user s Manual 8201052 IDigora OptimeUser s Manualfor theDigora Optime Standard versionand theDigora Optime Multiconnect versionThis unit is marked according to the Medical Device Directive93/42/EEC (if the unit contains the CE mark) user s Manual for theOptimeDocument number 8201052 rev. 2 (0509)Original approved English language versionManufactured by BOX 148 FIN-04301 TUUSULA, FINLANDTel. +358 10 394 820 Fax. + 358 9 701 5261 IIUser s Manual 8201052 Digora OptimeSoredex endeavours to produce product documentation that is accurate and upto date. However, our policy of continual product development may result inchanges to products that are not reflected in the product documentation. There-fore, this document should not be regarded as an infallible guide to currentproduct specifications.

II User’s manual 8201052 Digora Optime Soredex endeavours to produce product documentation that is accurate and up to date. However, our policy of continual product development may result in

Tags:

  Manual, User, User s manual, Soredex

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of User’s Manual - Dental Concepts

1 user s Manual 8201052 IDigora OptimeUser s Manualfor theDigora Optime Standard versionand theDigora Optime Multiconnect versionThis unit is marked according to the Medical Device Directive93/42/EEC (if the unit contains the CE mark) user s Manual for theOptimeDocument number 8201052 rev. 2 (0509)Original approved English language versionManufactured by BOX 148 FIN-04301 TUUSULA, FINLANDTel. +358 10 394 820 Fax. + 358 9 701 5261 IIUser s Manual 8201052 Digora OptimeSoredex endeavours to produce product documentation that is accurate and upto date. However, our policy of continual product development may result inchanges to products that are not reflected in the product documentation. There-fore, this document should not be regarded as an infallible guide to currentproduct specifications.

2 soredex maintains the right to make changes andalterations without prior s Manual 8201052 IIID igora OptimeContents1. Introduction .. The Warnings and Handling Imaging Positioning the unit .. 32. Optime - Standard Main parts and controls .. Indicator Using the 7 Preparing the image read-out system .. 7 Preparing the imaging plate .. 8 Taking an exposure using the imaging 9 Reading the imaging plate .. 10 Retrieve last 14 Image quality warning 15 Removing imaging plates from the plate 16 Shutting down the 16 Erasing an Imaging Plate .. 173. Optime - Multi-Connect version .. Main parts and controls .. Indicator Using the 22 Preparing the image read-out system .. 22 Preparing the imaging plate.

3 23 Taking an exposure using the imaging 24 Reading the imaging plate .. 25 Retrieve last 29 Image quality warning 30 IVUser s Manual 8201052 Digora OptimeRemoving imaging plates from the plate 30 Shutting down the 31 Erasing an Imaging Plate .. 314. Setup Tooth 34 Selecting the tooth numbering 34 Using the tooth numbering Image 355. Cleaning and disinfecting the unit .. Cleaning the unit .. Disinfecting 366. Maintenance, repair and disposal .. 377. Symbols that appear on the unit .. 388. Technical OpTime Unit .. Imaging Computer Main 42 user s Manual 82010521 Introduction Digora Optime1. The OptimeThis Manual describes how to operate the CranexOptime intraoral imaging plate unit is designed to read 0, 1, 2, and 3 sizeintraoral imaging plates supplied by unit works in conjunction with a PC in whichcompatible Dental imaging software, such as Digorafor Windows or above, has been Optime can be configured to work with a singlePC, the Standard version or in a shared environmentwhere the unit is accessed from different PCs.

4 TheMulti-Connect the manufacturer s authorized personnel areallowed to install and configure the Optime intraoralimaging plate s Manual 8201052 Digora Optime Warnings and precautionsCAUTION - CLASS 3B LASER RADIATION WHEN OPENAVOID EXPOSURE TO BEAMCAUTION - Use of controls or adjustments or performance of procedures other than those specifiedherein may result in hazardous laser radiation exposure When handling imaging plates, plate covers and protective bags always take the appropriatehygiene measures and precautions to prevent cross contamination. The imaging plates are harmful if swallowed. Do not move or knock the unit when it is reading an imaging plate. The use of control or adjustment or performance of procedures other than those specified hereinmay result in hazardous laser radiation exposure.

