T-Series Straight-Tube Coriolis Flow and Density Meters

Product Data SheetPS-00371, Rev. IAugust 2015 micro motion T-Series Meters offer the best performance of any Straight-Tube Coriolis meter. The single Straight-Tube design is self-draining, andallows the meter to be cleaned or sterilized in place (CIP/SIP). The straightflow path also resists plugging, and can be flow measurement in a single straight tube flow meter Built-in balance bar provides the best single straight tube mass flowmeasurement to reduce variability in process controlComprehensive hygienic application coverage Easy to CIP and SIP with EHEDG certified, 3-A authorized,ASME BPE design Diameter matches standard process tubing for draining in anyorientation Fast product change-over with self-draining design and no profileeffects Single flow path is easy to clean mechanically Highly-polished surface finish for ultra-pure fluidsSuperior reliability No moving parts to wear or replace minimizes maintenance forlong-term reliabilityMicro motion T-SeriesCoriolis Flow and Density MetersHigh performancecompact drainableCorioli

Product Data Sheet PS-00371, Rev. I August 2015 Micro Motion® T-Series meters offer the best performance of any straight- tube Coriolis meter. The single straight-tube design is self-draining, and allows the meter to be cleaned or sterilized in place (CIP/SIP).

Tags:

  Series, Micro, Tubes, Straight, Motion, Coriolis, Micro motion, 174 t series, Series straight tube coriolis, Straight tube coriolis

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of T-Series Straight-Tube Coriolis Flow and Density Meters

1 Product Data SheetPS-00371, Rev. IAugust 2015 micro motion T-Series Meters offer the best performance of any Straight-Tube Coriolis meter. The single Straight-Tube design is self-draining, andallows the meter to be cleaned or sterilized in place (CIP/SIP). The straightflow path also resists plugging, and can be flow measurement in a single straight tube flow meter Built-in balance bar provides the best single straight tube mass flowmeasurement to reduce variability in process controlComprehensive hygienic application coverage Easy to CIP and SIP with EHEDG certified, 3-A authorized,ASME BPE design Diameter matches standard process tubing for draining in anyorientation Fast product change-over with self-draining design and no profileeffects Single flow path is easy to clean mechanically Highly-polished surface finish for ultra-pure fluidsSuperior reliability No moving parts to wear or replace minimizes maintenance forlong-term reliabilityMicro motion T-SeriesCoriolis Flow and Density MetersHigh performancecompact drainableCoriolis meterStraight tubefull-boreCoriolis meterHygienic compactdrainable CoriolismeterPeakperformanceCoriolis meterGeneral purposeflow-onlyCoriolis meterF-SeriesT-SeriesH-SeriesELITE R-SeriesExtreme low-flow CoriolismeterLF-SeriesPeak

2 Performancehigh capacitymeterELITE HC2T- series Straight-Tube Coriolis Flow and Density MetersMicro motion T-Series Straight-Tube flow and Density metersMicro motion Coriolis Meters meet a vast range of application needs, ranging from extreme low-flow up tohigh-flow, high-capacity lines. Cryogenic, hygienic, high-temperature, and high-pressure micro Motionmeters can handle them all. micro motion Meters are available with a variety of wetted parts to ensure thebest material Meters offer dramaticbenefits over traditional volumetric measurementtechnologies. Coriolis Meters : Deliver accurate and repeatable process dataover a wide range of flow rates and processconditions. Provide direct inline measurement of mass flowand Density , and also measure volume flow andtemperature all from a single device.

3 Have no moving parts, so maintenance costsare minimal. Have no requirements for flow conditioning orstraight pipe runs, so installation is simplifiedand less expensive. Provide advanced diagnostic tools for both themeter and the Coriolis Straight-Tube meter design is based on the ASMEB ioprocessing Equipment Standard. Withoptional sanitary fittings, micro MotionT- series Meters meet 3-A SanitaryStandards for Milk and Milk Products, areEHEDG clean-in-place approved, andfeature a standard surface finish of32 -inch Ra ( -meter) and15 -inch Ra ( -meter) is an available micro motion T-Series single straight -tubedesign makes these Meters self-draining, andallows them to be cleaned or sterilized in place(CIP/SIP).

4 The straight flow path also resistsplugging, and can be performance (liquid only)..4 Temperature approvals and of Straight-Tube Coriolis Flow and Density Meters3 Flow performanceMassVolume(1)(1) Specifications for volumetric flow rate are based on a process-fluid Density of 1 g/cc (1000 kg/m3). For fluids with Density other than1 g/cc (1000 kg/m3), the volumetric flow rate equals the maximum mass flow rate divided by the fluid s flow rateT025256803680T0501403800173800T07550 014,0006214,000T100110030,00013230,000T1 50320087,00038387,000 Mass flow accuracy(2)(2) Flow accuracy includes the combined effects of repeatability, linearity, and hysteresis. All specifications for liquids are based onreference conditions of water at 68 to 77 F (20 to 25 C) and 15 to 30 psig (1 to 2 bar), unless otherwise noted.

