Order no.: F M MAGFLO Electromagnetic flowmetersTransmitter types MAG 5000, MAG 6000 Operating ManualEdition 05/2006 - Revision 05s*083R9175*Technical Documentation (handbooks, instructions, manuals etc.) on the complete productrange SITRANS F can be found on the internet/intranet on the following links:English: []2 SITRANS F M MAGFLO Current outputActive current0-20 mA, 4-20 mA or 4-20 mA + alarm (Power supplied from flowmeter)Load< 800 ohmTime sec. adjustableDigital outputFrequency0-10 kHz, 50% duty cycleTime sec.

4 SITRANS F M MAGFLO®®® 2.1 Transmitter MAG 5000 & MAG 6000 connection diagram Special cable with indi-vidual wire shields (shown with dashed lines) are only

Tags:

  Strain, Magflo, Sitrans f m magflo

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of SITRANS F M MAGFLO - Lesman Instrument Company: …

1 Order no.: F M MAGFLO Electromagnetic flowmetersTransmitter types MAG 5000, MAG 6000 Operating ManualEdition 05/2006 - Revision 05s*083R9175*Technical Documentation (handbooks, instructions, manuals etc.) on the complete productrange SITRANS F can be found on the internet/intranet on the following links:English: []2 SITRANS F M MAGFLO Current outputActive current0-20 mA, 4-20 mA or 4-20 mA + alarm (Power supplied from flowmeter)Load< 800 ohmTime sec. adjustableDigital outputFrequency0-10 kHz, 50% duty cycleTime sec.

2 AdjustableActive pulse24 V DC, 30 mA, 1 K Rload 10 K , short-circuit-protected (Power supplied from flowmeter)Passive pulse3-30 V DC, max. 110 mA, 200 Rload 10 K (Powered from connected equipment)RelayTime constantChangeover relay, time constant same as current time constantLoad42 V AC/2 A, 24 V DC/1 ADigital input11-30 V DC, Ri = K Activation time50 V DC = mA, I30 V DC = 7 mAFunctionsFlowrate, 2 totalizers, low flow cut-off, empty pipe cut-off, flow direction, error system, operating time,uni/bidirectional flow, limit switches, pulse output, control for cleaning unit and batching2)

3 Galvanic isolationAll inputs and outputs are galvanically isolatedCut-offLow of maximum flowEmpty pipeDetection of empty pipe 1)TotalizerTwo eight-digit counters for forward, net or reverse flowDisplayBackground illumination with alphanumerical text, 3 20 characters to indicate flowrate, totalizedvalues, settings and faultsReverse flow indicated by negative signTime constantTime constant as current output time constantZero point adjustmentAutomaticElectrode input impedance> 1 x 1014 Excitation frequencySensor size depending pulsating DC current (125 mA)Ambient temperatureDisplay version during operation.

4 5 to 120 FBlind version during operation: 5 to 140 FDuring storage: 40 to 160 F (Relative humidity max 95%)Custody transfer approvalPTB (cold water)DANAK OIML R752)DANAK OIML R1172)(hot water)(cold water/milk, beer etc.)CommunicationStandardPrepared for client mounted add-on modules2)OptionalHART, Profibus PA & DP, Modbus RTU, CANopen, DeviceNet as add-on module2), HART (MAG 5000)Integral mountEnclosure materialFiberglass-reinforced polyamideEnclosure ratingNEMA 4X / 6 (3 ft.

5 Submersion for 30 min)Mecanical load18-1000 Hz random, G rms in all directions to EN 60068-2-36 Rack mountEnclosure materialStandard rack mount of aluminum/steel (DIN 41494)Width: inchHeight: inchEnclosure ratingNEMA 2 Mechanical loadVersion: 1 G, 1-800 Hz sinusoidal in all directions to EN 60068-2-36 EMC performanceEmission: EN 50081-1 (Light industry)Immunity: EN 50082-2 (Industry)Power supply115-230 V AC +10% to 15%, 50-60 Hz11-30 V DC or 11-24 V ACFuse: 250 V 500 mA TPower consumption230 V AC: 17 VA24 V DC: 9 W, IN = 380 mA, start-up peak current = 8A (30 msec.

6 12 V DC: 11 W, IN = 920 mA, start-up peak current = 4A (250 msec.)ApprovalsFM Class 1, division 2, ULc general purpose1) Special cable required in separate mounted installation, 2) MAG 6000 only1. Transmitter MAG 5000 & MAG 6000 (1/4" to 78")MAG 5000 accuracy 6000 accuracy ( for MAG 3100 W sensor) 3 SITRANS F M MAGFLO 1. SpecificationsOutput characteristicsBidirectional modeUnidirectional mode0-20 mA4-20 mAFrequencyPulse outputRelayPower downActiveError relayNo errorErrorLimit switch or1 set point2 set pointsdirection switchLow flowIntermediate flow(Reverse flow)High flowHigh flow/(Forward flow)Low flowBatch on digitaloutput(MAG 6000 only)Batch on relayHoldBatch(MAG 6000 only) characteris-tics MAG 5000 &MAG Sensor cables andconductivity Minimum acceptdata for cableCoil cableElectrode cableBasic dataNo.

7 Of conductors23 Min. sqr. mm2/20 mm2/22 gageScreenYesYesMax. cable loopMedia temperature:< 210 C40 < 390 C6 ofCompact installation: Liquids with an electrical conductivity 5 a conductivity between 5 and 10 S/cm, the repeatability may degrade to of actual :NoteFor detection of empty sensor the min. conductivity must always be 20 S/cm and the of electrode cable when remote mounted is 150 ft. Special cable must be remote mounting in Ex applications special cable cannot be used, empty sensor cannotbe detected and the electrically conductivity must be 30 remote mounted CT installations the max.

8 Cable length is 600 cableSpecial cable4 SITRANS F M MAGFLO TransmitterMAG 5000 & MAG 6000connection diagramSpecial cable with indi-vidual wire shields (shownwith dashed lines) are onlyrequired when using emptypipe function with lowconductivity process (see"Specifications")Safety NoteOnly qualified personnelshould perform wiring orrepairs, and only when thetransmitter is not transmitter inaccordance with allrelevant NEC and Electrical connectionPotential Hazards / GroundingThe mains protective earth wire must be connected to the PE terminal in accordance with thediagram (class 1 power supply).

9 Mechanical countersWhen mounting a mechanical counter to terminals 57 and 58 (active output), a 1000 F capacitormust be connected to the terminals 56 and + is connected to terminal 56 and capacitor to terminal cablesIf long cables in noise environment, we recommend to use screened cablesDotted connections only to be when using special electrode supply 115 to 230 V AC from building installation Class II. A switch or circuit-breaker ( A) shall be included in the building installation. It must be in close proximity to the equipmentand within easy reach of the operator, and it shall be marked as the disconnecting device for theequipment.

10 5 SITRANS F M MAGFLO 3. Installation of Integral installationMAG 5000 andMAG 6000 Step 1 Remove and discard the terminal box lid of the PG cable glands for the supply andoutput 2 Remove the two black plug assemblies for coiland electrode cables in the terminal box andconnect them to their corresponding terminalnumbers on the connection 3 Connect an earth wire between PE on connec-tion board and bottom of terminal the 2 pin connector and 3 pin connectoras earlier version the 3 pin connector was a 5