Pag 1 of 56 REPORT REPORTIEC EN 61010-1 Safety requirements for electrical equipment for measurement, control, andlaboratory usePart 1: General requirementsReport Reference No.. :55790 TRFSAFT ested byGuido Roman Approved byRoberto TarroDate of :2005-12-19 Contents ..:56 PagesTesting : (+39) 39 (+39) 39 :Via Trento e Trieste ,116 20046 Biassono (MI) - ItalyTesting location/procedure ..:CBTL [ ] SMT [ ]TMP[ ] :Via Trento e Trieste ,116 20046 Biassono (MI) - ItalyApplicant s :Via F. Rovereto, 9/1120014 S.

pag 8 of 56 Report No. 55790TRFSAF TABLE 3: - List of components and circuits relied on for safety Unique component reference or location Application /

Tags:

  Requirements, Safety, Safety requirements

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of TEST REPORT IEC EN 61010-1 Safety requirements …

1 Pag 1 of 56 REPORT REPORTIEC EN 61010-1 Safety requirements for electrical equipment for measurement, control, andlaboratory usePart 1: General requirementsReport Reference No.. :55790 TRFSAFT ested byGuido Roman Approved byRoberto TarroDate of :2005-12-19 Contents ..:56 PagesTesting : (+39) 39 (+39) 39 :Via Trento e Trieste ,116 20046 Biassono (MI) - ItalyTesting location/procedure ..:CBTL [ ] SMT [ ]TMP[ ] :Via Trento e Trieste ,116 20046 Biassono (MI) - ItalyApplicant s :Via F. Rovereto, 9/1120014 S.

2 Ilario di Nerviano (Mi) - ItalyTest specification:Standard ..:IEC/EN 61010 1 : 2001 (2nd Edition)Test :CENon-standard test method ..:Test REPORT Form No..:IECEN61010_1 CTRF Originator ..:Master TRF ..:Copyright 2001 IEC System for Conformity Testing and Certification of Electrical Equipment (IECEE), Geneva,Switzerland. All rights publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged as copyright owner andsource of the material. IECEE takes no responsibility for and will not assume liability for damages resulting from the reader's interpretation of thereproduced material due to its placement and item :ON-OFF LEVEL SWITCHEST rademark.

3 (s) ..:Rated voltage 300 Vdc maxor 300 Vac 50Hz max 80W / VApag 2 of 56 REPORT according to additional information: National requirements : Other requirements : Environmental conditions:Operating temperature range -30 to +55 C;Relative Humidity: 80% maxStorage temperature: -40 +65 CPollution Degree 2 Operating conditions:ContinuousConnection to supply mains:Permanent with screw terminal block solder on thePWBC able cross section: see instruction sheetDegree of mobility:Fixed equipmentOverall size of the equipment: width; depth.

4 Height:See instruction sheetMass of the equipment:1 kgSpecial protection to IEC 60529IP 65 Accessories and detachable parts included in theevaluation:Options included:noneGeneral remarks:"(See remark #)" refers to a remark appended to the REPORT ."(See appended table)" refers to a table appended to the REPORT ."(See Annex_#)" refers to an annex appended to the this REPORT a point is used as the decimal test results presented in this REPORT relate only to the object REPORT shall not be reproduced except in full without the written approval of the testing test case verdicts:P = Pass, F = Fail, N = Not applicable.

5 Placed in the column to the right (Verdict)Copyright 2001 IEC System for Conformity Testing and Certification of Electrical Equipment (IECEE),Geneva, Switzerland. All rights publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged ascopyright owner and source of the material. IECEE takes no responsibility for and will not assume liability for damages resulting fromthe reader's interpretation of the reproduced material due to its placement and 3 of 56 REPORT of marking plate not in scale:pag 4 of 56 REPORT INFORMATIONpag 5 of 56 REPORT test REPORT cover also the following modelsS1 ; S1 ; S1 ; S1 ; S1 ; S1 ; S1 ;S1 ; S1 ; S1 ; S1 ; S1 ; S1 ; S1 ;S1 ; L1 ; L1 ; L1 ; L1 ; L1 ; L1 ; L1 ;L1 ; L1 ; L1 ; L1 ; L1 ; L1 ; L1.

