Example: tourism industry

www.madur

Madur Polska sp. z CHARACTERISTICFEATURESTECHNICAL is a full size CEMS system equipped with up to 6 electrochemical sensors (more cells on demand), that are backed up with up to 3 NDIR is divided into 3 modules: the analyser, the conditioner and the power supply three are mounted in a server-type cabinet with a cart to move it data logger with SD card allows to collect measurement results for weeks time. continuous measuring madur Polska sp. z TECHNICAL DATASENSORSEQUIPMENTAPPEARANCE CMS-7 is a modular CEMS systemIt consists of three modules in 19 rack standard:Conditioner module - height 5 UAnalyser module - height 4 UPower supply module - height 2 UAll above are mounted in 19 cabinet of 12 U heightCabinet s front door is glass to protect steering panels, yet allows to view and control the work of the CEMSD river and steering of heated hose and heated filterGas dryer (Nafion or condensation type)Two fibreglass filters installed in the front panelIn the back door, at the bottom, there is a main panel of electric and gas connectors, and at the side a connection of the heated hoseBack door is hinged, what allows to easily access the gas and electrical connections of all analyser s modulesCONDITIONER INCLUDESFlow indicator installed in the front panel

sales@madur.com CMS7-EN-v01.00 madur Polska sp. z o.o. TECHNICAL DATA SENSORSEQUIPMENTAPPEARANCE Drying method Cooler temperature Under pressure in Nafion® collar

Tags:

  Pressure

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of www.madur

1 Madur Polska sp. z CHARACTERISTICFEATURESTECHNICAL is a full size CEMS system equipped with up to 6 electrochemical sensors (more cells on demand), that are backed up with up to 3 NDIR is divided into 3 modules: the analyser, the conditioner and the power supply three are mounted in a server-type cabinet with a cart to move it data logger with SD card allows to collect measurement results for weeks time. continuous measuring madur Polska sp. z TECHNICAL DATASENSORSEQUIPMENTAPPEARANCE CMS-7 is a modular CEMS systemIt consists of three modules in 19 rack standard:Conditioner module - height 5 UAnalyser module - height 4 UPower supply module - height 2 UAll above are mounted in 19 cabinet of 12 U heightCabinet s front door is glass to protect steering panels, yet allows to view and control the work of the CEMSD river and steering of heated hose and heated filterGas dryer (Nafion or condensation type)Two fibreglass filters installed in the front panelIn the back door, at the bottom, there is a main panel of electric and gas connectors, and at the side a connection of the heated hoseBack door is hinged, what allows to easily access the gas and electrical connections of all analyser s modulesCONDITIONER INCLUDESFlow indicator installed in the front panelGas pump (diaphragm) and condensate pump (peristaltic)Solenoid valve for ventilationParamagnetic oxygen sensor (if it is installed)

2 ANALYSER MODULE INCLUDESSet of electrochemical cells (O2 + up to 5 toxic sensors)Set of NDIR sensors (up to 3)Other measurement systems (temperature measurements, pressure , flow, etc.)Large (320*240), graphical, monochromatic LCD display and steering keyboardCPU administrative computerAnalogue outputsCommunication interfaceData-logger with SD cardMAIN CONNECTION PANEL - LOCATED AT THE BOTTOMOF THE CABINET S BACK DOORVIEW AT THE BACK PANELS OF ALL CMS-7 ARE ACCESSIBLE AFTER OPENING THE CABINET S BACK madur Polska sp. z TECHNICAL DATASENSORSEQUIPMENTAPPEARANCEPOWER SUPPLY UNITGAS ANALYSER UNITGAS SAMPLE CONDITIONER UNITCARTHEATED HOSE CONNECTIONCART FOR CMS-7 madur Polska sp. z TECHNICAL DATASENSORSEQUIPMENTAPPEARANCED imensions (W * H * D)Casing materialOperating conditionsStoring temperaturePower supply: input | maximal power consumptionData memory: size | number of resultsDisplayGas pump | gas flow32 kB | 30 reports + 10 banks (1024 sets of data)Graphical LCD 320*240, with variable contrast and backlightingPurging pump for CO sensorCommunication interface with PC computerGas filteringRS-232 CDiaphragm, max 2l/min | 90l/h (1,5l/min)470 mm * 310 mm * 160 mmAluminium padded with foam and fabric (polyester)Diaphragm, max 1,5l/minWeight (without accessories) T: 10 C 50 C, RH: 5% 90% (non-condensing) 0 C 55 C12,0 kg 12,8 kgANALYSER UNIT115V AC or 230V AC | 90W (without heated hose)1.