5 This unit must only be used to read soredex OPTIME image plates and must not be used forany other purpose. NEVER use imaging plates or protective covers from other manufacturers. This unit, or its accessories, must not be modified, altered or remanufactured in any way. Only the manufacturer s authorized service personnel are authorized to carry out maintenanceand repair. There are no user serviceable parts inside the unit. Infection control procedures must be observed when using accessories, such as film holders, x-ray tube guides and imaging plates. When using accessories always follow the manufacturer sinstructions on how to use the accessory and prevent cross contamination from one patient toanother. This device can interfere with other devices due to its EMC characteristics.

6 Other devices can interfere with this device due to their EMC characteristics. This device complies with IEC 60601-1 standard. Accessory equipment connected to this devicemust be in compliance with the related nationally harmonized IEC standards. Equipment not suitable for use in the presence of flammable anaesthetic mixture with air or withoxygen or nitrous oxide. Only use the SKYNET SNP-PA59-M power supply unit that is supplied with theunit. Do not use any other power supply units with the unit. If this device will be used with 3rd party imaging application software not supplied by soredex ,the 3rd party imaging application software must comply with all local laws on patient informationsoftware. This includes, for example, the Medical Device Directive 93/42/EEC and/or FDA s Manual 82010523 Introduction Digora Handling Imaging PlatesAfter exposing an imaging plate and removing theimaging plate and cover from its protective bag, doNOT expose the imaging plate and cover to verybright light before reading the imaging read the imaging plate as quickly as possibleafter it has been not leave read or unread imaging plates in theirprotective bags for long periods of handle the imaging plates with care and avoidsituations that could scratch imaging plates are not being used.

7 Store themin their protective box below 33 Positioning the unitDo NOT position the unit so that the door is in or nearvery bright s Manual 8201052 Digora Optime 2. Standard Version2. Optime - Standard Main parts and controlsUser s Manual 820105252. Standard Version Digora Indicator lightsThe different statuses of the indicator lights are de-scribed light onUnit is shut down (OFF).Press stand-by key to switch unit onUnit ready to read an imaging imaging slow flashingUnit in the energy save the Start key to enter ready fast flashingUnit is reading/erasing an imaging onDfW not opened or untransferred image in DfW and the image will automatically slow flashingThere is no Ethernet fast flashingUnit carrying out its s Manual 8201052 Digora Optime 2.

8 Standard VersionGREEN on, YELLOW onUnit is in the Erase flashing, YELLOW onUnit is in the Erase mode and is erasing an imagingplateGREEN on, YELLOW flashingUnit is entering the energy save modeRED flashingERROR. Switch the unit off and then on again. If thered light comes on again, contact your dealer for s Manual 820105272. Standard Version Digora Using the unitPreparing the image read-out system1. Switch on the computer that is connected to Open Digora for Windows or the Dental imagingsoftware you are If you are using the unit for the first time you maywish to check and/or change the setup section 4 Setup options for more Open the patient card where you wish to place thenew image, or open a new patient Press the Stand-by key to switch the unit on.

9 TheYELLOW stand-by light will turn off and unit willcarry out a selftest during which the YELLOW warning light will flash. The selftest will last about30 the plate carrier slides out of the unit doorand the GREEN light comes on the unit is ready the GREEN light does not come on it indicatesthat the unit is not connected to the the RED light comes on contact your dealer s Manual 8201052 Digora Optime 2. Standard VersionPreparing the imaging plateIMPORTANT NOTEIf the imaging plate is being used for the very first timeor if it has not been used within the last five days, itMUST be erased. See section 3. Erasing an imagingplate, for information on how to do handling imaging plates, plate covers andprotective bags always take the appropriate hygienemeasures and precautions to prevent cross Place the imaging plate you plan to use onto thelong half of the plate cover.

10 The blue (active) sideof the imaging plate must face the plate cover andthe metal disk, on the back of the imaging plate,must be positioned near the outer edge of theplate Fold the short half of the plate cover over onto theimaging Slide the plate cover and imaging plate into theprotective bag as far as it will go. Make sure thatthe black side of the imaging plate, the back, isagainst the tape (black) side of the s Manual 820105292. Standard Version Digora Optime4. Remove the cover paper from the sealing the flap over onto the sealing tape. A scoreline on the front of the bag will help you to fold the flap from the centre out to ensure thatthe seal does not an exposure using the imaging plate1. Place the sealed protective bag in the patient smouth in the required position for the image youwish to take.