Servo-i Ventilator System Service Manual CRITICAL CARE

Servo-i Ventilator SystemService ManualCRITICAL CARES ervo-i Ventilator SystemImportant1 - 2 Service ManualRevision 02 NotesImportantServo-i Ventilator SystemRevision 02 Service Manual1 - 3 Contents1. Introduction ..3. Description of Disassembling and Service procedures ..6. Preventive Ventilator SystemImportant1 - 4 Service ManualRevision 02 General Service documentation for the Servo-i VentilatorSystem consists of: User's Manual . The User's Manual is anindispensable complement to the ServiceManual for proper servicing. Service Manual Installation Instructions spare Parts information. Documentation for all optional equipmentincluded in the Servo-i System is also available. The Servo-i Ventilator System is referred to asthe Servo-i troughout this Manual . There are two serial number labels on the unit: One label is attached to the Patient Unit closeto the supply gas inlets.

The User's manual is an indispensable complement to the Service Manual for proper servicing. – Service Manual – Installation Instructions – Spare Parts information. – Documentation for all optional equipment included in the Servo-i System is also available. • The Servo-i Ventilator System is referred to as the Servo-i troughout this ...

Tags:

  Manual, Part, Spare, Spare parts

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Servo-i Ventilator System Service Manual CRITICAL CARE

1 Servo-i Ventilator SystemService ManualCRITICAL CARES ervo-i Ventilator SystemImportant1 - 2 Service ManualRevision 02 NotesImportantServo-i Ventilator SystemRevision 02 Service Manual1 - 3 Contents1. Introduction ..3. Description of Disassembling and Service procedures ..6. Preventive Ventilator SystemImportant1 - 4 Service ManualRevision 02 General Service documentation for the Servo-i VentilatorSystem consists of: User's Manual . The User's Manual is anindispensable complement to the ServiceManual for proper servicing. Service Manual Installation Instructions spare Parts information. Documentation for all optional equipmentincluded in the Servo-i System is also available. The Servo-i Ventilator System is referred to asthe Servo-i troughout this Manual . There are two serial number labels on the unit: One label is attached to the Patient Unit closeto the supply gas inlets.

2 The serial numberstated on this label is the ID number of thePatient Unit. The serial number is also stored inthe SW memory as the ' System ID'. One label is attached to the rear side of theUser Interface close to the On/Off switch. Theserial number stated on this label is the IDnumber of the User Interface. System version number can be found in theStatus window on the User Interface. Make surethat the version of the User's manualcorresponds to the System inside a box is used to highlight importantinformation. In addition to the Important information givenhere and in the related documents (e. g. in theUser's Manual ), always pay attention toapplicable local and national regulations. Responsibility for the safe functioning of theequipment reverts to the owner or user in allcases in which Service or repair has been doneby a non-professional or by persons who are notemployed by or authorized by Maquet, and whenthe equipment is used for other than its notices Before disassembling or assembling of theServo-i, make sure that the: On/Off switch is set to Off.

3 Mains power cable is disconnected. Gas supply is disconnected (wall and/orcylinder). Battery modules are disconnected. The Servo-i is cleaned according toinstructions in the User's Manual , chapter'Routine cleaning' and chapter 'Regularmaintenance', section 'Extended cleaning ofinspiratory channel'. With power supply connected to the Servo-i ,there are energized electrical components insidethe unit. All personnel must exercise extremecaution if fault tracing or adjustments areperformed with power supply connected andwith user interface and patient unit used in this Manual ESD sensitive components. Whenhandling ESD-sensitive devices,established procedures must beobserved to prevent damage. Special waste. Discard disposable,replaced and left-over parts inaccordance with appropriateindustrial and environmentalstandards. Recycling. Recycle if facilities may not beavailable in all areas.

4 Technical training. Refers to theTechnical training supplied byMaquet. Service contract. Refers to theService contract supplied Ventilator SystemRevision 02 Service Manual1 - 5To the responsible Service personnel The contents of this document are not any significant difference is found between theproduct and this document, please contactMaquet for further information. We reserve the right to modify products withoutamending this document or advising the user. Only personnel trained and authorizedby Maquet shall be permitted toperform installation, Service ormaintenance of the Maquet genuine spare parts mustbe used. PC boards ( spare parts) must always bekept in a package for sensitive electronicdevices. Maquet will not otherwise assumeresponsibility for the materials used, the workperformed or any possible consequences ofsame. The device complies to standards andrequirements as stated in the ' Servo-i VentilatorSystem User's Manual '.

5 Installation Only personnel trained and authorizedby Maquet shall be permitted to installthe Servo-i . The installation andhanding over procedures are describedin the ' Servo-i Ventilator System Installation Instructions'.Functional check After any installation, maintenance or serviceintervention in the Servo-i , perform a 'Pre-usecheck' according to instructions in the 'Servo-iVentilator System User's Manual '. Service The Servo-i must be serviced at regularintervals by personnel trained andauthorized by maintenance or Service must benoted in a log book provided. It is recommended that maintenanceand Service is done as a part of aservice contract with Maquet. Preventive maintenance must be performed atleast once every year as long as the unit is notused more than normal. Normal operation isestimated to correspond to approx. hoursof operation. Details are found in this ServiceManual, chapter Preventive maintenance.