3 Heated filter included in the heated hose2. Built-in final filter (behind the gas dryer) with replaceable insertAnalogue outputs (optional)Eight current (0/4mA 20mA) and eight voltage (0V 10V) outputs Drying methodCooler temperatureCooler temperature hysteresisMaximum gas flow for efficient dryingCondensate pumpHeated hose temperatureHeated hose temperature hysteresisHeated hose power consumption+120 C electronically stabilised~5 CHeated hose thermocouple wires360W (max)Water condensation by rapid cooling down+5 C electronically stabilisedK-type (S-type optionally)Cooler type~1 C100l/hBased on Peltier elementGAS CONDITIONING UNIT WITH CONDENSATION DRYER, HEATED HOSE DRIVER, HEATED HOSEP eristaltic, 38ml/minHeated hose length3m (optionally 5m or madur Polska sp. z TECHNICAL DATASENSORSEQUIPMENTAPPEARANCED rying methodCooler temperatureUnder pressure in Nafion collarReady to operate afterCondensate pumpHeated hose temperatureHeated hose temperature hysteresisHeated hose power consumption+120 C electronically stabilised~5 CHeated hose thermocouple wires360W (max)Water transfer through Nafion membrane driven by partial vapour pressure differential - first order kinetic reactionn/aK-type (S-type optionally)Cooler type~500 mbar1 minuteBased on Nafion exchangerGAS CONDITIONING UNIT WITH NAFION DRYER, HEATED HOSE DRIVER, HEATED HOSE n/aHeated hose length3m (optionally 5m or 10m)MethodRange | ResolutionAccuracyTime (T))

4 90 MEASUREMENTSV ariableT gas - gas temperature10 secK-type thermocouple 2 C-10 1000 C | 0,1 C T gas - gas temperature10 secS-type thermocouple 2 C-10 1500 C | 0,1 C T amb - boiler intake air temperature10 secPT500 resistive sensor 2 C-10 100 C | 0,1 C 10 secSilicon piezoresistive pressure sensor 2Pa abs. or 5% hPa +25 hPa |1 Pa (0,01hPa)10 secIndirect, with Pitot tube & pressure sensor0,3 m/s 5% 50 m/s | 0,1 m/sLambda - excess air number10 secCalculated 5% 10 | 0,01qA - stack loss10 secCalculated 5% 100% | 0,1%Eta - combustion efficiency10 secCalculated 5% 120% | 0,1%T & T external temperatures13T & T external temperatures13T & T external temperatures24 Differential pressureGas flow velocityU & U analogue inputs (voltage)12I & I analogue inputs (current)1210 secK-type thermocouple 2 C-10 1000 C | 0,1 C 10 secS-type thermocouple 2 C-10 1500 C | 0,1 C 10 secPT500 resistive sensor 2 C-10 100 C | 0,1 C 10 secDelta-sigma ADC 2 +20V | 0,01V 10 secDelta-sigma ADC 2 +20mA | 0,01mA madur Polska sp.

5 Z TECHNICAL DATASENSORSEQUIPMENTAPPEARANCENO - NITRIC OXIDEE lectrochemical sensor 5 000 ppm | 1 ppm 5 ppm abs. or 5% secElectrochemical sensor1 000 ppm | 1 ppm 5 ppm abs. or 5% secElectrochemical sensor5 000 ppm | 1 ppm 5 ppm abs. or 5% secCH TOTAL HYDROCARBONSxyNDIR5% | 0,01% 0,05% abs. or 5% secNDIR10% | 0,01% 0,05% abs. or 5% secNDIR25% | 0,01% 0,05% abs. or 5% secNDIR50% | 0,01% 0,05% abs. or 5% secNDIR100% | 0,1% 0,5% abs. or 5% secO - OXYGEN2CO - CARBON MONOXIDEE lectrochemical sensor20 000 ppm | 1 ppm 5 ppm abs. or 5% sec4 000 ppm | 1 ppm 5 ppm abs. or 5% secElectrochemical, partial pressure20,95% | 0,01% 0,1% abs. or 5% secElectrochemical, partial pressure25,00% | 0,01% 0,1% abs. or 5% sec10% | 10 ppm 50 ppm abs. or 5% sec10% | 0,01% 0,05% abs. or 5% sec100% | 0,1% 0,5% abs. Or 5% secElectrochemicalwith H compensation2 Electrochemical sensorNDIR (as a second sensor only)Electrochemical, partial pressure100,00% | 0,01% 0,1% abs.

