Standard Practice for Design, Manufacture and Material ...

Designation: E 747 97An American National StandardStandard Practice forDesign, Manufacture and Material grouping Classification ofWire Image Quality Indicators (IQI) Used for Radiology1 This Standard is issued under the fixed designation E 747; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or This practice2covers the design , Material groupingclassification, and Manufacture of wire image quality indica-tors (IQI) used to indicate the quality of radiologic This Practice is applicable to X-ray and This Practice covers the use of wire penetrameters as thecontrolling image quality indicator for the Material thicknessrange from to 152 mm ( to in.)

Designation: E 747 – 97 An American National Standard Standard Practice for Design, Manufacture and Material Grouping Classification of Wire Image Quality Indicators (IQI) Used for Radiology1 This standard is issued under the fixed designation E 747; the number immediately following the designation indicates the year of

Tags:

  Design, Practices, Standards, Material, Manufacture, Standard practice for design, Manufacture and material grouping classification, Grouping, Classification

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Standard Practice for Design, Manufacture and Material ...

1 Designation: E 747 97An American National StandardStandard Practice forDesign, Manufacture and Material grouping Classification ofWire Image Quality Indicators (IQI) Used for Radiology1 This Standard is issued under the fixed designation E 747; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or This practice2covers the design , Material groupingclassification, and Manufacture of wire image quality indica-tors (IQI) used to indicate the quality of radiologic This Practice is applicable to X-ray and This Practice covers the use of wire penetrameters as thecontrolling image quality indicator for the Material thicknessrange from to 152 mm ( to in.)

2 The values stated in inch-pound units are to be regardedas Standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this Standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to Referenced standards :B 139 Specification for Phosphor Bronze Rod, Bar, andShapes3B 150 Specification for Aluminum Bronze Rod, Bar, andShapes3B 161 Specification for Nickel Seamless Pipe and Tube4B 164 Specification for Nickel-Copper Alloy Rod, Bar, andWire4B 166 Specification for Nickel-Chromium-Iron Alloys(UNS N06600, N06601, and N06690) and Nickel-Chromium-Cobalt-Molybdenum Alloy (UNS N06617)Rod, Bar, and Wire4E 1025 Practice for design , Manufacture , and MaterialGrouping Classification of Hole-Type Image Quality Indi-cators (IQI) Used for Radiology5E 1316 Terminology for Nondestructive Examinations53.

3 The definitions of terms in TerminologyE 1316, Section D, relating to gamma and x-radiology, shallapply to the terms used in this Wire IQI The quality of all levels of examination shall be deter-mined by a set of wires conforming to the following require- Wires shall be fabricated from materials or alloysidentified or listed in accordance with Other materials maybe used in accordance with The IQI consists of sets of wires arranged in order ofincreasing diameter. The diameter sizes specified in Table 1 are1 This Practice is under the jurisdiction of ASTM Committee E-7 on Nonde-structive Testing and is the direct responsibility of Subcommittee onRadiographic Practice and edition approved July 10, 1997.

4 Published November 1997. Originallypublished as E 747 80. Last previous edition E 747 ASME Boiler and Pressure Vessel Code applications see related PracticeSE-747 in Section II of that Book of ASTM standards , Vol Book of ASTM standards , Vol Book of ASTM standards , Vol 1 Wire IQI Sizes and Wire Identity NumbersSET ASET BWire Diameterin. (mm)Wire IdentityWire Diameterin. (mm)Wire ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( )11 SET CSET DWire Diameterin. (mm)Wire IdentityWire Diameterin. (mm)Wire ( ) ( ) h( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) (8)21 AThe wire may be used to establish a special quality level as agreedupon between the purchaser and the 2 Wire Diameter Tolerances, (mm)Wire Diameter (d), mmTolerance, <d# <d# <d# <d# <d# <d# ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United American Society for Testing and MaterialsLicensed by Information Handling ServicesCOPYRIGHT American Society for Testing and MaterialsLicensed by Information Handling Servicesestablished from a consecutive series of numbers taken ingeneral from the ISO/R 10 series.

