12950 W. Eight Mile Road Oak Park, MI 48237-3288 248/399-8000 FAX: 248/399-7246 Thermometers15 DefinitionA thermometer is an instrument designed to measure and indicate thetemperature of a specific application or condition. A Dial thermometer (filledsystem thermometer ) can either be read at the point of measurement orfrom a remote location using a desired length of capillary of OperationTrerice Dial Thermometers operate using a filled thermal system. Thissystem consists of capillary tubing and a sensing bulb, which are filled withan expandable chemical compound. The fill is contained within the sealedthermal system, and is affected (expands or contracts) by temperaturechanges at the sensing bulb. As temperature increases, expanding filltravels via the capillary tube system to the bourdon tube within theinstrument s case. The expansion causes the bourdon tube to flex and theresulting motion is transmitted as a temperature measurement through amechanical movement to the pointer and is a Dial thermometer ?

A Dial Thermometer (filled system thermometer) can either be read at the point of measurement or from a remote location using a desired length of capillary tubing.

Tags:

  What, Dali, Thermometer, What is a dial thermometer

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of What is a Dial Thermometer? - Control Products Inc

1 12950 W. Eight Mile Road Oak Park, MI 48237-3288 248/399-8000 FAX: 248/399-7246 Thermometers15 DefinitionA thermometer is an instrument designed to measure and indicate thetemperature of a specific application or condition. A Dial thermometer (filledsystem thermometer ) can either be read at the point of measurement orfrom a remote location using a desired length of capillary of OperationTrerice Dial Thermometers operate using a filled thermal system. Thissystem consists of capillary tubing and a sensing bulb, which are filled withan expandable chemical compound. The fill is contained within the sealedthermal system, and is affected (expands or contracts) by temperaturechanges at the sensing bulb. As temperature increases, expanding filltravels via the capillary tube system to the bourdon tube within theinstrument s case. The expansion causes the bourdon tube to flex and theresulting motion is transmitted as a temperature measurement through amechanical movement to the pointer and is a Dial thermometer ?

2 Continued next pageCASEPOINTERMOVEMENTBOURDON TUBE O RING GASKETADJUSTABLE UNION TYPESENSING BULB ASSEMBLYSENSING BULBWINDOWCAPILLARY TUBINGDIALFACE12950 W. Eight Mile Road Oak Park, MI 48237-3288 248/399-8000 FAX: 248/399-7246 ThermometersSelecting a Dial ThermometerThermal System ActuationTrerice Dial Thermometers are available in one of three system actuationfills: Vapor, Liquid, and Mercury. Vapor ActuationTrerice Vapor Actuated Dial Thermometers are the industry standard andare noted for their economical cost and excellent speed of response. Thephysical principles of vapor actuation require that thedialface be printed with a nonlinear, progressivelygraduated temperature scale. These instruments areavailable for direct mounting, or for remote mountingwith capillary lengths up to 100 feet. Sensing bulblength is dependent upon the capillary length selected(a longer capillary length will require a longer sensingbulb length).

3 Vapor Dial Thermometers are available intemperature ranges up to 450 F (232 C). Note: Erratic performancemay be encountered if the measured process temperature rapidlycrosses ambient ActuationTrerice Liquid Actuated Dial Thermometers have a goodresponse time and are furnished with a temperature scaleof linear graduation. These instruments are available fordirect mounting, or for remote mounting with capillarylengths up to 20 feet. Sensing bulb length is consistentand not affected by capillary length or temperature Dial Thermometers are available in temperatureranges up to 300 F (150 C) and are ideally suited formeasuring process temperatures which routinely crossambient. However, care should be taken to insure againstthe exposure of the capillary to temperatures above orbelow the factory calibration temperature of 75 F (24 C).Mercury ActuationTrerice Mercury Actuated Dial Thermometers providean average speed of response and are designed for hightemperature applications (up to 1000 F) or thoserequiring extreme accuracy (optional accuracy of 1/2scale division).

4 These instruments are available for directmounting, or for remote mounting with capillary lengthsup to 100 feet. Capillary lengths over 50 feet require aspecial temperature compensated capillary, whichreduces any ambient temperature error. Sensing bulblength is dependent upon the temperature range spanAll Trerice Filled System Dial Thermometers should be carefully selected to meet the demands of the particular application. Theinformation contained in this catalog is offered only as a guide to assist in making the proper selection. Improper application maycause failure of the instrument, resulting in possible personal injury or property damage. For applications where the process mediamay be corrosive or contained under pressure, the use of a thermowell is required to prevent damage to the thermometer andfacilitate its removal from the : Vapor Dial Thermometers should be installed with thecase, capillary tubing, and sensing bulb located at a similarelevation to avoid measurement inaccuracies.