6 Or 5% secNDIR (as a second sensor only)CO - CARBON DIOXIDE2 NDIR1% | 0,01% 0,05% abs. or 5% secNDIR5% | 0,01% 0,05% abs. or 5% secNDIR25% | 0,1% 0,05% abs. or 5% secElectrochemical sensor20,95% | 0,01% 0,1% abs. or 5% secNDIR100% | 0,1% 0,5% abs. or 5% secMethodRange | ResolutionAccuracyTime madur Polska sp. z TECHNICAL DATASENSORSEQUIPMENTAPPEARANCEE lectrochemical sensor1 000 ppm | 1 ppm 5 ppm abs. or 5% secHS- HYDROGEN SULPHIDE2 Electrochemical sensor2 000 ppm | 1 ppm 10 ppm abs. or 5% secH- HYDROGEN 2 Electrochemical sensor20 000 ppm | 1 ppm 10 ppm abs. or 5% secElectrochemical sensor250 ppm | 1 ppm 5 ppm abs. or 5% secCL- CHLORINE2 Electrochemical sensor100 ppm | 1 ppm 5 ppm abs. or 5% secHCl- HYDROGEN CHLORINE NDIR2,5% | 0,01% 0,05 ppm abs. or 5% secCHF- FLUOROFORM (REFRIGERANT R23)3 NDIR1% | 0,01% 0,05 ppm abs.

7 Or 5% secSO- SULPHUR DIOXIDE2 NDIR1% | 0,01% 0,05 ppm abs. or 5% secNO- NITROGEN DIOXIDE2 TECHNICAL DATAAPPEARANCEEQUIPMENTSTANDARD EQUIPMENT SUPPLIED ALONG WITH THE DEVICET hree modules installed in 19 12U cabinet:Power supply unitGas sample conditioning unit with selected type of gas dryerGas analyser unit with selected types of sensorsSENSORSA mbient temperature sensor with 3m cable and magnetic holderAnalyser's cartSD card for data-loggerSD card reader (USB type to connect with PC computer)CD with software and manualsMethodRange | ResolutionAccuracyTime (T)90 Conformity3m PVC hose for outlet madur Polska sp. z ADDITIONAL EQUIPMENT NECESSARY FOR THE ANALYSER TO WORKGas probe pipeExchangeable gas probe pipe mounted on the probe holder or heated hose with M30x1 has thermocouple type K (in some configurations type S) for measurement of gas temperature and a threaded fixing cone.

8 With the heated hose is a complete gas are many probe pipes available. They differ in length and working hoseHeated hose with heated gas filter supplies gas sample to the the analyser s conditioning module. Hose has M30x1 threaded connection to fix gas probe length of hose is 3m, it is possible to order other lengths of is especially advised when dealing with high humidity and and other gases highly reactive with , NO 22 Stationary gas probeGas probe designed specially for stationary purposes. Probe is available in different lengths and is equipped with suitable holder (different types are available).Optionally it may also be equipped with:- Thermocouple for measurements of gas Sintered stainless-steel filter (cleanable) - especially recommended when dealing with high concentration of dust and Blow-back cleaning option - valve that allows to switch between measured gas and the compressed air inlet that is used for cleaning the sintered filterHeated filter is installed right after the gas is best when it is paired with heated hose to prevent vapour from dust filter's size in the heated filter was enlarged (in comparison with a filter in the heated hose)to allow longer maintenance-free work.

9 OPTIONAL EQUIPMENTP itot tubePitot tube is an accessory that allows to perform measurement of the flow velocity of the gas stream. The measurement is performed indirectly Pitot tube is connected to analyser s differential pressure sensor. Analyser recalculates the differential pressure on the Pitot tube s outlets to few length of tubes are available. Pitot tube has 2m gas tubings to connect it with the analyser. It may be provided with a suitable holder for stationary madur Polska sp. z TECHNICAL DATASENSORSEQUIPMENTAPPEARANCEGAS ANALYSER MODULE - FRONT PANELE thernetZasilanieLCD DISPLAY QVGA (320*240) MONOCHROMATIC GAS SAMPLE CONDITIONING UNIT - FRONT PANELPOWER SUPPLY MODULE - FRONT PANELCONDENSATE PUMPFLOW INDICATORPOWER ON/ OFF SWITCH 2 GAS FILTERSPOWER ON/ OFF SWITCH KEYPAD, 16 KEYSFUNCTION KEYSSC CARD madur Polska sp. z 10 TECHNICAL DATASENSORSEQUIPMENTAPPEARANCEEXAMPLE PRINTSCREENS10 GAS AND ELECTRIC CONNECTORS PANEL - BOTTOM OF THE CABINET S BACK #: :13M+ :04XL1 CONO2SO2H2Cl2---NOH2 SNH3 HClNO---22 ppm13 ppm220 ppm45 ppm15 ppm--- ppm10 mg/m12 mg/m160 mg/m286 mg/m0 mg/m--- mg/m 333333 Temperature stabilizing Please C / C410:13M+(+)Options(-)M003F1T=2s0 % %--- --- --- %COCO2 TgasTambqACO IRNO IRSO2 IRNOx0 ppm0 ppm0 ppm0 ppm-180-120-60039 Pdif[Pa]4-6 DIFFERENTIAL pressure INLETSFRESH AIRCONDENSATE OUTLETAMBIENT TEMPERATURE SENSOR SOCKETRS232 CGAS OUTLET(WASTE GAS)AIR INTAKE FANS OF THE GAS DRYERCHARACTERISTICFEATURES


Related search queries