5 Of rate(3)(3) When flow rate < (zero stability / ), then mass flow accuracy = [(zero stability / flow rate) 100]% of rate andrepeatability = [ (zero stability / flow rate) 100]% of flow accuracy(2) of rate(4)(4) When flow rate < (zero stability / ), then volume flow accuracy = [ (zero stability / flow rate) 100]% of rate, andrepeatability = [ (zero stability / flow rate) 100]% of flow accuracy(2) of rate(5)(5) When flow rate < (zero stability / ), then gas flow accuracy = [(zero stability / flow rate) 100]% of rate, andrepeatability = [ (zero stability / flow rate) 100]% of of ratelb/minkg/hgal/minl/hZero Straight-Tube Coriolis Flow and Density MetersFlow performancecontinuedDensity performance (liquid only)Typical accuracy, turndown, and pressure dropPressure drop is dependent on process conditions.

6 To determine accuracy, turndown, and pressure drop with your processvariables, use micro motion s product selector, available from maximumflow rate100:120:110:11:1 Accuracy ( %) drop~ ~ (1)(1) Density accuracy includes the combined effects of repeatability, linearity, and hysteresis. All specifications for liquids are based onreference conditions of water at 68 to 77 F (20 to 25 C) and 15 to 30 psig (1 to 2 bar), unless otherwise noted. g/cc kg/m3 Repeatability g/cc kg/m3 Range0 5000 :120:110:11:1 Maximum flow rate (%)Accuracy (%) T-Series Straight-Tube Coriolis Flow and Density Meters5 Temperature specificationsAccuracy C of reading in CRepeatability CTemperature limits(1) (2) (3)(1) Temperature limits may be further restricted by hazardous area approvals.

7 See page 12.(2) When ambient temperature is below 40 F ( 40 C), a core processor must be heated to bring its local ambient temperature tobetween 40 F ( 40 C) and +140 F (+60 C). Long-term storage of electronics at ambient temperatures below 40 F ( 40 C) is notrecommended.(3) For the purposes of selecting electronics options, this graph should be used only as a general guide. If your process conditions areclose to the gray areas, it may be inappropriate to use electronics options other than a junction box. Consult with your micro Motionrepresentative. 60 ( 51) 20 ( 29) 40 ( 40)20 (7)60 (16)100 (38)140 (60) 60 ( 51) 20 ( 29)20 (7)60 (16)100 (38)140 (60)200 (93)240 (116)280 (138)320 (160)300 (149) (52)122(50)Mount transmitter remotely; use junction boxBelow 40 F ( 40 C),mount transmitter remotely.

8 Use junction boxAmbient temperature of core processor or transmitter in F ( C)Maximum process temperature in F ( C)6T- series Straight-Tube Coriolis Flow and Density MetersPressure ratingsAll pressure ratings are based on ASME pressure/temperature rating with ASME F316/316L socket weld flangesPED complianceSensors comply with council directive 97/23/EC of 29 May 1997 on Pressure secondarycontainment rating(1)(1) Housing is not rated for pressure containment below 20 F ( 29 C).Models without purge fittings1450100 Models with purge fittings72550 Models T025T through T150T; Models T075F through T150F Temperature ( F)Pressure (psig) T-Series Straight-Tube Coriolis Flow and Density Meters7 Pressure ratingscontinuedSensor pressure/temperature rating with Tri-Clamp compatibleTi Grade 1/304L clad hygienic fittingsSensor pressure/temperature rating with VCO Ti Grade 1/304L clad fittingsModels T025T through T150T.

9 Models T025F through T150 FModels T025T and T050T Temperature ( F)Pressure (psig) Temperature ( F)Pressure (psig)8T- series Straight-Tube Coriolis Flow and Density MetersPressure ratingscontinuedSensor pressure/temperature rating with JIS 2220 F316/316L socket weld flangesSensor pressure/temperature rating with DIN 11851 Ti Grade 1/304L clad hygienic couplingModels T025T through T150T; Models T075F through T150 FModels T025T through T150T; Models T025F through T150F Temperature ( C)Pressure (bar) Temperature ( C)Pressure (bar) T-Series Straight-Tube Coriolis Flow and Density Meters9 Pressure ratingscontinuedSensor pressure/temperature rating with DIN 11864-1 Ti Grade 1/304L clad hygienic couplingSensor pressure/temperature rating with SMS 1145 316L hygienic couplingModels T025T through T150T.

10 Models T025F through T150 FModels T075T, T150T, T075F, and T150F Temperature ( C)Pressure (bar) Temperature ( C)Pressure (bar)10T- series Straight-Tube Coriolis Flow and Density MetersPressure ratingscontinuedSensor pressure/temperature rating with ISO 2853 (IDF) 316L process fittingsSensor pressure/temperature rating with EN1092-1 & DIN F316/316L socket weld flangesModels T075T, T150T, T075F, and T150 FModels T025T through T150T; Models T075F through T150F Temperature ( C)Pressure (bar) Temperature ( C)Pressure (bar) T-Series Straight-Tube Coriolis Flow and Density Meters11 Environmental effectsVibration limitsSanitary standardsProcess temperature effectProcess temperature effect is defined as the worst-case zero offset due to process fluidtemperature change away from the zeroing of maximum flow rate per CPressure effectPressure effect is defined as the change in sensor flow sensitivity due to process pressurechange away from the calibration pressure.