6 The Battery modules shall be replaced everythree years. The internal Lithium batteries (on PC 1771 andPC 1772) shall be replaced every five years. Worn-out batteries must be recycled ordisposed of properly according to localregulations. Recycle facilities may notbe available in all areas. Batteries must not be disposed of withordinary waste. Discard all otherdisposable, replaced and left-over partsin accordance with appropriateindustrial and environmental standards. When working with ESD sensitivecomponents, always use a groundedwrist band and a grounded worksurface. Adequate Service tools mustalways be Ventilator SystemImportant1 - 6 Service ManualRevision 02 Environmental declarationPurposeThis environmental declaration is for a Servo-ibasic unit including the carrier and one codes within brackets refers to theFunctional Block Diagram in chapter with special environmental concernComponents listed below shall be disposed of inan environmentally safe circuit boards PC 1770 Main back-plane PC 1771 Control, including a Lithium battery (C) PC 1772 Monitoring, including Lithium battery (M) PC 1775 Plug-and-Play back-plane (P) PC 1777 Panel (U) PC 1778 DC/DC & Standard connectors (P) PC 1780 Pneumatic back-plane (I) PC 1781 Pressure transducer, 2 pcs (T) PC 1784 Expiratory channel (F) PC 1785 Expiratory channel connector (E) PC 1786 Expiratory channel cassette (E) PC 1789 Remote alarm connector (A)

7 Other electronics TFT assembly including backlight (U) Touch screen (U) O2 cell, containing Pb (I) Air module, containing multiple PC boards (I) O2 module, containing multiple PC boards (I) AC/DC Converter, containing PC boards (P) Expiratory cassette (E) Expiratory valve coil (E) Safety valve pull magnet (I)Construction materialsThe construction materials used in Servo-i in % ofthe total total 77% Aluminium 70% Steel, zink, brass 8%Polymeric material total 9% PA (Polyamide) POM (Polyoxymethylene) SI (Silicone) TPE (Thermoplastic elastomer) PUR (Polyurethane) ABS (Acrylicnitrilebutadienstyrene) EPDM (Ethylenepropylenedienemonomer) PTFE (Polytetrafluoroethylene) FPM (Fluororubber) NBR (Nitrilerubber) PP (Polypropylene) PVC (Polyvinyl chloride) PS (Polystyrene)Electronics total 14% Accumulators Nickel Metalhydride Printed circuit boards, cables very small amounts Sterile filter paper of glass fibre1 ImportantImportantServo-i Ventilator SystemRevision 02 Service Manual1 - 7 Articles of consumption1.

8 Bacteria filter2. Filters for the gas modules3. Filter for the inspiration pressure transducer4. Filter for the O2 cell5. Nozzle units for the gas modules6. Battery modules7. Lithium batteries8. Expiratory cassette9. Expiratory cassette membrane10. O2 cell11. Backlight 1: Consumption approximately 250 2 5: Changed approx. every 6 7: Changed approx. every 8 11: Changed when consumptionThe power consumption depends on the operatingmode and whether the internal batteries are beingfast or trickle chargingTrickle chargingIn operation70 W38 WStandby65 W33 WOff35 W6 WNoise levelLess than 50 materialsThe amounts of packing materials will varydepending on customer for packing: Loading pallet. Fulfils the USA requirements7 CFR May 25 th 1995. Corrugated cardboard Stretch film of Polyethylene, PE. Shock-absorbing material of expandedpolyethylene, EPE, or expanded polypropylene,EPP.

9 Clamps of Polyethylene, Ventilator SystemImportant1 - 8 Service ManualRevision 02 Notes1 Servo-i Ventilator SystemIntroductionRevision 02 Service Manual2 - 122. IntroductionMain units .. 2 - 2 User Interface .. 2 - 4 Patient 2 - 6 Servo-i software 2 - 9 General .. 2 - 2 - 2 - 2 - 9 System ID .. 2 - 9 Only personnel trained and authorizedby Maquet shall be permitted to performinstallation, Service or maintenance ofthe sure to prepare the Servo-i properly beforedisassembling and assembling. Refer to section'Hazard notices' in chapter 'Important'.Any Service or maintenance must be noted in alog disposable, replaced and left-over partsin accordance with appropriate industrial andenvironmental any installation, maintenance or serviceintervention in the Servo-i , perform a 'Pre-usecheck'. Refer to the ' Servo-i Ventilator System User's Manual ' for Ventilator System2 - 2 Service ManualRevision 022 Main unitsThe Servo-i is available in different mainconfigurations: Infant Adult UniversalThese main configurations are as standard equippedwith a number of ventilation modes suitable for eachpatient category.

10 Further ventilation modes can beinstalled via software Option Servo-i can be divided into the following mainunits: User Interface. The User Interface contains allcontrols used to set the ventilation and monitoringparameters. Ventilation parameters as well as otherimportant information are shown on the UserInterface display. Patient Unit. The Patient Unit contains pneumaticsand electronics for gas supply to the supply and battery back-up is alsocontained in the Patient Control cable connects the User Interface andthe Patient Servo-i shown in the illustration is mounted ontothe optional Servo-i Mobile UnitUser InterfaceServo-iMobile CartServo-i Ventilator SystemIntroductionRevision 02 Service Manual2 - 32A number of optional equipment can be added to theServo-i Ventilator System . For further information,refer to the documents listed Mobile cart Mobile cart with drawers Mobile cart without drawers Mobile cart for Compressor to: Servo-i User's Manual Servo-i Mobile cart Installation InstructionsServo Ultra Nebulizer, Servo-iRefer to: Servo-i User's Manual Servo Ultra Nebulizer, Servo-i InstallationInstructionsCompressor MiniRefer to: Servo-i User's Manual Compressor Mini Operating Manual Compressor Mini Service Manual Compressor Mini Installation InstructionsServo-i HolderRefer to: Servo-i User's Manual Servo-i Holder Installation InstructionsServo-i Shelf baseRefer to.