Digital aI/A Series WorkstationaFoxCom/4 to 20 mA aHand-Held TerminalaPersonal ComputeraOptional Pushbuttons-Ta HART/ 4 to 20mA aHART Communicator aWorkstationaPersonal Computer-FaFoundation Fieldbus aWorkstation-Aa 4 to 20mA a Standard Pushbuttons-Va 1-5 Vdc aStandard PushbuttonsOutput signal and configuration:Contact Swanson Flo | | 800-288-7926 Pressure1-3 Physical SpecificationsMaterial Combinations and Value Package: Refer to How to Order for material versions available.

6 For exceptional value and corrosion resistance, the standard material combination with the lowest price is 316L ss Process Connection with 316L ss Fill Fluid: Dow Corning dimethylsiloxane (DC 200) or fluorinated hydrocarbon (3M Fluorinert FC77), as Classification: Meets IEC IP66 & NEMA Type SpecificationsAccuracy (Includes Linearity, Hysteresis, and Repeatability): Version Output Signal Accuracy in % of Calib. Span-D or TDigital 4 to 20 mA to 20 mA 1 to 5 Vdc to Product Specification Sheets for accuracies atsmall spans (less than 10% of URL).IAP10 How to Order Specify model number IAP10 Electronic Versions and Output Signals4-20 to 20 mA/HART ..TFoundation Fieldbus ..F4 to 20 mA ..A1 to 5 V dc ..VStructure Code Select from one of the following eight Only (no seals)ProcessSensorConnection Sensor Fill Fluid Connection Type316L ss Co-Ni-Cr Silicone NPT External Thread, NPT Internal Thread.

7 20316L ss Co-Ni-Cr Fluorinert NPT External Thread, NPT Internal Thread ..21316L ss 316L ss Silicone NPT External Thread, NPT Internal Thread ..22316L ss 316L ss Fluorinert NPT External Thread, NPT Internal Thread ..23316L ss Hastelloy C Silicone NPT External Thread, NPT Internal Thread ..30316L ss Hastelloy C Fluorinert NPT External Thread, NPT Internal Thread ..31 Hastelloy C Hastelloy C Silicone NPT External Thread, NPT Internal Thread ..32 Hastelloy C Hastelloy C Fluorinert NPT External Thread, NPT Internal Thread .. transmitter Only (no seals) Process SensorConnection Sensor Fill Fluid Connection Type 316L ss 316L ss Silicone NPT External Thread, NPT Internal Thread ..52 316L ss 316L ss Fluorinert NPT External Thread, NPT Internal Thread.

8 53 316L ss Hastelloy C Silicone NPT External Thread, NPT Internal Thread ..60 316L ss Hastelloy C Fluorinert NPT External Thread, NPT Internal Thread ..61 Hastelloy C Hastelloy C Silicone NPT External Thread, NPT Internal Thread ..62 Hastelloy C Hastelloy C Fluorinert NPT External Thread, NPT Internal Thread .. with Sanitary Connection(1) Process SensorConnection Sensor Fill Fluid Connection Type316L ss 316L ss NEOBEE M-20 in Tri-Clamp ..TA316L ss 316L ss NEOBEE M-20 in Tri-Clamp ..T2316L ss 316L ss NEOBEE M-20 in Tri-Clamp ..T3316L ss Hastelloy C276 NEOBEE M-20 in Tri-Clamp ..TB316L ss Hastelloy C276 NEOBEE M-20 in Tri-Clamp ..T4316L ss Hastelloy C276 NEOBEE M-20 in Tri-Clamp ..T5316L ss 316L ss NEOBEE M-20 Mini Tank Spud Type, 1 in extension.

9 M1316L ss 316L ss NEOBEE M-20 Mini Tank Spud Type, 6 in extension ..M6316L ss 316L ss NEOBEE M-20 Mini Tank Spud Type, 9 in extension ..M9316L ss 316L ss NEOBEE M-20 1 in Threaded Spud Type ..PX316L ss 316L ss NEOBEE M-20 in Threaded Spud Type ..PZContact Swanson Flo | | 800-288-7926 Pressure11-4 IAP10 4. transmitter with Pulp & Paper Connection(1) Process Sensor Connection Sensor Fill Fluid Connection Type 316L ss 316L ss Silicone Sleeve Type, 1 inch nominal ..PA 316L ss 316L ss Silicone Threaded Type, 1 inch nominal ..PB 316L ss 316L ss Silicone Sleeve Type, 1 inch nominal ..PC 316L ss 316L ss Silicone Threaded Type, 1 inch nominal ..PD 316L ss Hastelloy C276 Silicone Sleeve Type, 1 inch nominal ..PE 316L ss Hastelloy C276 Silicone Threaded Type, 1 inch nominal.

10 PF 316L ss Hastelloy C276 Silicone Sleeve Type, 1 inch nominal ..PG 316L ss Hastelloy C276 Silicone Threaded Type, 1 inch nominal ..PH 316L ss Hastelloy C276 Silicone Threaded Type, 1 inch nominal ..PJ (fits Ametek spud) 5. transmitter Prepared for Foxboro Model Coded Seals(2) transmitter Prepared for Foxboro Direct Connect Seal; Silicone Fill in Sensor(3) ..D1 transmitter Prepared for Foxboro Remote Mount Seal; Silicone Fill in Sensor(4) ..S3 6. Transmitters Prepared for non-Foxboro Seals transmitter Prepared for Remote Seal; Silicone Fill in Sensor ..SC transmitter Prepared for Remote Seal; Fluorinert Fill in Sensor ..SD 7. Flameproof transmitter Prepared for Foxboro Model Coded Seals(2) Flameproof transmitter Prepared for Direct Connect Seal; Silicone Fill in Sensor(3).