TECHNICAL MANUAL RESPIRATORY HUMIDIFIERS …

TECHNICAL MANUALRESPIRATORY HUMIDIFIERS MODELS:MR730MR720MR700MR480 Revision FIssued March 2001 Fisher & Paykel Healthcare LtdAuckland, New ZealandiiRef. 185040646 *MR730/720/700/480 TECHNICAL MANUAL Revision F Issued March 2001 National Office:Fisher & Paykel Healthcare Box 14348, PanmureAuckland, New ZealandTelephone:+64-9-574 0100 Facsimile:+64-9-574 0158 TECHNICAL MANUAL for RESPIRATORY HumidifierMR730 Dual ServoMR720 Dual Servo (anaesthesia)MR700 Dual ServoMR480 Single ServoFisher & Paykel Healthcare have a policy of continued product improvement and reserve the right to alter specificationswithout F changes,SectionDescription of Performance test shortened (Dynamic test removed). Compatability list history updatedAppendix BMaintenance Schedule Changed for Temperature ProbesAppendix DCleaning Procedures RewrittenAppendix EDynamic performance check made an optional extraiiiRef.

TECHNICAL MANUAL RESPIRATORY HUMIDIFIERS MODELS: MR730 MR720 MR700 MR480 Revision F Issued March 2001 Fisher & Paykel Healthcare Ltd Auckland, New Zealand

Tags:

  Manual, Technical, Respiratory, Humidifiers, Technical manual respiratory humidifiers

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of TECHNICAL MANUAL RESPIRATORY HUMIDIFIERS …

1 TECHNICAL MANUALRESPIRATORY HUMIDIFIERS MODELS:MR730MR720MR700MR480 Revision FIssued March 2001 Fisher & Paykel Healthcare LtdAuckland, New ZealandiiRef. 185040646 *MR730/720/700/480 TECHNICAL MANUAL Revision F Issued March 2001 National Office:Fisher & Paykel Healthcare Box 14348, PanmureAuckland, New ZealandTelephone:+64-9-574 0100 Facsimile:+64-9-574 0158 TECHNICAL MANUAL for RESPIRATORY HumidifierMR730 Dual ServoMR720 Dual Servo (anaesthesia)MR700 Dual ServoMR480 Single ServoFisher & Paykel Healthcare have a policy of continued product improvement and reserve the right to alter specificationswithout F changes,SectionDescription of Performance test shortened (Dynamic test removed). Compatability list history updatedAppendix BMaintenance Schedule Changed for Temperature ProbesAppendix DCleaning Procedures RewrittenAppendix EDynamic performance check made an optional extraiiiRef.

2 185040646 *MR730/720/700/480 TECHNICAL MANUAL Revision F Issued March SERIES GENERAL MR730/720 Electrical Specifications (All Models).. and of CONTROLS AND Temperature Control (Heater Wire On Mode Only).. Temp Button (MR700/720/730 only).. Wire Switch (MR720, MR730 only).. Up and Version Wire On Mode (All models except MR480).. in Heater Wire Off Mode and Up up mode exiting conditions (low temperature alarm enabled):.. (Version , Software).. (All software versions), MR720 (Version Only).. Standby Mode (Version and software only).. AND Probe In Heater Wire Mode (MR700/720/730 only).. Wire temperature High temperature Low Temp High Temp Low Alarm Low Alarms in Non Heater Wire Mode (includes MR480).

3 Wire Temperature High Temperature Low Protection (All Models).. Watch 185040646 *MR730/720/700/480 TECHNICAL MANUAL Revision F Issued March Analog Switches U8 and to Digital Converter U10 b,d, Control Line Interrupt (Revisions D/F PCBs).. Line Interrupt (Revision H PCB and later).. C D/F H AND J Temperature versus Thermistor Procedure for Models with Revision D PCBs or Procedure for Models with Revision F PCBs or Check MR730/720 Setup Notes:.. Control Check MR700/720 check MR700/720 Check (MR480).. Control Check (MR480).. check - Probe Temperature Circuit Board with screw-in Torque 185040646 *MR730/720/700/480 TECHNICAL MANUAL Revision F Issued March , PART LISTS AND MECHANICAL Series Circuit Layout Block Layout - Revision H and J Layout - Revision F Pad Locations for Programming Control Options - Revision H and J Diagrams - Revision H and J PCB Control Diagrams - Revision H and J PCB Power Diagrams - Revision F PCB Control Diagrams - Revision F PCB Power Parts Board Components: (Revision h and later PCBs).

4 Board Components: (Revision h and later PCBs).. Parts Mechanical Diagrams: Mechanical Diagrams: MR730/720 Assembly Parts (MR730/720/700/480).. Diagrams:Heaterplate Assembly MR730/720/700/480 (Inverted View).. Identification PCB Harness Connection CHANGE Number Number display functions: Heater Wire Functions: Non Heater Wire Control Series Humidifier Serial Data Data Data Interface Or In Accordance With Relevant Country Schedule Guide For The MR700 Series RESPIRATORY 185040646 *MR730/720/700/480 TECHNICAL MANUAL Revision F Issued March Water In circuit water Series Fault Flow Chart For No Go of Humidifier Heater of Airway Temperature Dynamic Performance Check of Check MR730/720 Test Sequence (MR730/720/700).