5 The IQI shall be fabricatedin accordance with the requirements specified in Figs. 1-8 andTables 1-3. IQIs previously manufactured to the requirementsof Annex A1 may be used as an alternate provided all otherrequirements of this Practice are Image quality indicator (IQI) designs other than thoseshown in Figs. 1-8 and Annex A1 are permitted by contractualagreement. If an IQI set as listed in Table 1 or Annex A1 ismodified in size, it must contain the grade number, set identity,and essential wire. It must also contain two additional wiresthat are the next size larger and the next size smaller asspecified in the applicable set listed in Table Each set must be identified using letters and numbersmade of industrial grade lead or of a Material of similarradiographic density.

6 Identification shall be as shown on or Annex A1, unless otherwise specified by Image Quality Indicator (IQI) When selecting IQI s for procurement, the followingfactors should be Determine the alloy group(s) of the Material to Determine the thickness or thickness range of thematerial(s) to be Select the applicable IQI s that represent the requiredIQI thickness(s) and alloy(s).6. Image Quality The quality level required using wire penetrametersshall be equivalent to the 2-2T level of Practice E 1025 forhole-type IQI s unless a higher or lower quality level is agreedupon between purchaser and supplier. Table 4 provides a list ofvarious hole-type IQI s and the diameter of wires of corre-sponding equivalent penetrameter sensitivity (EPS) with theapplicable 1T, 2T, and 4T holes in the IQI.

7 This table can beused for determining 1T, 2T, and 4T quality levels. AppendixX1 gives the equation for calculating other equivalencies In specifying quality levels, the contract, purchase order,product specification, or drawing should clearly indicate thethickness of Material to which the quality level applies. Carefulconsideration of required quality levels is particularly Material Materials have been designated in eight groups basedon their radiographic absorption characteristics: groups 03, 02,and 01 for light metals and groups 1 through 5 for The light metal groups, magnesium (Mg), aluminum(Al), and titanium (Ti) are identified 03, 02, and 01 respec-tively, for their predominant alloying constituent.

8 The materialsare listed in order of increasing radiation The heavy metal groups, steel, copper-base, nickel-base, and kindred alloys are identified 1 through 5. Thematerials increase in radiation absorption with increasingnumerical Common trade names or alloy designations have beenTABLE 3 Wire Diameter Tolerances, (in.)Wire Diameter (d), , <d# <d# <d# <d# <d# <d# 1 Set A/Alternate 1E 7472 COPYRIGHT American Society for Testing and MaterialsLicensed by Information Handling ServicesCOPYRIGHT American Society for Testing and MaterialsLicensed by Information Handling Servicesused for clarification of the pertinent The materials from which the IQI for the group are tobe made are designated in each case and these IQI s areapplicable for all materials listed in that group.

9 In addition, anygroup IQI may be used for any Material with a higher groupnumber, provided the applicable quality level is Group 01 Image quality indicators (IQI s) shall be made oftitanium or titanium shall be the predominant alloying Use on all alloys of which titanium is the predomi-nant alloying Group 02 Image quality indicators (IQI s) shall be made ofaluminum or aluminum shall be the predominant Use on all alloys of which aluminum is the predomi-nant alloying Group 03 Image quality indicators (IQI s) shall be made ofmagnesium or magnesium shall be the predominant Use on all alloys of which magnesium is thepredominant alloying Group 1 Image quality indicators (IQI s) shall be made ofcarbon steel or Type 300 series stainless 2 Set A/Alternate 2 FIG.

10 3 Set B/Alternate 1E 7473 COPYRIGHT American Society for Testing and MaterialsLicensed by Information Handling ServicesCOPYRIGHT American Society for Testing and MaterialsLicensed by Information Handling Use on all carbon steel, low-alloy steels, stainlesssteels, and manganese-nickel-aluminum bronze (Superston). Group 2 Image quality indicators (IQI s) shall be made ofaluminum bronze (Alloy No. 623 of Specification B 150) orequivalent, or nickel-aluminum bronze (Alloy No. 630 ofSpecification B 150) or Use on all aluminum bronzes and all nickel-aluminum Group 3 Image quality indicators (IQI s) shall be made ofnickel-chromium-iron alloy (UNS No.)