6 L1 ABOUT OTHER STANDARDS / DOCUMENTS CONSIDERED:TESTED ACCORDING TO NATIONAL requirements FOR THE FOLLOWINGCOUNTRIES:All CENELEC countriesLIST OF APPENDIXES / ENCLOSURES TO THE TEST REPORT :APPENDIX 1 : PHOTOGRAPHSpag 6 of 56 REPORT : 2 - Test equipment listUsedTypeEquipmentCalibration NumberXVariac 0-600 VacBellotti 225515 XDigital power meterYokogawa 253107/200507/200625AW0064 Digital power meterVoltech PM3000A07/200508/20066872-0021 XDigital power meterEDC 3188 MT07/200508/2006009 XStabiliser3 RTE XOscilloscopeYokogawa DL 154002/200502/200625WY-1600 LXTemperature recorderFluke Hydra 262507/200507/20065506661 XThermocouples type KNemko testerAOIP OM2107/200507/200691124221 XMultimeterPhilips PM253407/200508/2006DY0101292 XMultimeterFluke

7 7507/200508/200652230540 XDielectric strength testElektrotech lab. UH2707/200507/200600246 XLeakage TestNemko07/200508/200600815 XEarth Continuity TesterEDC 9170B07/200507/200600220 XElectronic loads 3 phaseBelotti D 25843 00645 XOvenCarbolite Tersid NR 60F07/200506/200721648 XBall pressure Test apparatusATS 00912 XHumidity chamberVotsch VC715007/200507/2006 XMicrometerMitutoyo 103-13702/200502/2006 XTorque screwdriverTorqueleader 6F502597602/200502/2007 SEQ100 XTorque screwdriverTorqueleader 6K50495902/200502/2007 SEQ101 XInsulation transformerNemko XImpact hammerPTL ( )

8 02/200502/20069612478 XMass Ball XCalliperMitutoyo CD 15 DC02/200502/20071253866 XForce gaugeAmetek Accuforce III XRigidity test (12mm)ATS02/200502/200800503 XAccessibility (finger pin)ATS02/200502/200800542 Inclined planesATS 00519pag 7 of 56 REPORT recorder one chanel Tersid 70507/200507/2006014-ELD-MDielectric strength testHipotronics HD100 XImpulse test +Coupler networkSchaffner NSG651 Schaffner CD11604/12/200504/12/2006 xData LoggerTesto 175-H207/200507/200620013013DC Power SupplyPHILIPS PE1510 1) or interval between 8 of 56 REPORT 3.

9 - List of components and circuits relied on for safetyUnique componentreference orlocationApplication /FunctionManufacturer(NOTE 1)Part numberRATING(NOTE 2)Evidenceofacceptance(NOTE 3)PCBW iring printed -ComponentAGM SrlZPMV2FR-4 single layer94V-0T=110 CUL(E161090)Terminal blockplasticPlasticsComponentWIECONType 8191/8291300V, 16 AVDE,S+,D,UR,CSAI nternal wireInternal cable (forresistancesensor)Cavi HT300/500 VacT=105 Csection 20-20 CEI 20-11 NOTE 1 - List all manufacturers 2 - Electrical, mechanical, flammability, 3 - Licence number, file number or other documentary evidence of acceptancepag 9 of 56 REPORT + TestResult RemarkVerdict5 MARKING AND DOCUMENTATION Required equipment markings are: visible:PFrom the exterior; orPAfter removing a cover.

10 OrNOpening a doorNAfter removal from a rack or panelNNot put on parts which can be removed by anOPERATORNL etter symbols (IEC 60027) usedPGraphic symbols (IEC 61010-1 : Table 1) Equipment is identified by: )Manufacturer s or supplier s name or )Model number, name or other location supplySee page 1 Equipment is marked as follows: )Nature of supply:AC or DC 1) RATED mains frequency or range offrequencies:50 HzP2) with symbol )RATED supply voltage(s) or range:See page )Max. RATED power (W or VA)or input current:PThe measured value not more than 110 %(see Form )PIf more than one voltage range: Separate values marked.