5 If the sensing bulbmust be installed at a different elevation than the case, pleaseadvise the factory when ordering so that the instrument can becalibrated : Temperature indication error will be introducedwhenever the capillary tubing is exposed to ambienttemperatures above or below 75 F. The following formula MUSTbe considered when specifying liquid actuation:Where:S = thermometer range span in FL = capillary length in feetT = capillary temperature variation from 75 FError = xS xL xTExample:S = 210 (30 to 240 F)L = 20T = 10 (85 F)Error = x210 x20 x10 = CAUTION: Temperature indication error will be introducedwhenever NONCOMPENSATED capillary tubing is exposed toambient temperatures above or below 75 F. The followingformula MUST be considered when specifying liquid actuationwithout compensated capillary tubing:Where:S = thermometer range span in FL = capillary length in feetT = capillary temperature variation from 75 FError = xS xL xTExample:S = 210 (30 to 240 F)L = 20T = 10 (85 F)Error = x210 x20 x10 = continued next pageWhat is a Dial thermometer ?

6 Continued12950 W. Eight Mile Road Oak Park, MI 48237-3288 248/399-8000 FAX: 248/399-7246 Thermometers17selected (a narrower range span will require a longer sensing bulb length).Mercury Dial Thermometers are furnished with a temperature scale of lineargraduation. Thermal System Actuation ComparisonConsiderationVapor ActuationLiquid ActuationMercury ActuationPriceEconomicalMidrangePremiumR esponse TimeExcellentGoodAverageDialfaceNon-line arLinearLinearMaximum Temperature Range450 F (232 C)300 F (149 C)1000 F (538 C)Cross Ambient ApplicationsNot recommendedRecommendedRecommendedAvailab le Thermal SystemsAll (except averaging)AllStainless steel only(averaging not available)Bulb SizeDependent on capillary lengthConsistentDependent on temperature rangeMaximum Capillary Length100 feet20 feet100 feet (compensated capillaryrequired over 50 feet)Accuracy One scale division One scale division One scale division standard 1/2scale division optionalCaseCases are made from stainless steel or cast aluminum, in sizes from 31/2"through 81/2".

7 Trerice Dial Thermometers can be directly mounted, orremotely mounted using capillary tubing, so that the measurement can beread from a convenient viewing location. Direct mounted thermometers areavailable with adjustable angle or universal angle connections, while remotemounted thermometers can be ordered for almost any surface or panelmounting and RingThe window is normally held in place by a ring or snapped directly to thecase of the thermometer . Plastic and clear glass are typical windowmaterials. Ring styles include threaded, friction and hinged, dependingupon the casetype accuracy of a dial thermometer is expressed as a variance (plus orminus) in scale divisions. All Trerice Dial Thermometers are accurate towithin one scale division of the measured range. Mercury Actuated DialThermometers can be ordered with an optional accuracy of 1/2of onescale division at additional cost.

8 Ambient temperature conditions andelevation variances may effect measurement next pageWhat is a Dial Thermometer? continued12950 W. Eight Mile Road Oak Park, MI 48237-3288 248/399-8000 FAX: 248/399-7246 ThermometersMeasurement Range and DialTrerice Dial Thermometers are available in Fahrenheit, Celsius, and DualScale temperature ranges from -40 F through 1000 F. Ranges are indeliblypresented in black figures and markings upon a white finished aluminumdialface. The physical principles of vapor actuation require the dialface tohave a non-linear, progressively graduated temperature scale; therefore,the temperature range should be selected so that the intended measuringpoint falls within the upper two-thirds of the range scale. Both LiquidActuated and Mercury Actuated Dial Thermometers have temperaturescales of linear graduation, thus, the intended measuring point should fallwithin the middle third of the range System Bulb Trerice Dial Thermometers are furnished with copper, brass orstainless steel sensing bulbs, depending upon the system actuationand the requirements of the application.

9 A fixed union connection isstandard, with an adjustable union connection optionally fixed union connection is furnished with a standard sensing bulblength of 13/4" to 55/8", depending upon the actuation, capillary length,and temperature range specified. The adjustable union connectionmay be adjusted over a 24-inch length prior to initial insertion. Thisallows the sensing bulb to be installed at any desired insertion length(U-length). Plain and Teflon covered bulbs are available for open tankapplications. Other bulb styles, including averaging and air-sensing,can be furnished on some models. Please see the Dial ThermometerSensing Bulb Section for complete bulb specifications. Capillary Trerice Dial Thermometers can be specified with variouscapillary materials and special covers to meet the requirements ofany application. Mercury Actuated Thermometers are supplied onlywith stainless steel tubing and can be specified with self-compensating capillary tubing (required on thermal system over 50feet), which will reduce any ambient temperature applications where the process media may be corrosive or containedunder pressure, the use of a Trerice Thermowell is required to preventdamage to the thermometer and facilitate its removal from the are available in various lengths, connections, sizes, andmaterials.

10 Please consult the Thermowell Section of this next pageTo ensure minimum responsetime, Trerice Heat Transfer Pasteshould be applied to the sensingportion of the bulb beforeinstallation into a oz. tube: Item No. 107-0001 what is a Dial Thermometer? continu