5 Check (MR480).. Test Procedure (MR480)..581 Ref. 185040646 MR730/720/700/480 TECHNICAL MANUAL Revision F Issued March SERIES GENERAL SPECIFICATIONS_____This documentation defines the TECHNICAL specifications of the MR730, MR720, MR700 and MR480 respiratoryhumidifiers, it also includes maintenance and repair MR730/720 models are Heated Wire HUMIDIFIERS with the option of disabling the Heater MR720 has a lower alarm limit and an extended warmup mode for use in MR700 model is exclusively a Heated Wire MR480 model is a non Heated Wire MR730/720/700 Supply Voltage230 25 V127 12 V115 V100 10 VSupply Frequency50 or 60 HzSupply A maximum at 230 A maximum at 127 A maximum at 115 A maximum at 100 VHeater Wire Capacity60 MR480 Supply Voltage.

6 230 25 V115 V100 10 VSupply Frequency50 or 60 HzSupply A maximum at 230 A maximum at 115 A maximum at 100 ELECTRICAL SPECIFICATIONS (ALL MODELS)Heaterplate Power150 WHeaterplate Overheat ProtectorOperates at 118 6 AND APPROVALSC lassification , UL2601 IEC601-1, , EN60601-1 Class 1 Type BDrip ProofContinuous OperationNot to be used in the presence of flammable CONTROLPID (Proportional Integral Derivative) control of Heater Wire (in Heated Wire mode operation only) and SENSINGT emperature measurement using standard characteristic thermistors (YSI 400 compatible) mounted in fast responsetemperature probes, located at the patient end of the delivery tube and at the humidifying chamber 185040646 MR730/720/700/480 TECHNICAL MANUAL Revision F Issued March DISPLAYT hree digit 14mm 7 segment LED (Light Emitting Diode) to C.

7 Below C displays Lo . Above C displays Hi .Accuracy C to OF ABBREVIATIONSEOHThe End Of Hose temperature difference of the TOP from the temperature Set potentiometer. Sets the controlling temperature of the temperature at the Top of the humidification Shut 185040646 MR730/720/700/480 TECHNICAL MANUAL Revision F Issued March CONTROLS AND TEMPERATURE CONTROLS elects the airway controlling temperature. The range of adjustment is from 31 to 40 C over a linear scale marked indegrees on the front temperature can be set with a resolution of C. The set temperature is flashed on the display when the controlknob is rotated and continues to be flashed for five seconds after the last CONTROL (HEATER WIRE ON MODE ONLY)The temperature of the gas leaving the chamber is regulated by the Chamber Control knob, it sets the difference betweenthe chamber outlet temperature and the airway temperature delivered to the patient.

8 The setting of the Chamber Control,when added to the setting of the Temperature Control, determines the chamber outlet set temperature. This ChamberControl has a linear range of adjustment from -5 to +2 C marked in one degree steps and can be set with a resolution C. The value is flashed on the Airway display as the control knob is rotated and continues to be flashed for fiveseconds after the last humidifier is also supplied with a fixed chamber control knob. This prevents inadvertent adjustment of thechamber control by preadjusting the existing knob to the desired setting and then replacing the normal knob with thefixed BUTTONA momentary push will disable an audible alarm for 3 minutes (except low temperature alarm in non heater wire modewhich is 10 minutes).

9 A secondary function is that in conjunction with the Chamber Temp switch enables entry to the engineering menu (referAppendix A2).On every initial push of the mute button a momentary audio beep indicates function TEMP BUTTONMR700 Series HUMIDIFIERS with Version Software:Operation of this switch causes the airway temperature display to momentarily blank then show the chamber outlettemperature. A green LED adjacent to the switch is enabled to indicate that this function is active. On release of thisswitch the display reverts immediately to show the airway button is used in conjunction with the mute to enter the engineering menu (refer Appendix A2).MR700 Series HUMIDIFIERS with Version and Version Software:Operation of this switch causes the airway temperature display to momentarily blank then show the chamber outlettemperature.

10 A green LED adjacent to the switch is enabled to indicate that this function is active. On release of thisswitch the display continues to show the chamber outlet temperature for 2 seconds before blanking momentarily andreverting to show the airway button is used in conjunction with the mute to enter the engineering menu (refer Appendix A2).4 Ref. 185040646 MR730/720/700/480 TECHNICAL MANUAL Revision F Issued March BUTTON (MR700/720/730 ONLY)This button has to be depressed for one second to enter or exit the standby mode. When the adjacent amber LED startsto flash (100/minute) the button can then be released and the Humidifier remains in the standby exit Standby mode press the Standby button for 1 second, the amber LED will then turn off.